A biodegradable slotted tube stent based on Poly(L-lactide) and poly(4-hydroxybutyrate) for rapid balloon-expansion

被引:82
作者
Grabow, Niels
Buenger, Carsten M.
Schultze, Christine
Schmohl, Kathleen
Martin, David P.
Williams, Simon F.
Sternberg, Katrin
Schmitz, Klaus-Peter
机构
[1] Univ Rostock, Inst Biomed Engn, D-18057 Rostock, Germany
[2] Univ Rostock, Dept Surg, D-18057 Rostock, Germany
[3] Tepha Inc, Lexington, MA 02421 USA
关键词
PLLA; P4HB; recoil; collapse; laser machining; mechanical properties; degradation;
D O I
10.1007/s10439-007-9376-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Safe vascular stent application requires rapid expansion of the stent to minimize the risk of procedural ischemia. While high expansion speeds can be achieved with metallic stents, they are not necessarily feasible with biodegradable polymeric stents due to the viscoelastic material behavior. This study reports on a novel biodegradable polymer blend material based on poly(L-lactide) (PLLA) and poly(4-hydroxybutyrate) (P4HB), and describes the mechanical properties and in vitro degradation behavior of a balloon-expandable slotted tube stent concept. The stent prototypes with nominal dimensions of 6.0 x 25 mm were manufactured by laser machining of solution cast PLLA/P4HB tubes (I.D. = 2.8 mm, d = 300 mu m). The stents were expanded within 1 min by balloon inflation to 8 bar, after 5 min preconditioning in 37 degrees C water. Recoil and collapse pressure were 4.2% and 1.1 bar, respectively. During in vitro degradation collapse pressure initially increased to a maximum at 4 w and then decreased thereafter. After 48 w, molecular weight was decreased by 82%. In summary, the PLLA/P4HB slotted tube stents allowed for rapid balloon-expansion and exhibited adequate mechanical scaffolding properties suitable for a broad range of vascular and non-vascular applications.
引用
收藏
页码:2031 / 2038
页数:8
相关论文
共 29 条
  • [11] Peripheral artery stent visualization and in-stent stenosis analysis in 16-row computed tomography: an in-vitro evaluation
    Herzog, C
    Grebe, C
    Mahnken, A
    Balzer, JO
    Mack, MG
    Zangos, S
    Ackermann, H
    Schaller, S
    Seifert, T
    Ohnesorge, B
    Vogl, TJ
    [J]. EUROPEAN RADIOLOGY, 2005, 15 (11) : 2276 - 2283
  • [12] Cause of death with bare metal and sirolimus-eluting stents
    Holmes, David R., Jr.
    Moses, Jeffrey W.
    Schofer, Joachim
    Morice, Marie-Claude
    Schampaert, Erick
    Leon, Martin B.
    [J]. EUROPEAN HEART JOURNAL, 2006, 27 (23) : 2815 - 2822
  • [13] A Bioabsorbable self-expandable, self-reinforced poly-L-lactic acid urethral stent for recurrent urethral strictures:: Long-term results
    Isotalo, T
    Talja, M
    Välimaa, T
    Törmälä, P
    Tammela, TLJ
    [J]. JOURNAL OF ENDOUROLOGY, 2002, 16 (10) : 759 - 762
  • [14] Biocompatibility properties of a new braided biodegradable urethral stent:: a comparison with a biodegradable spiral and a braided metallic stent in the rabbit urethra
    Isotalo, TM
    Nuutine, JP
    Vaajanen, A
    Martikainen, PM
    Laurila, M
    Törmälä, P
    Talja, M
    Tammela, TL
    [J]. BJU INTERNATIONAL, 2006, 97 (04) : 856 - 859
  • [15] Pathology of drug-eluting stents in humans - Delayed healing and late thrombotic risk
    Joner, Michael
    Finn, Aloke V.
    Farb, Andrew
    Mont, Erik K.
    Kolodgie, Frank D.
    Ladich, Elena
    Kutys, Robert
    Skorija, Kristi
    Gold, Herman K.
    Virmani, Renu
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (01) : 193 - 202
  • [16] Self-expandable chitosan stent: design and preparation
    Lauto, A
    Ohebshalom, M
    Esposito, M
    Mingin, J
    Li, PS
    Felsen, D
    Goldstein, M
    Poppas, DP
    [J]. BIOMATERIALS, 2001, 22 (13) : 1869 - 1874
  • [17] Leadership skills and the group performance: Situational demands, behavioral requirements, and planning
    Marta, S
    Leritz, LE
    Mumford, MD
    [J]. LEADERSHIP QUARTERLY, 2005, 16 (01) : 97 - 120
  • [18] Poly(3-hydroxybutyrate-co-4-hydroxybutyrate) produced by Comamonas acidovorans
    Mitomo, H
    Hsieh, WC
    Nishiwaki, K
    Kasuya, K
    Doi, Y
    [J]. POLYMER, 2001, 42 (08) : 3455 - 3461
  • [19] First-in-human implantation of a fully bioabsorbable drug-eluting stent: The BVS poly-L-lactic acid everolimus-eluting coronary stent
    Ormiston, John A.
    Webster, Mark W. I.
    Armstrong, Guy
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (01) : 128 - 131
  • [20] Characteristic mechanical properties of balloon-expandable peripheral stent systems
    Schmidt, W
    Andresen, R
    Behrens, P
    Schmitz, KP
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (11): : 1430 - 1437