Optimization of extracranial stereotactic radiation therapy of small lung lesions using accurate dose calculation algorithms

被引:39
作者
Dobler, Barbara [1 ,2 ]
Walter, Cornelia [1 ]
Knopf, Antje [1 ]
Fabri, Daniella [1 ]
Loeschel, Rainer [3 ]
Polednik, Martin [1 ]
Schneider, Frank [1 ]
Wenz, Frederik [1 ]
Lohr, Frank [1 ]
机构
[1] Heidelberg Univ, Mannheim Med Ctr, Dept Radiat Oncol, D-6800 Mannheim, Germany
[2] Univ Regensburg, Med Ctr, Dept Radiotherapy, Regensburg, Germany
[3] Univ Appl Sci, Dept Math & Comp Sci, Regensburg, Germany
关键词
D O I
10.1186/1748-717X-1-45
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The aim of this study was to compare and to validate different dose calculation algorithms for the use in radiation therapy of small lung lesions and to optimize the treatment planning using accurate dose calculation algorithms. Methods: A 9-field conformal treatment plan was generated on an inhomogeneous phantom with lung mimics and a soft tissue equivalent insert, mimicking a lung tumor. The dose distribution was calculated with the Pencil Beam and Collapsed Cone algorithms implemented in Masterplan (Nucletron) and the Monte Carlo system XVMC and validated using Gafchromic EBT films. Differences in dose distribution were evaluated. The plans were then optimized by adding segments to the outer shell of the target in order to increase the dose near the interface to the lung. Results: The Pencil Beam algorithm overestimated the dose by up to 15% compared to the measurements. Collapsed Cone and Monte Carlo predicted the dose more accurately with a maximum difference of -8% and -3% respectively compared to the film. Plan optimization by adding small segments to the peripheral parts of the target, creating a 2-step fluence modulation, allowed to increase target coverage and homogeneity as compared to the uncorrected 9 field plan. Conclusion: The use of forward 2-step fluence modulation in radiotherapy of small lung lesions allows the improvement of tumor coverage and dose homogeneity as compared to non-modulated treatment plans and may thus help to increase the local tumor control probability. While the Collapsed Cone algorithm is closer to measurements than the Pencil Beam algorithm, both algorithms are limited at tissue/lung interfaces, leaving Monte-Carlo the most accurate algorithm for dose prediction.
引用
收藏
页数:11
相关论文
共 25 条
[1]  
Aarup L, 2006, RADIOTHER ONCOL, V81, pS99
[2]   A PENCIL BEAM MODEL FOR PHOTON DOSE CALCULATION [J].
AHNESJO, A ;
SAXNER, M ;
TREPP, A .
MEDICAL PHYSICS, 1992, 19 (02) :263-273
[3]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[4]   Dose calculations for external photon beams in radiotherapy [J].
Ahnesjö, A ;
Aspradakis, MM .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (11) :R99-R155
[5]   Experimental verification of convolution /superposition photon dose calculations for radiotherapy treatment planning [J].
Aspradakis, MM ;
Morrison, RH ;
Richmond, ND ;
Steele, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (17) :2873-2893
[6]   Verification of lung dose in an anthropomorphic phantom calculated by the collapsed cone convolution method [J].
Butson, MJ ;
Elferink, R ;
Cheung, T ;
Yu, PKN ;
Stokes, M ;
Quach, KY ;
Metcalfe, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :N143-N149
[7]   Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (01) :60-62
[8]   Absorption spectra variations of EBT radiochromic film from radiation exposure [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (13) :N135-N140
[9]   Post-irradiation colouration of Gafchromic EBT radiochromic film [J].
Cheung, T ;
Butson, MJ ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (20) :N281-N285
[10]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253