Strain analysis of different diameter Morse taper implants under overloading compressive conditions

被引:3
作者
Castro, Carolina Guimaraes [1 ]
Zancope, Karla [1 ]
Verissimo, Crisnicaw [2 ]
Soares, Carlos Jose [2 ]
das Neves, Flavio Domingues [1 ]
机构
[1] Univ Fed Uberlandia, Sch Dent, Dept Occlus Fixed Prostheses & Dent Mat, Uberlandia, MG, Brazil
[2] Univ Fed Uberlandia, Sch Dent, Dept Operat Dent & Dent Mat, Uberlandia, MG, Brazil
关键词
Dental Implants; Dental Implant-Abutment Design; Mechanical Phenomena; MAXIMAL BITE FORCE; FRACTURE-RESISTANCE; DENTAL IMPLANTS; CONNECTION; MECHANICS; TEETH;
D O I
10.1590/1807-3107BOR-2015.vol29.0028
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aim of this study was to evaluate the amount of deformation from compression caused by different diameters of Morse taper implants and the residual deformation after load removal. Thirty Morse taper implants lacking external threads were divided into 3 groups (n = 10) according to their diameter as follows: 3.5 mm, 4.0 mm and 5.0 mm. Two-piece abutments were fixed into the implants, and the samples were subjected to compressive axial loading up to 1500 N of force. During the test, one strain gauge remained fixed to the cervical portion of each implant to measure the strain variation. The strain values were recorded at two different time points: at the maximum load (1500 N) and 60 seconds after load removal. To calculate the strain at the implant/abutment interface, a mathematical formula was applied. Data were analyzed using a one-way Anova and Tukey's test (a = 0.05). The 5.0 mm diameter implant showed a significantly lower strain (650.5 mu S +/- 170.0) than the 4.0 mm group (1170.2 mu S +/- 374.7) and the 3.5 mm group (1388.1 mu S +/- 326.6) (p < 0.001), regardless of the load presence. The strain values decreased by approximately 50% after removal of the load, regardless of the implant diameter. The 5.0 mm implant showed a significantly lower strain at the implant/abutment interface (943.4 mu S +/- 504.5) than the 4.0 mm group (1057.4 mu S +/- 681.3) and the 3.5 mm group (1159.6 mu S +/- 425.9) (p < 0.001). According to the results of this study, the diameter influenced the strain around the internal and external walls of the cervical region of Morse taper implants; all diameters demonstrated clinically acceptable values of strain.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 14 条
[1]   Mechanics of the tapered interference fit in dental implants [J].
Bozkaya, D ;
Müftü, S .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1649-1658
[2]   Fracture resistance and mode of failure of various types of root filled teeth [J].
Castro, C. G. ;
Santana, F. R. ;
Roscoe, M. G. ;
Simamoto, P. C., Jr. ;
Santos-Filho, P. C. F. ;
Soares, C. J. .
INTERNATIONAL ENDODONTIC JOURNAL, 2012, 45 (09) :840-847
[3]  
Cosme DC, 2005, INT J PROSTHODONT, V18, P328
[4]   Strain Gauge Analysis of the Effect of Porcelain Firing Simulation on the Prosthetic Misfit of Implant-Supported Frameworks [J].
De Vasconcellos, Diego Klee ;
Oezcan, Mutlu ;
Maziero Volpato, Claudia Angela ;
Bottino, Marco Antonio ;
Yener, Esra Salihoglu .
IMPLANT DENTISTRY, 2012, 21 (03) :225-229
[5]  
Fernandes AS, 2001, INT J PROSTHODONT, V14, P355
[6]  
Haack J E, 1995, Int J Oral Maxillofac Implants, V10, P529
[7]   Bone Microstrain Values of 1-Piece and 2-Piece Implants Subjected to Mechanical Loading [J].
Harel, Noga ;
Eshkol-Yogev, Inbar ;
Piek, Dana ;
Livne, Shiri ;
Lavi, David ;
Ormianer, Zeev .
IMPLANT DENTISTRY, 2013, 22 (03) :277-281
[8]   Prospective Evaluation of 2,549 Morse Taper Connection Implants: 1- to 6-Year Data [J].
Mangano, Carlo ;
Mangano, Francesco ;
Shibli, Jamil A. ;
Tettamanti, Lucia ;
Figliuzzi, Michele ;
d'Avila, Susana ;
Sammons, Rachel L. ;
Piattelli, Adriano .
JOURNAL OF PERIODONTOLOGY, 2011, 82 (01) :52-61
[9]  
Merz BR, 2000, INT J ORAL MAX IMPL, V15, P519
[10]  