A microassembled low-profile three-dimensional microelectrode array for neural prosthesis applications

被引:61
作者
Yao, Ying [1 ]
Gulari, Mayurachat Ning
Wiler, James A.
Wise, Kensall D.
机构
[1] Univ Michigan, Engn Res Ctr Wireless Integrat MicroSyst, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Kresge Hearing Res Inst, Dept Otolaryngol, Ann Arbor, MI 48109 USA
基金
美国国家科学基金会;
关键词
biomedical electrodes; microassembly;
D O I
10.1109/JMEMS.2007.896712
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes the design and microassembly process of a low-profile 3-D microelectrode array for mapping the functional organization of targeted areas of the central nervous system and for possible application in neural prostheses. The array consists of multiple planar complimentary metal-oxide-semiconductor stimulating probes and 3-D assembly components. Parylene-encapsulated gold beams supported by etch-stopped silicon braces allow the backends of the probes to be folded over to reduce the height of the array above the cortical surface. A process permitting parylene to be used at wafer level with bulk-silicon wet release has been reported. Spacers are used to fix the microassembled probes in position and are equipped with interlocking structures to facilitate the assembly process and increase yield. Four-probe 256-site 3-D arrays operate from 5 V with an average per-channel power dissipation of 97 mu W at full range stimulation with pulse widths of 100 mu s at 500-Hz frequency. Thirty-two sites can be stimulated simultaneously with maximum currents of 127 mu A and a current resolution of 1 mu A. The microassembly techniques allow a variety of 3-D microstructures to be created from planar components.
引用
收藏
页码:977 / 988
页数:12
相关论文
共 32 条
[1]   BATCH-FABRICATED THIN-FILM ELECTRODES FOR STIMULATION OF THE CENTRAL AUDITORY-SYSTEM [J].
ANDERSON, DJ ;
NAJAFI, K ;
TANGHE, SJ ;
EVANS, DA ;
LEVY, KL ;
HETKE, JF ;
XUE, XL ;
ZAPPIA, JJ ;
WISE, KD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (07) :693-704
[2]   Single-unit neural recording with active microelectrode arrays [J].
Bai, Q ;
Wise, KD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (08) :911-920
[3]  
BAI Q, 1999, THESIS U MICHIGAN AN
[4]   Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays [J].
Biran, R ;
Martin, DC ;
Tresco, PA .
EXPERIMENTAL NEUROLOGY, 2005, 195 (01) :115-126
[5]  
Chapin J. K., 2000, NEURAL PROSTHESES RE
[6]  
DECHEV N, 2004, P SPIE MICR MICR PHO, P134
[7]   PERFORMANCE OF PLANAR MULTISITE MICROPROBES IN RECORDING EXTRACELLULAR SINGLE-UNIT INTRACORTICAL ACTIVITY [J].
DRAKE, KL ;
WISE, KD ;
FARRAYE, J ;
ANDERSON, DJ ;
BEMENT, SL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (09) :719-732
[8]   Microelectrode array fabrication by electrical discharge machining and chemical etching [J].
Fofonoff, TA ;
Martel, SM ;
Hatsopoulos, NG ;
Donoghue, JP ;
Hunter, IW .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (06) :890-895
[9]  
FRANK K, 1964, PHYS TECHNIQUES BIOL, V5
[10]   A BiCMOS wireless interface chip for micromachined stimulating microprobes [J].
Ghovanloo, M ;
Beach, K ;
Wise, KD ;
Najafi, K .
2ND ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2002, :277-282