Dew-point measurements at high water vapour pressure

被引:4
作者
Lomperski, S
Dreier, J
机构
[1] Paul Scherrer Institut
关键词
D O I
10.1088/0957-0233/7/5/003
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A dew-point meter capable of measuring humidity at high vapour pressure and high temperature has been constructed and tested. Humidity measurements in pure steam were made over the temperature range 100-150 degrees C and a vapour pressure range of 1-4 bar. The dew-point meter performance was assessed by comparing measurements with a pressure transmitter and agreement between the two was within +/-2.5% relative humidity. Humidity measurements in steam-air mixtures were also made and the dew-point meter readings were compared to those of a zirconia oxygen sensor. for these tests the dew-point meter readings were generally within +/-2.5% relative humidity of the oxygen sensor measurements.
引用
收藏
页码:742 / 745
页数:4
相关论文
共 7 条
[1]  
DADACHANJI F, 1992, MEAS CONTROL, V25, P48
[2]  
DOEBELIN EO, 1990, MEASUREMENT SYSTEMS
[3]  
GOLDSTEIN S, 1990, DRYING TECHNOL, V9, P833
[4]   HUMIDITY SENSORS [J].
KULWICKI, BM .
JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 1991, 74 (04) :697-708
[5]  
LOMPERSKI S, 1994, UNPUB
[6]  
Moseley P.T., 1991, TECHNIQUES MECH GAS, VVolume 234
[7]  
TANAKA H, 1988, P 1 KSME JSME THERM, V1, P462