An improved method for the detection of changes in brain extracellular glutamate levels

被引:33
作者
Kohno, T
Asai, S
Iribe, Y
Hosoi, I
Shibata, K
Ishikawa, K
机构
[1] Nihon Univ, Sch Med, Dept Pharmacol, Itabashi Ku, Tokyo 173, Japan
[2] Tokyo Gas Co Ltd, Frontier Technol Res Inst, Biomed Res Team, Tsurumi Ku, Yokohama, Kanagawa 230, Japan
关键词
anoxia; dialysis electrode; enzyme electrode; glutamate; microdialysis; striatum;
D O I
10.1016/S0165-0270(98)00041-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We developed a method for in vivo real-time monitoring of the concentration of extracellular glutamate ([Glu]e) in the brain under anoxic conditions. A dialysis electrode (Sycopel Int., UK) was employed as a sensing device to measure the concentration of glutamate by enzyme amperometry, and an electron mediator, ferrocene, was introduced into the electrode together with glutamate oxidase, The ferrocene was covalently conjugated with a high molecular weighs molecule, bovine serum albumin, to avoid outward diffusion through the dialysis membrane. With this set-up, the amperometric response was independent of the pO(2) around the electrode in vitro up to 400 mu M glutamate. Using this method, we investigated the dynamics of [Glu]e in the rat striatum during anoxia. [Glu]e increased rapidly at 102 +/- 5.4 s (n = 6) after the start of nitrogen inhalation. The increase continued for about 30 s, and then [Glu]e decreased. The peak value of Delta[Glu]e was 141 +/- 37 mu M. [Glu]e subsequently underwent another gradual increase, reaching 213 +/- 69 mu M at 15 min after the start of nitrogen inhalation. This distinct biphasic profile was reproducible. We conclude that this method is very useful for monitoring [Glu]e in the brain under low pO(2) conditions. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:199 / 205
页数:7
相关论文
共 14 条
  • [1] THE DIALYSIS ELECTRODE - A NEW METHOD FOR INVIVO MONITORING
    ALBERY, WJ
    BOUTELLE, MG
    GALLEY, PT
    [J]. JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1992, (12) : 900 - 901
  • [2] ALI ZS, 1992, B ELECTROCHEM, V8, P171
  • [3] Real time monitoring of biphasic glutamate release using dialysis electrode in rat acute brain ischemia
    Asai, S
    Iribe, Y
    Kohno, T
    Ishikawa, K
    [J]. NEUROREPORT, 1996, 7 (05) : 1092 - 1096
  • [4] NONVESICULAR RELEASE OF NEUROTRANSMITTER
    ATTWELL, D
    BARBOUR, B
    SZATKOWSKI, M
    [J]. NEURON, 1993, 11 (03) : 401 - 407
  • [5] STABILITY OF WATER-SOLUBLE CARBODIIMIDES IN AQUEOUS-SOLUTION
    GILLES, MA
    HUDSON, AQ
    BORDERS, CL
    [J]. ANALYTICAL BIOCHEMISTRY, 1990, 184 (02) : 244 - 248
  • [6] DIRECT MEASUREMENT OF GLUTAMATE RELEASE IN THE BRAIN USING A DUAL ENZYME-BASED ELECTROCHEMICAL SENSOR
    HU, YB
    MITCHELL, KM
    ALBAHADILY, FN
    MICHAELIS, EK
    WILSON, GS
    [J]. BRAIN RESEARCH, 1994, 659 (1-2) : 117 - 125
  • [7] THE EARLY EVENTS OF OXYGEN AND GLUCOSE DEPRIVATION - SETTING THE SCENE FOR NEURONAL DEATH
    MARTIN, RL
    LLOYD, HGE
    COWAN, AI
    [J]. TRENDS IN NEUROSCIENCES, 1994, 17 (06) : 251 - 257
  • [8] A NEW ENZYMATIC CYCLING TECHNIQUE FOR GLUTAMATE DETERMINATION IN BRAIN MICRODIALYSATES
    MITANI, A
    KUBO, H
    IGA, K
    IMON, H
    KADOYA, F
    KATAOKA, K
    [J]. JOURNAL OF NEUROCHEMISTRY, 1990, 54 (02) : 709 - 711
  • [9] OBRENOVITCH TP, 1990, PHARMACOLOGY OF CEREBRAL ISCHEMIA 1990, P23
  • [10] CEREBRAL VASCULAR AUTO-REGULATION OF BLOOD-FLOW AND TISSUE PO-2 IN DIABETIC RATS
    RUBIN, MJ
    BOHLEN, HG
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1985, 249 (03): : H540 - H546