CALIBRATION OF PHOTOEMISSION PYROMETER WITHOUT USING STANDARD TEMPERATURE RADIATORS

被引:2
|
作者
Viazava, K. A. [1 ]
Kasparov, K. N. [1 ]
Miatselskaya, N. S. [2 ]
Penyazkov, O. G. [1 ]
机构
[1] Natl Acad Sci Belarus, Lykov Inst Heat & Mass Transfer, Minsk, BELARUS
[2] Natl Acad Sci Belarus, Stepanov Phys Inst, Minsk, BELARUS
关键词
photoemission temperature measurements; calibration; photoemission of first dynode of the photomultiplier;
D O I
10.1007/s11018-017-1238-8
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The possibility of calibrating a photoemission pyrometer without using a standard temperaturel radiator is examined. The validity of such calibration is shown, since the calculated calibration curve agrees with that obtained with the use of a ribbon filament lamp.
引用
收藏
页码:578 / 582
页数:5
相关论文
共 50 条
  • [1] Calibration of Photoemission Pyrometer without Using Standard Temperature Radiators
    K. A. Viazava
    K. N. Kasparov
    N. S. Miatselskaya
    O. G. Penyazkov
    Measurement Techniques, 2017, 60 : 578 - 582
  • [2] Hidden Uncertainties of Temperature Measurements in the Calibration of a Pyrometer
    Khodunkov, V. P.
    MEASUREMENT TECHNIQUES, 2019, 61 (11) : 1126 - 1133
  • [3] Hidden Uncertainties of Temperature Measurements in the Calibration of a Pyrometer
    V. P. Khodunkov
    Measurement Techniques, 2019, 61 : 1126 - 1133
  • [4] A pyrometer for calibration traceable to a future thermodynamic temperature scale
    Lowe, D.
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2013, 24 (01)
  • [5] Research on the Non-Source Temperature Calibration of Multispectral Pyrometer
    Sun Xiao-gang
    Sun Kun
    Dai Jing-min
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32 (01) : 273 - 277
  • [6] Investigation of the temperature dependence of the effective wavelength of a standard photoelectric pyrometer
    R. P. Sergienko
    I. S. Romodanov
    Measurement Techniques, 2007, 50 : 991 - 995
  • [7] Investigation of the temperature dependence of the effective wavelength of a standard photoelectric pyrometer
    Sergienko, R. P.
    Romodanov, I. S.
    MEASUREMENT TECHNIQUES, 2007, 50 (09) : 991 - 995
  • [8] ACCURATE DETERMINATION OF THERMODYNAMIC TEMPERATURE BY TUNGSTEN STRIP LAMPS USING INFRARED STANDARD RADIATION PYROMETER
    SINGH, YP
    ZAIDI, ZH
    INDIAN JOURNAL OF PURE & APPLIED PHYSICS, 1994, 32 (03) : 238 - 242
  • [9] Comparative calibration without a gold standard
    Lu, Y
    Ye, KY
    Mathur, AK
    Hui, S
    Fuerst, TP
    Genant, HK
    STATISTICS IN MEDICINE, 1997, 16 (16) : 1889 - 1905
  • [10] Absolute calibration of the OMEGA streaked optical pyrometer for temperature measurements of compressed materials
    Gregor, M. C.
    Boni, R.
    Sorce, A.
    Kendrick, J.
    Mccoy, C. A.
    Polsin, D. N.
    Boehly, T. R.
    Celliers, P. M.
    Collins, G. W.
    Fratanduono, D. E.
    Eggert, J. H.
    Millot, M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (11):