Effectiveness of anti-oxidant polyethylene: What early retrievals can tell us

被引:18
作者
Currier, Barbara H. [1 ]
Currier, John H. [1 ]
Holdcroft, Lindsay A. [1 ]
Van Citters, Douglas W. [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
关键词
antioxidant polyethylene; retrieval analysis; in vivo oxidation; uniaxial tensile toughness; comparison of retrieved conventional; highly cross-linked and antioxidant tibial inserts; MOLECULAR-WEIGHT POLYETHYLENE; CROSS-LINKED POLYETHYLENE; TOTAL HIP-ARTHROPLASTY; CRACK PROPAGATION RESISTANCE; RANDOMIZED CONTROLLED-TRIAL; IN-VIVO OXIDATION; VITAMIN-E; MECHANICAL-PROPERTIES; IRRADIATED UHMWPE; ALPHA-TOCOPHEROL;
D O I
10.1002/jbm.b.33840
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The optimum UHMWPE orthopedic implant bearing surface must balance wear, oxidation, and fatigue resistance. Antioxidant polyethylene addresses free radicals, resulting from irradiation used in cross-linking, that could oxidize and potentially lead to fatigue damage under cycles of in vivo use. This study evaluates what short-term antioxidant UHMWPE retrievals can reveal about: (1) oxidation-resistance and (2) fatigue-resistance of these new materials. Retrievals of three different antioxidant polyethylene materials (n=25) were analyzed by FTIR and uniaxial tensile tests and compared to conventional (n=20) and remelted highly cross-linked (n=30) polyethylene retrievals of similar in vivo duration. Maximum oxidation values differed significantly across material types (p=0.018). No antioxidant retrieval exhibited a subsurface oxidation peak, in contrast to conventional gamma-sterilized (55%) and highly cross-linked (37%) retrievals that exhibited subsurface oxidation peaks over the same in vivo time. Trans-vinylene index (TVI) correlated positively with nominal irradiation dose (p<0.001). Tensile toughness correlated negatively with increasing TVI (p<0.001). The antioxidant materials in this study prevented in vivo oxidation more effectively than remelted HXL polyethylene at least over the in vivo period represented. The comparison of antioxidant retrieval tensile properties can be used as a guide for clinicians in choosing appropriate materials for the applications represented by their patients. (c) 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 353-359, 2018.
引用
收藏
页码:353 / 359
页数:7
相关论文
共 39 条
[1]  
[Anonymous], 2006, 882 ASTM D
[2]  
[Anonymous], 2010, D63810 ASTM
[3]  
[Anonymous], 2013, F2633-13, P1, DOI DOI 10.1520/F2792-12A.2
[4]  
ASTM, 2010, ASTM F238110 STAND T
[5]  
ASTM International, 2008, F13608E1 ASTM INT, P1, DOI [10.1520/F2066, DOI 10.1520/F036-08]
[6]   Gamma-irradiation Sterilization in an Inert Environment: A Partial Solution [J].
Berry, Daniel J. ;
Currier, Barbara H. ;
Mayor, Michael B. ;
Collier, John P. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (07) :1805-1813
[7]   KINETIC STUDIES IN THE CHEMISTRY OF RUBBER AND RELATED MATERIALS .1. THE THERMAL OXIDATION OF ETHYL LINOLEATE [J].
BOLLAND, JL .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1946, 186 (1005) :218-236
[8]   Improving synthetic hindered phenol antioxidants:: learning from vitamin E [J].
Breese, KD ;
Lamèthe, JF ;
DeArmitt, C .
POLYMER DEGRADATION AND STABILITY, 2000, 70 (01) :89-96
[9]   Comparison of cross-linked polyethylene materials for orthopaedic applications [J].
Collier, JP ;
Currier, BH ;
Kennedy, FE ;
Currier, JH ;
Timmins, GS ;
Jackson, SK ;
Brewer, RL .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (414) :289-304
[10]   Oxidation in orthopaedic UHMWPE sterilized by gamma-radiation and ethylene oxide [J].
Costa, L ;
Luda, MP ;
Trossarelli, L ;
del Prever, EMB ;
Crova, M ;
Gallinaro, P .
BIOMATERIALS, 1998, 19 (7-9) :659-668