Multispectral imaging of the ocular fundus using light emitting diode illumination

被引:59
作者
Everdell, N. L. [1 ]
Styles, I. B. [2 ]
Calcagni, A. [2 ]
Gibson, J. [3 ]
Hebden, J. [1 ]
Claridge, E. [2 ]
机构
[1] UCL, Dept Med Phys, London WC1E 6BT, England
[2] Univ Birmingham, Sch Comp Sci, Birmingham B15 2TT, W Midlands, England
[3] Aston Univ, Sch Life & Hlth Sci, Birmingham B4 7ET, W Midlands, England
基金
英国工程与自然科学研究理事会;
关键词
biomedical optical imaging; CCD image sensors; diseases; eye; light emitting diodes; DISEASES;
D O I
10.1063/1.3478001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We present an imaging system based on light emitting diode (LED) illumination that produces multispectral optical images of the human ocular fundus. It uses a conventional fundus camera equipped with a high power LED light source and a highly sensitive electron-multiplying charge coupled device camera. It is able to take pictures at a series of wavelengths in rapid succession at short exposure times, thereby eliminating the image shift introduced by natural eye movements (saccades). In contrast with snapshot systems the images retain full spatial resolution. The system is not suitable for applications where the full spectral resolution is required as it uses discrete wavebands for illumination. This is not a problem in retinal imaging where the use of selected wavelengths is common. The modular nature of the light source allows new wavelengths to be introduced easily and at low cost. The use of wavelength-specific LEDs as a source is preferable to white light illumination and subsequent filtering of the remitted light as it minimizes the total light exposure of the subject. The system is controlled via a graphical user interface that enables flexible control of intensity, duration, and sequencing of sources in synchrony with the camera. Our initial experiments indicate that the system can acquire multispectral image sequences of the human retina at exposure times of 0.05 s in the range of 500-620 nm with mean signal to noise ratio of 17 dB (min 11, std 4.5), making it suitable for quantitative analysis with application to the diagnosis and screening of eye diseases such as diabetic retinopathy and age-related macular degeneration. (c) 2010 American Institute of Physics. [doi:10.1063/1.3478001]
引用
收藏
页数:9
相关论文
共 17 条
[1]   New spectral imaging techniques for blood oximetry in the retina [J].
Alabboud, Ied ;
Muyo, Gonzalo ;
Gorman, Alistair ;
Mordant, David ;
McNaught, Andrew ;
Petres, Clement ;
Petillot, Yvan R. ;
Harvey, Andrew R. .
NOVEL OPTICAL INSTRUMENTATION FOR BIOMEDICAL APPLICATIONS III, 2007, 6631
[2]   Retinal blood oxygen saturation mapping by multispectral imaging and morphological angiography [J].
Arimoto, Hidenobu ;
Furukawa, Hiromitsu .
2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, :1627-1630
[3]   Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation [J].
Beach, JM ;
Schwenzer, KJ ;
Srinivas, S ;
Kim, D ;
Tiedeman, JS .
JOURNAL OF APPLIED PHYSIOLOGY, 1999, 86 (02) :748-758
[4]   Fundus reflectance - historical and present ideas [J].
Berendschot, TTJM ;
DeLint, PJ ;
van Norren, D .
PROGRESS IN RETINAL AND EYE RESEARCH, 2003, 22 (02) :171-200
[5]   Maximum permissible exposures for ocular safety (ANSI 2000), with emphasis on ophthalmic devices [J].
Delori, Francois C. ;
Webb, Robert H. ;
Sliney, David H. .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2007, 24 (05) :1250-1265
[6]   A frequency multiplexed near-infrared topography system for imaging functional activation in the brain [J].
Everdell, NL ;
Gibson, AP ;
Tullis, IDC ;
Vaithianathan, T ;
Hebden, JC ;
Delpy, DT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (09)
[7]  
Greaney MJ, 2002, INVEST OPHTH VIS SCI, V43, P140
[8]   Hyperspectral imaging for the detection of retinal diseases [J].
Harvey, AR ;
Lawlor, A ;
McNaught, AI ;
Williams, JW ;
Fletcher-Holmes, DW .
IMAGING SPECTROMETRY VIII, 2002, 4816 :325-335
[9]   Snapshot hyperspectral imaging in ophthalmology [J].
Johnson, William R. ;
Wilson, Daniel W. ;
Fink, Wolfgang ;
Humayun, Mark ;
Bearman, Greg .
JOURNAL OF BIOMEDICAL OPTICS, 2007, 12 (01)
[10]   A new compact design interferometer based spectral imaging system for bio-medical applications [J].
Lavi, M ;
Milman, U ;
Cabib, D ;
Garini, Y ;
Gil, A ;
Juta, T ;
Adel, M .
THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING V, PROCEEDINGS OF, 1998, 3261 :313-321