A free-air ionization chamber with a large aperture diaphragm

被引:1
|
作者
Takata, N. [1 ]
机构
[1] AIST, NMIJ, Ionizing Radiat Sect, Tsukuba, Ibaraki 3058568, Japan
关键词
Parallel-plate free-air ionization chamber; Ionization volume; Charge collection volume; Electric field distortion; Guard electrode; Scattered X-rays; Air kerma; Exposure; X-RAYS;
D O I
10.1016/j.nima.2010.02.083
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Calculations of the electric field distributions in free-air ionization chambers reveal that the distortion of the charge collection volume is small even for wide X-ray beam passage if the diaphragm and the X-ray shielding box are kept at a potential equal to half that applied to the high-voltage electrode. Applying this potential to the diaphragm and the shielding box permits a larger aperture diaphragm to be used. This will allow a wider X-ray beam to enter the chamber, thus generating a larger signal. In addition, the distance between the diaphragm and the charge collection volume can be shortened to reduce the amount of X-ray attenuation. It is also possible to calibrate a dosimeter against a free-air ionization chamber that has a diaphragm whose aperture size is equal to the size of the dosimeter in an X-ray field that is collimated to the same size. This is important since free-air ionization chambers are not sensitive to X-rays that are incident at large angles, such as those scattered by the collimator, filters and air. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:216 / 220
页数:5
相关论文
共 30 条
  • [21] Enhanced ozone strongly reduces the carbon sink strength of adult beech - resume from the free-air fumigation study at Kranzberg Forest
    Matyssek, R.
    GEFAHRSTOFFE REINHALTUNG DER LUFT, 2011, 71 (03): : 103 - 107
  • [22] Responses of Antioxidant Enzymes to Chronic Free-Air Ozone Stress in Rice (Oryza sativa L.) Cultivars with Different Ozone-Sensitivities
    Wang, JunLi
    Liu, Gang
    Liu, FuXing
    Zhu, JianGuo
    BULLETIN OF ENVIRONMENTAL CONTAMINATION AND TOXICOLOGY, 2019, 103 (03) : 428 - 434
  • [23] Impact of Elevated Ozone on Nutrient Uptake and Utilization of Chinese Hybrid Indica Rice (Oryza Sativa) Cultivars under Free-Air Ozone Enrichment
    Kou, T. J.
    Xu, G. W.
    Zhu, J. G.
    COMMUNICATIONS IN SOIL SCIENCE AND PLANT ANALYSIS, 2017, 48 (06) : 635 - 645
  • [24] Evaluation of an in-duct bipolar ionization device on particulate matter and gas-phase constituents in a large test chamber
    Zeng, Yicheng
    Heidarinejad, Mohammad
    Stephens, Brent
    BUILDING AND ENVIRONMENT, 2022, 213
  • [25] Enhanced ozone strongly reduces carbon sink strength of adult beech (Fagus sylvatica) - Resume from the free-air fumigation study at Kranzberg Forest
    Matyssek, R.
    Wieser, G.
    Ceulemans, R.
    Rennenberg, H.
    Pretzsch, H.
    Haberer, K.
    Loew, M.
    Nunn, A. J.
    Werner, H.
    Wipfler, P.
    Osswaldg, W.
    Nikolova, P.
    Hanke, D. E.
    Kraigher, H.
    Tausz, M.
    Bahnweg, G.
    Kitao, M.
    Dieler, J.
    Sandermann, H.
    Herbinger, K.
    Grebenc, T.
    Blumenroether, M.
    Deckmyn, G.
    Grams, T. E. E.
    Heerdt, C.
    Leuchner, M.
    Fabian, P.
    Haeberle, K. -H.
    ENVIRONMENTAL POLLUTION, 2010, 158 (08) : 2527 - 2532
  • [26] Open top chamber and free air CO2 enrichment - approaches to investigate tree responses to elevated CO2
    Machacova, K.
    IFOREST-BIOGEOSCIENCES AND FORESTRY, 2010, 3 : 102 - 105
  • [27] Study on the lower absolute measurement limit of air kerma from 137Cs source using a large-size cavity chamber
    Teng, Zhongbin
    Song, Mingzhe
    Ni, Ning
    Wei, Kexin
    Liu, Yuntao
    Liu, Senlin
    RADIATION PHYSICS AND CHEMISTRY, 2021, 183
  • [28] Intra-annual growth response of adult Norway spruce (Picea abies [L.] KARST.) and European beech (Fagus sylvatica L.) to an experimentally enhanced, free-air ozone regime
    Wipfler, Philip
    Seifert, Thomas
    Biber, Peter
    Pretzsch, Hans
    EUROPEAN JOURNAL OF FOREST RESEARCH, 2009, 128 (02) : 135 - 144
  • [29] Practical dosimetry procedure of air kerma for kilovoltage X-ray imaging in radiation oncology using a 0.6-cc cylindrical ionization chamber with a cobalt absorbed dose-to-water calibration coefficient
    Hidenobu Tachibana
    Ryo Takahashi
    Takayuki Kogure
    Shiro Nishiyama
    Tomoyuki Kurosawa
    Radiological Physics and Technology, 2022, 15 : 264 - 270
  • [30] Practical dosimetry procedure of air kerma for kilovoltage X-ray imaging in radiation oncology using a 0.6-cc cylindrical ionization chamber with a cobalt absorbed dose-to-water calibration coefficient
    Tachibana, Hidenobu
    Takahashi, Ryo
    Kogure, Takayuki
    Nishiyama, Shiro
    Kurosawa, Tomoyuki
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2022, 15 (03) : 264 - 270