normal phase TLC;
densitometry;
videodensitometry;
videoscanning;
bezaflbrate;
ciprofibrate;
D O I:
10.1556/JPC.18.2005.3.3
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
New simple TLC methods with densitometry and videoscanning have been developed for the quantitation of bezafibrate in Bezamidin tablets and ciprofibrate in Lipanor capsules. Analysis was performed on HPTLC Dint F-254 plates with hexane-tetrahydrofu ran, 8 + 2, as mobile phase. Detection was performed by densitometry at lambda = 227 nm and videoscanning at A = 254 nm. Calibration plots were constructed in the range 5-30 mu g per spot for both drugs. The calibration data were tested using several regression models and the optimum models were selected (quadratic for videoscanning and nonlinear y = ax(m) + b for densitometry; R-2 was always > 0.995). The active substances were extracted from tablets with methanol. The linearity of the method was tested by spotting different amounts of extracted solution (15-30 mg). The recovery function was always sufficiently linear, with an insignificant intercept and slope very close to unity. Accuracy was tested by quantitating three fortified samples (50, 100, and 150%); this resulted in homogeneous results without significant differences. Recovery measured by use of densitometry was 100.3% (RSD 7.84%) for bezalibrate and 98.01% (RSD 6.12%) for ciprofibrate. Videodensitometry resulted in recovery of 96.16% (RSD 9.8%) and 97.8% (RSD 11.2%), respectively. The F-Snedecor test and t-test for two means showed there was no significant difference between the precision and accuracy of the methods.