Power measurement standards for high-power lasers:: comparison between the NIST and the PTB

被引:12
作者
Li, X
Scott, TR
Cromer, CL
Keenan, D
Brandt, F
Möstl, K
机构
[1] Natl Inst Stand & Technol, Boulder, CO 80303 USA
[2] PTB, D-38116 Braunschweig, Germany
关键词
D O I
10.1088/0026-1394/37/5/22
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report the results of the first laser high-power measurement comparison between the Physikalisch-Technische Bundesanstalt (PTB, Germany), and the National Institute of Standards and Technology (NIST, USA). Laser power transfer standards were calibrated at both national standards laboratories between 82 W and 127 W at 1.06 mum and between 85 W and 554 W at 10.6 mum. Relative agreement between the standards of the two laboratories was demonstrated to lie between 5 x 10(-3) and 7 x 10(-3), which is well within the combined uncertainties.
引用
收藏
页码:445 / 447
页数:3
相关论文
共 5 条
[1]  
BRANDT F, 1996, LASER FORSCHUNG TECH, P730
[2]   FAST DETERMINATION OF THE NONLINEARITY OF PHOTODETECTORS [J].
COSLOVI, L ;
RIGHINI, F .
APPLIED OPTICS, 1980, 19 (18) :3200-3203
[3]  
MOSTL K, 1989, NEW DEV APPL OPTICAL, V92, P11
[4]  
Scott T. R., 1988, Proceedings of the SPIE - The International Society for Optical Engineering, V888, P48, DOI 10.1117/12.944232
[5]  
WEST ED, 1977, 685 NBS