Implementation of total laboratory automation at a tertiary care hospital in Saudi Arabia: effect on turnaround time and cost efficiency

被引:14
作者
Ellison, Tracy Louise [1 ]
Alharbi, Maha [1 ]
Alkaf, Morad [1 ]
Elimam, Shamad [1 ]
Alfaries, Mariam [1 ]
Al Nounou, Randa [1 ]
Nasr, Rasheed [1 ]
Owaidah, Tarek [1 ]
机构
[1] King Faisal Specialist Hosp & Res Ctr, Dept Pathol & Lab Med, POB 3354, Riyadh 11211, Saudi Arabia
关键词
IMPACT;
D O I
10.5144/0256-4947.2018.352
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: Total laboratory automation (TLA) is a relatively new way of improving the management of high volume clinical laboratories. TLA may reduce staff, reduce operating costs, decrease testing time and provide enhanced process control. OBJECTIVES: Establish a cost efficient TLA that is less labor intensive, improves productivity and reduces turnaround time (TAT). DESIGN: Implementation of TLA for random glucose and troponin-T as sentinel tests to compare change in TAT. SETTING: Tertiary hospital with high volume of laboratory tests. METHODS: Routine patient samples for random glucose and troponin-T were used to capture TAT. Information on staff grades and schedules before and after implementing the TLA, and cost of contracts to deliver the service were collected. MAIN OUTCOME MEASURES: TAT, cost efficiency, and reduction in labor. RESULTS: The consolidation of contracts resulted in a reduction of 28.8 million SAR in direct costs. Staffing cost was reduced by 1.14 million SAR with less senior staff required; there were reductions in staff at both senior and junior level. The overall TAT for all tests was reduced by 32% in 2016 (after TLA implementation) compared to 2012 (before TLA implementation). The median TAT for random glucose tests was reduced by 21% (to 55.7 minutes in 2016 from 70.1 minutes in 2012). Evidence of test optimization by exploring the impact of stat tests, auto-dilutions and reruns on the overall TAT of the TLA is shown by comparing troponin T TATs after reclassifying stat tests (in 2016) to routine (in 2017). At the 75th percentile, there was a 27% reduction in TAT when comparing August 2016 to March 2017 with a 19% reduction in median TAT. CONCLUSION: By moving from stat to routine assays, the TAT was reduced, which is counter-intuitive. The use of stat assays slowed down the performance of the TLA. A careful review of the mix of assays should be conducted to maximize performance and to ensure that the system delivers what is required. LIMITATIONS: Room for improvement by systematically analyzing and reviewing the impact of making minor changes that could have significant impact on TAT.
引用
收藏
页码:352 / 357
页数:6
相关论文
共 7 条
[1]   The impact of automating laboratory request forms on the quality of healthcare services [J].
Al Dogether, Majed ;
Al Muallem, Yahya ;
Househ, Mowafa ;
Saddik, Basema ;
Khalifa, Mohamed .
JOURNAL OF INFECTION AND PUBLIC HEALTH, 2016, 9 (06) :749-756
[2]  
Archetti C, 2017, J PUBLIC HEALTH RES, V6, P31, DOI 10.4081/jphr.2017.881
[3]   Total laboratory automation: Do stat tests still matter? [J].
Dolci, Alberto ;
Giavarina, Davide ;
Pasqualetti, Sara ;
Szoke, Dominika ;
Panteghini, Mauro .
CLINICAL BIOCHEMISTRY, 2017, 50 (10-11) :605-611
[4]   Nonanalytic Laboratory Automation: A Quarter Century of Progress [J].
Hawker, Charles D. .
CLINICAL CHEMISTRY, 2017, 63 (06) :1074-1082
[5]   Designing and evaluating autoverification rules for thyroid function profiles and sex hormone tests [J].
Li, Jiancheng ;
Cheng, Bizhen ;
Ouyang, Huizhen ;
Xiao, Tongtong ;
Hu, Jing ;
Cai, Yingmu .
ANNALS OF CLINICAL BIOCHEMISTRY, 2018, 55 (02) :254-263
[6]  
Lou AH, 2017, CLIN BIOCHEM, DOI [10.1016/j.clinbio-chem.2017.04.019, DOI 10.1016/J.CLINBIO-CHEM.2017.04.019]
[7]   Evaluation of the impact of a total automation system in a large core laboratory on turnaround time [J].
Lou, Amy H. ;
Elnenaei, Manal O. ;
Sadek, Irene ;
Thompson, Shauna ;
Crocker, Bryan D. ;
Nassar, Bassam .