Comparison of Ambulatory Tonometric and Oscillometric Blood Pressure Monitoring in Hypertensive Patients

被引:6
作者
Hornstrup, Bodil Gade [1 ,2 ]
Rosenbaek, Jeppe Bakkestrom
Bech, Jesper Norgaard
机构
[1] Reg Hosp West Jutland, Univ Clin Nephrol & Hypertens, Holstebro, Denmark
[2] Aarhus Univ, Holstebro, Denmark
来源
INTEGRATED BLOOD PRESSURE CONTROL | 2020年 / 13卷
关键词
ambulatory blood pressure monitoring; oscillometric; tonometric; PROGNOSTIC VALUE; MANAGEMENT; MORTALITY; BURDEN;
D O I
10.2147/IBPC.S235228
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Aim: Correct measurement of blood pressure (BP) is important for optimal diagnosis and treatment of patients with hypertension. The aim of this study was to compare a wrist-worn device using tonometric measurements of BP to a conventional device using oscillometric measurements of 24 h BP, diagnosing of hypertension, and non-dipping. Methods: One-hundred patients in the Renal Outpatient Clinic had 24 h ambulatory BP monitoring performed with a tonometric device, BPro, and an oscillometric device, A&D, simultaneously. Results: Twenty-four-hour and daytime systolic BP was significantly lower using tonometric monitoring compared to oscillometric (7 and 6 mmHg, respectively, p< 0.001). In the population of patients diagnosed with hypertension, the tonometric device diagnosed 90% of patients with uncontrolled hypertension correctly (positive predictive value), whereas 49% of patients classified as normotensive were uncontrolled hypertensive (negative predictive value). The mean difference between relative nocturnal BP decrease between tonometric and oscillometric was 2+8% (p< 0.01), and 33% of patients classified as dippers were nondippers (negative predictive value). Conclusion: Using the BPro device for tonometric monitoring of BP and classification of hypertension and non-dipping in patients diagnosed with hypertension leads to misclassification of patients. Therefore, the BPro device is not suitable for clinical practice in hypertensive patients from a Renal Outpatient Clinic.
引用
收藏
页码:41 / 47
页数:7
相关论文
共 17 条
[1]   2003 World Health Organization (WHO)/International Society of Hypertension (ISH) statement on management of hypertension [J].
Afridi, I ;
Canny, J ;
Yao, CH ;
Christensen, B ;
Cooper, RS ;
Kadiri, S ;
Hill, S ;
Kaplan, N ;
Kuschnir, E ;
Lexchin, J ;
Mendis, S ;
Poulter, N ;
Psaty, BM ;
Rahn, KH ;
Sheps, SG ;
Whitworth, J ;
Yach, D ;
Bengoa, R ;
Ramsay, L ;
Kaplan, N ;
Mendis, S ;
Poulter, N ;
Whitworth, J .
JOURNAL OF HYPERTENSION, 2003, 21 (11) :1983-1992
[2]   Minimally sufficient numbers of measurements for validation of 24-hour blood pressure monitoring in chronic kidney disease [J].
Agarwal, Rajiv ;
Tu, Wanzhu .
KIDNEY INTERNATIONAL, 2018, 94 (06) :1199-1204
[3]   Superiority of ambulatory over clinic blood pressure measurement in predicting mortality - The Dublin Outcome Study [J].
Dolan, E ;
Stanton, A ;
Thijs, L ;
Hinedi, K ;
Atkins, N ;
McClory, S ;
Den Hond, E ;
McCormack, P ;
Staessen, JA ;
O'Brien, E .
HYPERTENSION, 2005, 46 (01) :156-161
[4]   BP AS A DETERMINANT OF CARDIAC LEFT-VENTRICULAR MUSCLE MASS [J].
DRAYER, JIM ;
WEBER, MA ;
DEYOUNG, JL .
ARCHIVES OF INTERNAL MEDICINE, 1983, 143 (01) :90-92
[5]   Selected major risk factors and global and regional burden of disease [J].
Ezzati, M ;
Lopez, AD ;
Rodgers, A ;
Vander Hoorn, S ;
Murray, CJL .
LANCET, 2002, 360 (9343) :1347-1360
[6]   Understanding Bland Altman analysis [J].
Giavarina, Davide .
BIOCHEMIA MEDICA, 2015, 25 (02) :141-151
[7]   Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use [J].
Komori, Takahiro ;
Eguchi, Kazuo ;
Hoshide, Satoshi ;
Williams, Bryan ;
Kario, Kazuomi .
BLOOD PRESSURE MONITORING, 2013, 18 (01) :57-62
[8]   Global burden of blood-pressure-related disease, 2001 [J].
Lawes, Carlene M. M. ;
Vander Hoorn, Stephen ;
Rodgers, Anthony .
LANCET, 2008, 371 (9623) :1513-1518
[9]  
Lewington S, 2002, LANCET, V360, P1903, DOI 10.1016/S0140-6736(02)11911-8
[10]   2013 ESH/ESC Practice Guidelines for the Management of Arterial Hypertension [J].
Mancia, Giuseppe ;
Fagard, Robert ;
Narkiewicz, Krzysztof ;
Redon, Josep ;
Zanchetti, Alberto ;
Boehm, Michael ;
Christiaens, Thierry ;
Cifkova, Renata ;
De Backer, Guy ;
Dominiczak, Anna ;
Galderisi, Maurizio ;
Grobbee, Diederick E. ;
Jaarsma, Tiny ;
Kirchhof, Paulus ;
Kjeldsen, Sverre E. ;
Laurent, Stephane ;
Manolis, Athanasios J. ;
Nilsson, Peter M. ;
Ruilope, Luis Miguel ;
Schmieder, Roland E. ;
Sirnes, Per Anton ;
Sleight, Peter ;
Viigimaa, Margus ;
Waeber, Bernard ;
Zannad, Faiez .
BLOOD PRESSURE, 2014, 23 (01) :3-16