Intermittent, erratic behaviour of an implantable cardioverter defibrillator secondary to a hidden magnetic source of interference

被引:7
作者
Beinart, Roy [1 ]
Guy, Mary L. [1 ]
Ellinor, Patrick T. [1 ]
机构
[1] Massachusetts Gen Hosp, Cardiac Arrhythmia Serv, Boston, MA 02114 USA
来源
EUROPACE | 2011年 / 13卷 / 10期
基金
美国国家卫生研究院;
关键词
D O I
10.1093/europace/eur152
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Interference between pacemakers or implantable cardioverter-defibrillators (ICDs) and electromagnetic fields are of great concern due to the potential harmful influence on the function of the device. We report a case of intermittent, erratic behaviour of an ICD secondary to electromagnetic interference (EMI) caused by small magnets that were incorporated into the patients clothing. © The Author 2011.
引用
收藏
页码:1508 / 1509
页数:2
相关论文
共 3 条
[1]  
Ryf Salome, 2008, Technol Health Care, V16, P13
[2]   Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators [J].
Wolber, Thomas ;
Ryf, Salome ;
Binggeli, Christian ;
Holzmeister, Johannes ;
Brunckhorst, Corinna ;
Luechinger, Roger ;
Duru, Firat .
HEART RHYTHM, 2007, 4 (01) :1-4
[3]  
Yerra Lakshminarayan, 2007, Cardiol Rev, V15, P304, DOI 10.1097/CRD.0b013e31813e0ba9