A biodegradable slotted tube stent based on Poly(L-lactide) and poly(4-hydroxybutyrate) for rapid balloon-expansion

被引:82
作者
Grabow, Niels
Buenger, Carsten M.
Schultze, Christine
Schmohl, Kathleen
Martin, David P.
Williams, Simon F.
Sternberg, Katrin
Schmitz, Klaus-Peter
机构
[1] Univ Rostock, Inst Biomed Engn, D-18057 Rostock, Germany
[2] Univ Rostock, Dept Surg, D-18057 Rostock, Germany
[3] Tepha Inc, Lexington, MA 02421 USA
关键词
PLLA; P4HB; recoil; collapse; laser machining; mechanical properties; degradation;
D O I
10.1007/s10439-007-9376-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Safe vascular stent application requires rapid expansion of the stent to minimize the risk of procedural ischemia. While high expansion speeds can be achieved with metallic stents, they are not necessarily feasible with biodegradable polymeric stents due to the viscoelastic material behavior. This study reports on a novel biodegradable polymer blend material based on poly(L-lactide) (PLLA) and poly(4-hydroxybutyrate) (P4HB), and describes the mechanical properties and in vitro degradation behavior of a balloon-expandable slotted tube stent concept. The stent prototypes with nominal dimensions of 6.0 x 25 mm were manufactured by laser machining of solution cast PLLA/P4HB tubes (I.D. = 2.8 mm, d = 300 mu m). The stents were expanded within 1 min by balloon inflation to 8 bar, after 5 min preconditioning in 37 degrees C water. Recoil and collapse pressure were 4.2% and 1.1 bar, respectively. During in vitro degradation collapse pressure initially increased to a maximum at 4 w and then decreased thereafter. After 48 w, molecular weight was decreased by 82%. In summary, the PLLA/P4HB slotted tube stents allowed for rapid balloon-expansion and exhibited adequate mechanical scaffolding properties suitable for a broad range of vascular and non-vascular applications.
引用
收藏
页码:2031 / 2038
页数:8
相关论文
共 29 条
  • [1] DISTAL CORONARY-ARTERY PERFUSION DURING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY
    ANDERSON, HV
    LEIMGRUBER, PP
    ROUBIN, GS
    NELSON, DL
    GRUENTZIG, AR
    [J]. AMERICAN HEART JOURNAL, 1985, 110 (04) : 720 - 726
  • [2] BIAMINO G, 2006, INT S END THER
  • [3] Iliac anastomotic stenting with a biodegradable poly-L-Lactide stent:: A preliminary study after 1 and 6 weeks
    Buenger, Carsten Michael
    Grabow, Niels
    Sternberg, Katrin
    Ketner, Lisa
    Kroeger, Christina
    Lorenzen, Bjoern
    Hauenstein, Karlheinz
    Schmitz, Klaus-Peter
    Kreutzer, Hans Juergen
    Lootz, Daniel
    Ince, Hueseyin
    Nienaber, Christoph A.
    Klar, Ernst
    Schareck, Wolfgang
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2006, 13 (04) : 539 - 548
  • [4] BUNGER C, 2007, IN PRESS J ENDOVASC
  • [5] INTRACORONARY STENTING WITHOUT ANTICOAGULATION ACCOMPLISHED WITH INTRAVASCULAR ULTRASOUND GUIDANCE
    COLOMBO, A
    HALL, P
    NAKAMURA, S
    ALMAGOR, Y
    MAIELLO, L
    MARTINI, G
    GAGLIONE, A
    GOLDBERG, SL
    TOBIS, JM
    [J]. CIRCULATION, 1995, 91 (06) : 1676 - 1688
  • [6] The road to bioabsorbable stents: Reaching clinical reality?
    Erne, P
    Schier, M
    Resink, TJ
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 29 (01) : 11 - 16
  • [7] INVESTIGATION OF STRUCTURE OF SOLUTION GROWN CRYSTALS OF LACTIDE COPOLYMERS BY MEANS OF CHEMICAL-REACTIONS
    FISCHER, EW
    STERZEL, HJ
    WEGNER, G
    [J]. KOLLOID-ZEITSCHRIFT AND ZEITSCHRIFT FUR POLYMERE, 1973, 251 (11): : 980 - 990
  • [8] Mechanical properties of laser cut poly(L-lactide) micro-specimens: Implications for stent design, manufacture, and sterilization
    Grabow, N
    Schlun, M
    Sternberg, K
    Hakansson, N
    Kramer, S
    Schmitz, KP
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (01): : 25 - 31
  • [9] Grabow N, 2007, ASME J MED DEVICES, V1, P84
  • [10] GRABOW N, 2006, BIOMATERIALIEN MUNCH, V7, P154