Laboratory and Field Evaluation of the Partec CyFlow MiniPOC for Absolute and Relative CD4 T-Cell Enumeration

被引:10
作者
Wade, Djibril [1 ,2 ,3 ]
Diaw, Papa Alassane [1 ]
Daneau, Geraldine [2 ]
Diallo, Abdoul Aziz [1 ]
Mboup, Souleymane [1 ]
Dieye, Tandakha Ndiaye [1 ]
Kestens, Luc [2 ,3 ]
机构
[1] Le Dantec Hosp, Immunol Unit, Lab Bacteriol Virol, Dakar, Senegal
[2] Inst Trop Med, Immunol Lab, Dept Biomed Sci, B-2000 Antwerp, Belgium
[3] Univ Antwerp, Dept Biomed Sci, B-2020 Antwerp, Belgium
关键词
HIV-INFECTED INDIVIDUALS; PERFORMANCE EVALUATION; FLOW CYTOMETER; SINGLE-PLATFORM; POINT; PIMA; SETTINGS; ANALYZER; COUNTS; SYSTEM;
D O I
10.1371/journal.pone.0116663
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background A new CD4 point-of-care instrument, the CyFlow miniPOC, which provides absolute and percentage CD4 T-cells, used for screening and monitoring of HIV-infected patients in resource-limited settings, was introduced recently. We assessed the performance of this novel instrument in a reference laboratory and in a field setting in Senegal. Methodology A total of 321 blood samples were obtained from 297 adults and 24 children, all HIV-patients attending university hospitals in Dakar, or health centers in Ziguinchor. Samples were analyzed in parallel on CyFlow miniPOC, FACSCount CD4 and FACSCalibur to assess CyFlow miniPOC precision and accuracy. Results At the reference lab, CyFlow miniPOC, compared to FACSCalibur, showed an absolute mean bias of -12.6 cells/mm(3) and a corresponding relative mean bias of -2.3% for absolute CD4 counts. For CD4 percentages, the absolute mean bias was -0.1%. Compared to FACSCount CD4, the absolute and relative mean biases were -31.2 cells/mm3 and -4.7%, respectively, for CD4 counts, whereas the absolute mean bias for CD4 percentages was 1.3%. The CyFlow miniPOC was able to classify HIV-patients eligible for ART with a sensitivity of >= 95% at the different ART-initiation thresholds (200, 350 and 500 CD4 cells/mm(3)). In the field lab, the room temperature ranged from 30 to 35 degrees C during the working hours. At those temperatures, the CyFlow miniPOC, compared to FACSCount CD4, had an absolute and relative mean bias of 7.6 cells/mm(3) and 2.8%, respectively, for absolute CD4 counts, and an absolute mean bias of 0.4% for CD4 percentages. The CyFlow miniPOC showed sensitivity equal or greater than 94%. Conclusion The CyFlow miniPOC showed high agreement with FACSCalibur and FACSCount CD4. The CyFlow miniPOC provides both reliable absolute CD4 counts and CD4 percentages even under the field conditions, and is suitable for monitoring HIV-infected patients in resource-limited settings.
引用
收藏
页数:14
相关论文
共 48 条
[1]  
[Anonymous], HIV AIDS DIAGN TECHN
[2]   A reliable and inexpensive EasyCD4 assay for monitoring HIV-infected individuals in resource-limited settings [J].
Balakrishnan, Pachamuthu ;
Solomon, Suniti ;
Mohanakrishnan, Janardhanan ;
Cecelia, Anitha J. ;
Kumarasamy, Nagalingeswaran ;
Murugavel, Kailapuri G. ;
Venkatakrishnan, Bhavani ;
Solomon, Sunil S. ;
Crowe, Suzanne M. ;
Ganesh, Aylur K. ;
Thyagarajan, Sadras P. ;
Flanigan, Timothy ;
Mayer, Kenneth H. .
JAIDS-JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES, 2006, 43 (01) :23-26
[3]   Performance of the PointCare NOW System for CD4 Counting in HIV Patients Based on Five Independent Evaluations [J].
Bergeron, Michele ;
Daneau, Geraldine ;
Ding, Tao ;
Sitoe, Nadia E. ;
Westerman, Larry E. ;
Stokx, Jocelijn ;
Jani, Ilesh V. ;
Coetzee, Lindi M. ;
Scott, Lesley ;
De Weggheleire, Anja ;
Boel, Luc ;
Stevens, Wendy S. ;
Glencross, Deborah K. ;
Peter, Trevor F. .
PLOS ONE, 2012, 7 (08)
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   REVISED SCHEME FOR EVALUATION OF AUTOMATIC INSTRUMENTS FOR USE IN CLINICAL CHEMISTRY [J].
BROUGHTON, PM ;
GOWENLOCK, AH ;
MCCORMACK, JJ ;
NEILL, DW .
ANNALS OF CLINICAL BIOCHEMISTRY, 1974, 11 (NOV) :207-218
[6]  
Cheng XH, 2007, JAIDS-J ACQ IMM DEF, V45, P257
[7]   Multisite Evaluation of a Point-of-Care Instrument for CD4+ T-Cell Enumeration Using Venous and Finger-Prick Blood: The PIMA CD4 [J].
Diaw, Papa Alassane ;
Daneau, Geraldine ;
Coly, Abdoul Aziz ;
Ndiaye, Birahim Pierre ;
Wade, Djibril ;
Camara, Makhtar ;
Mboup, Souleymane ;
Kestens, Luc ;
Dieye, Tandakha Ndiaye .
JAIDS-JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES, 2011, 58 (04) :E103-E111
[8]  
Dunn DT, 2006, AIDS, V20, P1289, DOI 10.1097/01.aids.0000232237.20792.68
[9]   The Partec CyFlow Counter® could provide an option for CD4+T-cell monitoring in the context of scaling-up antiretroviral treatment at the district level in Malawi [J].
Fryland, M. ;
Chaillet, P. ;
Zachariah, R. ;
Barnaba, A. ;
Bonte, L. ;
Andereassen, R. ;
Charrondiere, S. ;
Teck, R. ;
Didakus, O. .
TRANSACTIONS OF THE ROYAL SOCIETY OF TROPICAL MEDICINE AND HYGIENE, 2006, 100 (10) :980-985
[10]   Performance evaluation of the Pima™ point-of-care CD4 analyser using capillary blood sampling in field tests in South Africa [J].
Glencross, Deborah K. ;
Coetzee, Lindi M. ;
Faal, Mamsallah ;
Masango, Martin ;
Stevens, Wendy S. ;
Venter, W. D. Francois ;
Osih, Regina .
JOURNAL OF THE INTERNATIONAL AIDS SOCIETY, 2012, 15