Variability of lesion detectability and standardized uptake value according to the acquisition procedure and reconstruction among five PET scanners

被引:20
作者
Takahashi, Yasuyuki [1 ]
Oriuchi, Noboru [2 ]
Otake, Hidenori [3 ]
Endo, Keigo [2 ]
Murase, Kenya [4 ]
机构
[1] Gunma Prefectural Coll Hlth Sci, Dept Nucl Med Technol, Maebashi, Gunma 3710052, Japan
[2] Gunma Univ, Grad Sch Med, Dept Diagnost Radiol & Nucl Med, Maebashi, Gunma 371, Japan
[3] Gunma Univ Hosp, Dept Diagnost Radiol, Maebashi, Gunma, Japan
[4] Osaka Univ, Grad Sch Med, Dept Med Engn, Osaka, Japan
关键词
PET; CT; F-18-FDG; SUV;
D O I
10.1007/s12149-008-0152-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective The objective of this study was to assess differences in the semiquantitative values of F-18-fluorodeoxyglucose (F-18-FDG) uptake among different positron emission tomographic (PET) systems. Methods A phantom study was performed to compare standardized uptake value (SUV) in five PET scanners including a dedicated PET scanner and four PET/computed tomography (CT) scanners. Radioactivity simulating the SUV of 2.5 was filled in the hot spheres (8 mm, 11 mm, 14 mm, 18 mm, 22 mm, and 27 mm) that were set in the cylindrical phantom with the background SUV of 1.0. Data acquisition and reconstruction were performed according to routine and standardized conditions. The standardized condition was as follows: CT acquisition (120 kVp, 50 mA) and PET acquisition (2-min acquisition with a slice thickness of 2 mm); reconstruction was performed by ordered subsets expectation maximization + Fourier rebinning. Detectability of hot spheres and SUV was compared between routine condition and standardized condition with five PET scanners. Results On routine condition, two cameras could detect a 14-mm sphere clearly. On the other hand, the visualization of hot spheres by the standardized condition was remarkably variable. Semiquantitative evaluation revealed that a maximum of 45.7% error was recognized with the 27-mm sphere by the routine condition, although the standardized condition could reduce the error to 22.6%. Conclusions Detectability depends not only on the PET machine but also on the imaging protocol. The results indicate that SUV is variable with PET machines under routine conditions of data acquisition and reconstruction. Standardization of the reconditions can reduce variability and maximum difference in the SUV by half.
引用
收藏
页码:543 / 548
页数:6
相关论文
共 22 条
[1]   Optimization of a fully 3D single scatter simulation algorithm for 3D PET [J].
Accorsi, R ;
Adam, LE ;
Werner, ME ;
Karp, JS .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12) :2577-2598
[2]  
Beaulieu S, 2003, J NUCL MED, V44, P1044
[3]  
Boellaard R, 2004, J NUCL MED, V45, P1519
[4]   A row-action alternative to the EM algorithm for maximizing likelihoods in emission tomography [J].
Browne, J ;
DePierro, AR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (05) :687-699
[5]   PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients [J].
Burger, C ;
Goerres, G ;
Schoenes, S ;
Buck, A ;
Lonn, AHR ;
von Schulthess, GK .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (07) :922-927
[6]  
Calvo R, 2000, J NUCL MED, V41, P439
[7]   3D PET USING A CONVENTIONAL MULTISLICE TOMOGRAPH WITHOUT SEPTA [J].
CHERRY, SR ;
DAHLBOM, M ;
HOFFMAN, EJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (04) :655-668
[8]   Fast reconstruction of 3D PET data with accurate statistical modeling [J].
Comtat, C ;
Kinahan, PE ;
Defrise, M ;
Michel, C ;
Townsend, DW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :1083-1089
[9]   Exact and approximate rebinning algorithms for 3-D PET data [J].
Defrise, M ;
Kinahan, PE ;
Townsend, DW ;
Michel, C ;
Sibomana, M ;
Newport, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :145-158
[10]  
Fujiwara T, 1997, Ann Nucl Med, V11, P307