Synchrotron X-ray Irradiation of a Rat's Head Model: Monte Carlo Study of Chromatic Gel Dosimetry

被引:3
作者
Walg, Yarden Peleg [1 ]
Krutman, Yanai [2 ]
Berman, Amir [1 ]
Orion, Itzhak [3 ]
机构
[1] Ben Gurion Univ Negev, Dept Biotechnol Engn, IL-84105 Beer Sheva, Israel
[2] Soroka Univ, Med Ctr, Inst Oncol, Radiotherapy Dept, IL-84100 Beer Sheva, Israel
[3] Ben Gurion Univ Negev, Dept Nucl Engn, IL-84105 Beer Sheva, Israel
来源
APPLIED SCIENCES-BASEL | 2021年 / 11卷 / 16期
关键词
X-ray; gel dosimeters; EGS5; Monte Carlo simulations; percentage depth dose; polydiacetylene; DOSE DISTRIBUTIONS; RADIATION; PHOTON;
D O I
10.3390/app11167389
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Accurate treatment planning in radiotherapy essentially decreases damage to healthy tissue surrounding the tumor. Due to plans to use a direct, highly collimated, narrow beam with high intensity to treat small area tumors, researchers have studied microbeam radiation therapy extensively. Using a synchrotron beam as the radiation source may help to limit damage, but treatment planning using computerized simulations and dosimetry is still necessary to achieve optimal results. For this purpose, PDA-gel dosimeters were developed and their sensitivity around a 150 keV induced synchrotron X-ray radiation beam was examined via Monte Carlo simulations using the EGS5 code system. The microbeam development is now at the animal study stage. In this study, we simulate the irradiation of a rat's brain. The simulation results obtained spectra for two types of PDA-gel dosimeters that were compared with the spectrum obtained in a modelized brain tumor of a rat. Additionally, percentage depth dose curves were calculated for the brain tissue and the two gels. Correction equations for the dosimeters were obtained from the dose-difference plots. For further references, these equations can be used to calculate the actual dose in a brain tumor in a rat. The Monte Carlo simulations demonstrate that PDA-gel dosimeters can be used for treatment planning using synchrotron irradiations.
引用
收藏
页数:9
相关论文
共 19 条
[1]   Review of gel dosimetry: a personal reflection [J].
Baldock, C. .
MICRO-MINI & NANO-DOSIMETRY & INNOVATIVE TECHNOLOGIES IN RADIATION THERAPY (MMND&ITRO2016), 2017, 777
[2]   Polymer gel dosimetry [J].
Baldock, C. ;
De Deene, Y. ;
Doran, S. ;
Ibbott, G. ;
Jirasek, A. ;
Lepage, M. ;
McAuley, K. B. ;
Oldham, M. ;
Schreiner, L. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :R1-R63
[3]  
Briesmeister J.F., 1993, MCNP A GEN MONTE CAR
[4]   Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning [J].
Chetty, Indrin J. ;
Curran, Bruce ;
Cygler, Joanna E. ;
DeMarco, John J. ;
Ezzell, Gary ;
Faddegon, Bruce A. ;
Kawrakow, Iwan ;
Keall, Paul J. ;
Liu, Helen ;
Ma, C. -M. Charlie ;
Rogers, D. W. O. ;
Seuntjens, Jan ;
Sheikh-Bagheri, Daryoush ;
Siebers, Jeffrey V. .
MEDICAL PHYSICS, 2007, 34 (12) :4818-4853
[5]   SYNCHROTRON RADIATION AS AN ABSOLUTE STANDARD SOURCE [J].
EINFELD, D ;
STUCK, D .
NUCLEAR INSTRUMENTS & METHODS, 1980, 172 (1-2) :101-106
[6]   Age-dependent changes in material properties of the brain and braincase of the rat [J].
Gefen, A ;
Gefen, N ;
Zhu, QL ;
Raghupathi, R ;
Margulies, SS .
JOURNAL OF NEUROTRAUMA, 2003, 20 (11) :1163-1177
[7]  
Hirayama H, 2005, EGS5 CODE SYSTEM KEK
[8]  
Jabbari Keyvan, 2011, J Med Signals Sens, V1, P73
[9]  
Khan FaizM., 2016, Khans Treatment Planning in Radiation Oncology
[10]   Structural Transitions in Polydiacetylene Langmuir Films [J].
Lifshitz, Yevgeniy ;
Golan, Yuval ;
Konovalov, Oleg ;
Berman, Amir .
LANGMUIR, 2009, 25 (08) :4469-4477