Optimising the Windkessel Model for Cardiac Output Monitoring During Changes in Vascular Tone

被引:0
|
作者
Charlton, Peter [1 ,2 ,3 ]
Smith, John [2 ]
Camporota, Luigi [2 ]
Beale, Richard [2 ,3 ]
Alastruey, Jordi [1 ]
机构
[1] Kings Coll London, Dept Biomed Engn, London WC2R 2LS, England
[2] Guys & St Thomas NHS Fdn Trust, Dept Crit Care Med, London, England
[3] Kings Coll London, Sch Med, Div Asthma Allergy & Lung Biol, London WC2R 2LS, England
来源
2014 36TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC) | 2014年
关键词
ARTERIAL-BLOOD PRESSURE; VOLUME; PULSE;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Algorithms for estimating cardiac output (CO) from the arterial blood pressure wave have been observed to be inaccurate during changes in vascular tone. Many such algorithms are based on the Windkessel model of the circulation. We investigated the optimal analytical approaches and assumptions that make up each algorithm during changes in vascular tone. Several analytical approaches and assumptions were evaluated on data from 15 critically ill patients by comparison with thermodilution measurements. We found that the most accurate algorithms assumed a constant compliance for the duration of the beat. They produced a percentage error of +/- 31% by maintaining the compliance and outflow terms in the Windkessel model. For any algorithm, the following assumptions gave highest accuracy: (i) outflow pressure into the microcirculation is zero; (ii) end of systole is identified using the second derivative of pressure. None of the tested algorithms reached the clinically acceptable accuracy of +/- 30%.
引用
收藏
页码:3759 / 3762
页数:4
相关论文
共 12 条
  • [1] Monitoring cardiac output during hyperbaric oxygen treatment of haemodynamically unstable patients
    Hansen, Marco Bo
    Treschow, Frederik
    Skielboe, Martin
    Hyldegaard, Ole
    Jansen, Erik Christian
    Nielsen, Jonas Bille
    DIVING AND HYPERBARIC MEDICINE, 2013, 43 (01) : 3 - 8
  • [2] Serial changes in cardiac output during normal pregnancy: A Doppler ultrasound study
    Hennessy, TG
    MacDonald, D
    Hennessy, MS
    Maguire, M
    Blake, S
    McCann, HA
    Sugrue, DD
    EUROPEAN JOURNAL OF OBSTETRICS GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 1996, 70 (02): : 117 - 122
  • [3] Changes in Cardiac Output and Perfusion during Hemodialysis and Hemodiafiltration Treatments Determined by Cardiac Magnetic Resonance Imaging
    Blankestijn, Peter J.
    Davenport, Andrew
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2017, 28 (04): : 1013 - 1015
  • [4] A Pilot and Feasibility Study of Continuous Cardiac Output and Blood Pressure Monitoring during Intermittent Hemodialysis
    Spano, Sofia
    Maeda, Akinori
    Lam, Joey
    Chaba, Anis
    Phongphithakchai, Atthaphong
    Pattamin, Nuttapol
    Hikasa, Yukiko
    See, Emily
    Mount, Peter
    Bellomo, Rinaldo
    BLOOD PURIFICATION, 2024, 53 (11-12) : 928 - 936
  • [5] Cardiac output and peripheral vascular resistance during normotensive and hypertensive pregnancy - a systematic review and meta-analysis
    Mulder, E. G.
    de Haas, S.
    Mohseni, Z.
    Schartmann, N.
    Abo Hasson, F.
    Alsadah, F.
    van Kuijk, S. M. J.
    van Drongelen, J.
    Spaanderman, M. E. A.
    Ghossein-Doha, C.
    BJOG-AN INTERNATIONAL JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2022, 129 (05) : 696 - 707
  • [6] Relationship Between Plethysmographic Waveform Changes and Hemodynamic Variables in Anesthetized, Mechanically Ventilated Patients Undergoing Continuous Cardiac Output Monitoring
    Thiele, Robert H.
    Colquhoun, Douglas A.
    Patrie, James
    Nie, Sarah H.
    Huffmyer, Julie L.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2011, 25 (06) : 1044 - 1050
  • [7] A multimodal stacked ensemble model for cardiac output prediction utilizing cardiorespiratory interactions during general anesthesia
    Dervishi, Albion
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [8] Hemodynamic profiles by non-invasive monitoring of cardiac index and vascular tone in acute heart failure patients in the emergency department: External validation and clinical outcomes
    Harrison, Nicholas Eric
    Meram, Sarah
    Li, Xiangrui
    White, Morgan B.
    Henry, Sarah
    Gupta, Sushane
    Zhu, Dongxiao
    Pang, Peter
    Levy, Phillip
    PLOS ONE, 2022, 17 (03):
  • [9] Cardiac Output Monitoring by Long Time Interval Analysis of a Radial Arterial Blood Pressure Waveform with Correction for Arterial Compliance Changes using Pulse Transit Time
    Moslehpour, Mohsen
    Zhang, Guanqun
    Mukkamala, Ramakrishna
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 5496 - 5498
  • [10] A pilot feasibility study of continuous cardiac output and blood pressure monitoring during intermittent hemodialysis in patients recovering from severe acute kidney injury
    Spano, Sofia
    Maeda, Akinori
    Lam, Joey
    Chaba, Anis
    Phongphithakchai, Atthaphong
    Hikasa, Yukiko
    Pattamin, Nuttapol
    Kitisin, Nuanprae
    See, Emily
    Mount, Peter
    Bellomo, Rinaldo
    JOURNAL OF CRITICAL CARE, 2025, 88