Reference change value and measurement uncertainty in the evaluation of tumor markers

被引:3
作者
Yavuz, Hatice Bozkurt [1 ]
Bildirici, Mehmet Akif [2 ]
Yaman, Huseyin [3 ]
Karahan, Suleyman Caner [3 ]
Aliyazicioglu, Yuksel [3 ]
Orem, Asim [3 ]
机构
[1] Sebinkarahisar State Hosp, Giresun, Turkey
[2] Dogubayazit Doc Dr Yasar Eryilmaz State Hosp, Agri, Turkey
[3] Karadeniz Tech Univ, Dept Clin Biochem, Fac Med, Trabzon, Turkey
关键词
Uncertainty; reference values; tumor markers; reference change value AFP; CA; 125; 15-3; 19-9; CEA; BIOLOGICAL VARIATION; OVARIAN-CANCER; SERUM; CA-19-9; CA-125; LIMITS;
D O I
10.1080/00365513.2021.1979244
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The use of measurement uncertainty among clinical laboratories becomes widespread. Measurement uncertainty can be reported with the result, as well as be used in certain reference change value (RCV) calculation equations. RCV is especially recommended for use in tests with a low individuality index. In our study, we calculated the measurement uncertainty of AFP, CA 125, CA 15-3, CA 19-9, CEA tumor markers with the ISO TS 20914:2019. We compared results with limits. Two Beckman Coulter DXI-800 (Minnesota, USA) autoanalysers' results were used. We calculated the RCV values using the classical Fraser method, logarithmic Lund Method, and Clinical Laboratory Standards Institute (CLSI) method as Minimal Difference (MD). We found the same permissible measurement uncertainty limit as 15.97% for all five tumor markers. The highest RCV value was found as 90% upstream for AFP test with Lund logarithmic approach, the lowest RCV value was found as 12% for CEA with MD, all other RCV results were between these two values. We do not recommend the use of MD, as values for Biological variation are not used in the MD approach. We also recommend using the logarithmic approach, although it gives higher results. There are also clinical studies on the significance of tumor markers in a follow-up that show different results. These differences may be because the studies are conducted with different systems. Therefore, each laboratory needs to calculate its own RCV values. We also recommend informing the clinicians about the tests with high measurement uncertainty.
引用
收藏
页码:601 / 605
页数:5
相关论文
共 29 条
  • [11] European Federation of Clinical Chemistry and Laboratory Medicine, LIST BV EST
  • [12] Reference change values for brain natriuretic peptides revisited
    Fokkema, M. Rebecca
    Herrmann, Zuzana
    Muskiet, Frits A. J.
    Moecks, Joachim
    [J]. CLINICAL CHEMISTRY, 2006, 52 (08) : 1602 - 1603
  • [13] Fraser CG, 2012, CLIN CHEM LAB MED, V50, P807, DOI 10.1515/CCLM.2011.733
  • [14] FRASER CG, 1989, CRIT REV CL LAB SCI, V27, P409
  • [15] Permissible limits for uncertainty of measurement in laboratory medicine
    Haeckel, Rainer
    Wosniok, Werner
    Gurr, Ebrhard
    Peil, Burkhard
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (08) : 1161 - 1171
  • [16] HARRIS EK, 1983, CLIN CHEM, V29, P25
  • [17] CA-125 IN GYNECOLOGICAL PATHOLOGY - A REVIEW
    KENEMANS, P
    YEDEMA, CA
    BON, GG
    VONMENSDORFFPOUILLY, S
    [J]. EUROPEAN JOURNAL OF OBSTETRICS GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 1993, 49 (1-2): : 115 - 124
  • [18] False-Positive Elevations of Carcinoembryonic Antigen in Patients With a History of Resected Colorectal Cancer
    Litvak, Anya
    Cercek, Andrea
    Segal, Neil
    Reidy-Lagunes, Diane
    Stadler, Zsofia K.
    Yaeger, Rona D.
    Kemeny, Nancy E.
    Weiser, Martin R.
    Pessin, Melissa S.
    Saltz, Leonard
    [J]. JOURNAL OF THE NATIONAL COMPREHENSIVE CANCER NETWORK, 2014, 12 (06): : 907 - 913
  • [19] Calculation of limits for significant unidirectional changes in two or more serial results of a biomarker based on a computer simulation model
    Lund, Flemming
    Petersen, Per Hyltoft
    Fraser, Callum G.
    Soletormos, Gyorgy
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 2015, 52 (02) : 237 - 244
  • [20] Meier W, 1997, ANTICANCER RES, V17, P3019