Reconstruction of brachial artery pressure from noninvasive finger pressure measurements

被引:145
作者
Bos, WJW
vanGoudoever, J
vanMontfrans, GA
vandenMeiracker, AH
Wesseling, KH
机构
[1] UNIV ROTTERDAM HOSP, DEPT INTERNAL MED 1, ROTTERDAM, NETHERLANDS
[2] TNO BIOMED INSTRUMENTAT, AMSTERDAM, NETHERLANDS
关键词
pressure; arteries; waves; blood pressure;
D O I
10.1161/01.CIR.94.8.1870
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Pulse wave distortions, mainly caused by reflections, and pressure gradients, caused by flow in the resistive vascular tree, may cause differences between finger and brachial artery pressures. These differences may limit the use of finger pressure measurements. We investigated whether brachial artery pressure waves could be reconstructed from finger pressure measurements by correcting for the pressure gradient in addition to correction for pulse wave distortion with a previously described filter. Methods and Results Finger artery pressure (with Finapres), intra-arterial brachial artery pressure (BAP), Riva-Rocci/Korotkoff (RRK), oscillometric, and return-to-flow (RTF) measurements were simultaneously performed in 57 healthy elderly subjects and patients with vascular disease and/or hypertension. A generalized waveform filter was used to correct for pulse wave distortions. Correction equations for the pressure gradient, based on finger pressure, RRK, RTF, or oscillometric measurements, were obtained in 28 randomly selected subjects and tested in 29. Before reconstruction, Finapres underestimated mean and diastolic BAP (finger pressure minus BAP: systolic, -3.2+/-16.9 mm Hg; mean, -13.0+/-10.5 mm Hg; diastolic, -8.4+/-9.0 mm Hg [mean+/-SD]). After filtering, reconstructed BAP waves were similar to actual BAP in shape but not in pressure level. Optimal correction for the pressure gradient with an equation based on RTF measurements reduced the pressure differences to meet American Association for the Advancement of Medical Instrumentation criteria (reconstructed finger pressure minus BAP: systolic, 3.7+/-7.0 mm Hg; mean, 0.7+/-4.6 mm Hg; and diastolic, 1.0+/-4.9 mm Hg). Conclusions BAP waves can be reconstructed from noninvasive finger pressure registrations when finger pressure waves are corrected for pulse wave distortion and individual pressure gradients.
引用
收藏
页码:1870 / 1875
页数:6
相关论文
共 23 条
  • [1] THE RELIABILITY OF NONINVASIVE CONTINUOUS FINGER BLOOD-PRESSURE MEASUREMENT IN PATIENTS WITH BOTH HYPERTENSION AND VASCULAR-DISEASE
    BOS, WJW
    IMHOLZ, BPM
    VANGOUDOEVER, J
    WESSELING, KH
    VANMONTFRANS, GA
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 1992, 5 (08) : 529 - 535
  • [2] PSEUDOHYPERTENSION AND THE MEASUREMENT OF BLOOD-PRESSURE
    BOS, WJW
    VANGOUDOEVER, J
    WESSELING, KH
    RONGEN, GAPJM
    HOEDEMAKER, G
    LENDERS, JWM
    VANMONTFRANS, GA
    [J]. HYPERTENSION, 1992, 20 (01) : 26 - 31
  • [3] EFFECT OF REGIONAL AND SYSTEMIC CHANGES IN VASOMOTOR TONE ON FINGER PRESSURE AMPLIFICATION
    BOS, WJW
    VANDENMEIRACKER, AH
    WESSELING, KH
    SCHALEKAMP, MADH
    [J]. HYPERTENSION, 1995, 26 (02) : 315 - 320
  • [4] COMPARATIVE ACCURACIES OF A FINGER BLOOD-PRESSURE MONITOR AND AN OSCILLOMETRIC BLOOD-PRESSURE MONITOR
    EPSTEIN, RH
    HUFFNAGLE, S
    BARTKOWSKI, RR
    [J]. JOURNAL OF CLINICAL MONITORING, 1991, 7 (02): : 161 - 167
  • [5] AORTOFEMORAL TRANSFER-FUNCTION - A METHOD TO DETERMINE THE INSTANTANEOUS AORTIC-VALVE GRADIENT IN AORTIC-VALVE STENOSIS
    FITCHETT, DH
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (07) : 1909 - 1914
  • [6] DIRECT BLOOD-PRESSURE MEASUREMENT - DYNAMIC-RESPONSE REQUIREMENTS
    GARDNER, RM
    [J]. ANESTHESIOLOGY, 1981, 54 (03) : 227 - 236
  • [7] GIZDULICH P, 1990, P 12 INT C IEEE ENG, V12, P1046
  • [8] PRESSURE-FLOW CHARACTERISTICS OF CORONARY STENOSES IN UNSEDATED DOGS AT REST AND DURING CORONARY VASODILATION
    GOULD, KL
    [J]. CIRCULATION RESEARCH, 1978, 43 (02) : 242 - 253
  • [9] GUNDERSEN J, 1972, SCAND J CLIN LAB INV, V29, P371
  • [10] COMPARISON OF FINAPRES NONINVASIVE BEAT-TO-BEAT FINGER BLOOD-PRESSURE WITH INTRABRACHIAL ARTERY PRESSURE DURING AND AFTER BICYCLE ERGOMETRY
    IDEMA, RN
    VANDENMEIRACKER, AH
    IMHOLZ, BPM
    TVELD, AJMI
    SETTELS, JJ
    VANECK, HJR
    SCHALEKAMP, MADH
    [J]. JOURNAL OF HYPERTENSION, 1989, 7 : S58 - S59