Continuous monitoring for cyanide in waste water with a galvanic hydrogen cyanide sensor using a purge system

被引:41
作者
Hachiya, H
Ito, S
Fushinuki, Y
Masadome, T
Asano, Y
Imato, T
机构
[1] Ariake Natl Coll Technol, Omuta, Fukuoka 8368585, Japan
[2] DKK Corp, R&D Div, Musashino, Tokyo 1800001, Japan
[3] Kagoshima Pref Polece HQ, Sci Invest Res Lab, Kagoshima 8900064, Japan
[4] Kyushu Univ, Fac Engn, Higashi Ku, Fukuoka 8120053, Japan
关键词
galvanic cell; chemical sensor; cyanide; continuous monitoring; gas phase; industrial waste water;
D O I
10.1016/S0039-9140(98)00314-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A continuous monitoring system for cyanide with a galvanic hydrogen cyanide sensor and an aeration pump for purging was developed. Hydrogen cyanide evolved from cyanide solution using a purging pump was measured with the hydrogen cyanide sensor. The system showed good performance in terms of stability and selectivity. A linear calibration curve was obtained in the concentrating range from 0 to 15 mg dm(3) of cyanide ion with a slope of -0.24 mu A mg(-1) dm(-3). The lower detection limit was 0.1 mg dm(-3). The 90% response time of the sensor system was within 3.5 min for a 0.5 mg dm(-3) cyanide solution, when the flow rate of the purging air was 1 dm(3) min(-1). The system maintained the initial performance for 6 months in the field test. The developed galvanic sensor system was not subject to interference from sulfide and residual chlorine, compared with a potentiometric sensor system developed previously. The analytical results obtained by the present system were in good agreement with those obtained by the pyridine pyrazolone method. The correlation factor and regression line between both methods were 0.979 and Y = 2.30 x 10(-4) + 1.12X, respectively. This system was successfully applied for a continuous monitoring of cyanide ion in waste water. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:997 / 1004
页数:8
相关论文
共 14 条
[1]  
ASANO Y, 1987, Patent No. 1416967
[2]  
ASANO Y, 1972, CHEM ENG, V17, P90
[3]   EVALUATION OF EXPERIMENTAL CONDITIONS ON THE RESPONSE OF FLUORIDE AND CYANIDE SELECTIVE ELECTRODES IN FLOW-INJECTION POTENTIOMETRY [J].
ELSHOLZ, O ;
FRENZEL, W ;
LIU, CY ;
MOLLER, J .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 338 (02) :159-162
[4]   ESTIMATION OF MICROQUANTITIES OF CYANIDE [J].
EPSTEIN, J .
ANALYTICAL CHEMISTRY, 1947, 19 (04) :272-274
[5]   ELECTRODE INDICATOR TECHNIQUE FOR MEASURING LOW-LEVELS OF CYANIDE [J].
FRANT, MS ;
ROSS, JW ;
RISEMAN, JH .
ANALYTICAL CHEMISTRY, 1972, 44 (13) :2227-&
[6]   ION CHROMATOGRAPHY WITH COULOMETRIC DETECTION FOR THE DETERMINATION OF INORGANIC-IONS [J].
GIRARD, JE .
ANALYTICAL CHEMISTRY, 1979, 51 (07) :836-839
[7]  
*JAP IND STAN, K01021991 JIS, P119
[8]  
*JAP IND STAND, K01021991 JIS, P122
[9]   SELECTIVE DETERMINATION OF GASES BY 2-STAGE MEMBRANE-DIFFERENTIATED FLOW-INJECTION ANALYSIS - DETERMINATION OF TRACE HYDROGEN-CYANIDE IN THE PRESENCE OF LARGE CONCENTRATIONS OF HYDROGEN-SULFIDE [J].
KUBAN, V ;
DASGUPTA, PK .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1106-1112
[10]  
LATIMER WM, 1952, OXIDATION POTENTIALS, P191