A Simple Acquisition Strategy to Avoid Off-Resonance Blurring in Spiral Imaging with Redundant Spiral-In/Out k-Space Trajectories

被引:15
作者
Fielden, Samuel W. [1 ]
Meyer, Craig H. [1 ,2 ]
机构
[1] Univ Virginia, Dept Biomed Engn, Charlottesville, VA 22908 USA
[2] Univ Virginia, Dept Radiol, Charlottesville, VA 22908 USA
基金
美国国家卫生研究院;
关键词
spiral imaging; off-resonance correction; CONCOMITANT GRADIENT FIELD; BOLD FMRI; INHOMOGENEITY CORRECTION; RECONSTRUCTION; MRI; MAP;
D O I
10.1002/mrm.25172
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThe major hurdle to widespread adoption of spiral trajectories has been their poor off-resonance performance. Here we present a self-correcting spiral k-space trajectory that avoids much of the well-known spiral blurring during data acquisition. Theory and MethodsIn comparison with a traditional spiral-out trajectory, the spiral-in/out trajectory has improved off-resonance performance. By combining two spiral-in/out acquisitions, one rotated 180 degrees in k-space compared with the other, multishot spiral-in/out artifacts are eliminated. A phantom was scanned with the center frequency manually tuned 20, 40, 80, and 160 Hz off-resonance with both a spiral-out gradient echo sequence and the redundant spiral-in/out sequence. The phantom was also imaged in an oblique orientation in order to demonstrate improved concomitant gradient field performance of the sequence. Additionally, the trajectory was incorporated into a spiral turbo spin echo sequence for brain imaging. ResultsPhantom studies with manually tuned off-resonance agree well with theoretical calculations, showing that moderate off-resonance is well-corrected by this acquisition scheme. Blur due to concomitant fields is reduced, and good results are obtained in vivo. ConclusionThe redundant spiral-in/out trajectory results in less image blur for a given readout length than a traditional spiral-out scan, reducing the need for complex off-resonance correction algorithms. Magn Reson Med 73:704-710, 2015. (c) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:704 / 710
页数:7
相关论文
共 23 条
[1]  
Bernstein M.A., 2004, HDB MRI PULSE SEQUEN, DOI DOI 10.1016/B978
[2]   Fast Conjugate Phase Image Reconstruction Based on a Chebyshev Approximation to Correct for B0 Field Inhomogeneity and Concomitant Gradients [J].
Chen, Weitian ;
Sica, Christopher T. ;
Meyer, Craig H. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (05) :1104-1111
[3]   Semiautomatic off-resonance correction in spiral imaging [J].
Chen, Weitian ;
Meyer, Craig H. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (05) :1212-1219
[4]   Fast Concomitant Gradient Field and Field Inhomogeneity Correction for Spiral Cardiac Imaging [J].
Cheng, Joseph Y. ;
Santos, Juan M. ;
Pauly, John M. .
MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (02) :390-401
[5]   Spiral demystified [J].
Delattre, Benedicte M. A. ;
Heidemann, Robin M. ;
Crowe, Lindsey A. ;
Vallee, Jean-Paul ;
Hyacinthe, Jean-Noel .
MAGNETIC RESONANCE IMAGING, 2010, 28 (06) :862-881
[6]  
Feng X, 2011, P 19 ANN M ISMRM STO, P2620
[7]   Noncontrast Peripheral MRA with Spiral Echo Train Imaging [J].
Fielden, Samuel W. ;
Mugler, John P., III ;
Hagspiel, Klaus D. ;
Norton, Patrick T. ;
Kramer, Christopher M. ;
Meyer, Craig H. .
MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (03) :1026-1033
[8]   Spiral imaging in fMRI [J].
Glover, Gary H. .
NEUROIMAGE, 2012, 62 (02) :706-712
[9]   Spiral-in/out BOLD fMRI for increased SNR and reduced susceptibility artifacts [J].
Glover, GH ;
Law, CS .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (03) :515-522
[10]   Inhomogeneity correction using an estimated linear field map [J].
Irarrazabal, P ;
Meyer, CH ;
Nishimura, DG ;
Macovski, A .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (02) :278-282