Quantitative evaluation of metal artifacts using different CBCT devices, high-density materials and field of views

被引:82
作者
Codari, Marina [1 ,2 ,3 ]
Vasconcelos, Karla de Faria [1 ,2 ,4 ]
Pinheiro Nicolielo, Laura Ferreira [1 ,2 ]
Neto, Francisco Haiter [4 ]
Jacobs, Reinhilde [1 ,2 ]
机构
[1] Katholieke Univ Leuven, Fac Med, Dept Imaging & Pathol, OMFS IMPATH Res Grp,OIC, Leuven, Belgium
[2] Univ Hosp Leuven, Oral & Maxillofacial Surg, Leuven, Belgium
[3] Univ Milan, Dept Biomed Sci Hlth, Fac Med & Surg, Via Mangiagalli 31, I-20133 Milan, Italy
[4] Univ Estadual Campinas, Piracicaba Dent Sch, Dept Oral Diag, Div Oral Radiol, Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
artifacts; cone-beam computed tomography; density materials; field of view; BEAM COMPUTED-TOMOGRAPHY; HORIZONTAL ROOT FRACTURE; IN-VITRO ASSESSMENT; REDUCTION; TITANIUM; IMPLANTS; ABSENCE; TEETH; SIZE; POST;
D O I
10.1111/clr.13019
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
ObjectiveTo objectively compare the influence of different cone-beam computed tomography (CBCT) devices, high-density materials and field of views (FOVs) on metal artifact expression. Material and methodsFor this invitro study, three customized acrylic resin phantoms containing high-density materials cylinders: titanium, copper-aluminum alloy and amalgam were scanned on three CBCT devices using high-resolution protocols, same voxel size (0.2mm) and different FOVs. After fully automatic segmentation and image registration, the same region of interest was defined for the small and medium FOVs. The difference between the segmented and the real volume of the metal cylinders was assessed. Moreover for each segmented slice, the area difference between the segmented and the real axial section was determined. The artifacts on the background were measured as normalizing standard deviation of voxel values in the vicinity of the cylinder, in three different distances. ResultsConsiderable differences were observed in volume measurements for all CBCTs devices and materials for both FOV sizes (up to 67%). The slice per slice area analysis indicated higher artifacts at the edges of the metal cylinder. Within the materials, amalgam and titanium had, respectively, the worst and best artifact expression in all the CBCT devices. Standard deviation values varied differently between the three distances in each device. ConclusionOur invitro study showed that different CBCT devices, high-density materials and FOV should be considered while evaluating CBCT images. More carefully, diagnosis conclusions should be drawn in images containing amalgam and copper-aluminum alloy.
引用
收藏
页码:1509 / 1514
页数:6
相关论文
共 27 条
[1]   In vitro assessment of artifacts induced by titanium dental implants in cone beam computed tomography [J].
Benic, Goran I. ;
Sancho-Puchades, Manuel ;
Jung, Ronald E. ;
Deyhle, Hans ;
Haemmerle, Christoph H. F. .
CLINICAL ORAL IMPLANTS RESEARCH, 2013, 24 (04) :378-383
[2]   Quantitative analysis of metallic artifacts caused by dental metals: comparison of cone-beam and multi-detector row CT scanners [J].
Chindasombatjaroen, Jira ;
Kakimoto, Naoya ;
Murakami, Shumei ;
Maeda, Yoshinobu ;
Furukawa, Souhei .
ORAL RADIOLOGY, 2011, 27 (02) :114-120
[3]   Use of Large-volume Cone-Beam Computed Tomography in Identification and Localization of Horizontal Root Fracture in the Presence and Absence of Intracanal Metallic Post [J].
Costa, Felipe Ferreira ;
Gaia, Bruno Felipe ;
Umetsubo, Otavio Shoiti ;
Pinheiro, Lucas Rodrigues ;
Tortamano, Isabel Peixoto ;
Paraiso Cavalcanti, Marcelo Gusmao .
JOURNAL OF ENDODONTICS, 2012, 38 (06) :856-859
[4]   Detection of Horizontal Root Fracture with Small-volume Cone-Beam Computed Tomography in the Presence and Absence of Intracanal Metallic Post [J].
Costa, Felipe Ferreira ;
Gaia, Bruno Felipe ;
Umetsubo, Otavio Shoiti ;
Paraiso Cavalcanti, Marcelo Gusmao .
JOURNAL OF ENDODONTICS, 2011, 37 (10) :1456-1459
[5]   Detection of vertical root fractures by conventional radiographic examination and cone beam computed tomography - an in vitro analysis [J].
da Silveira, Priscila Fernanda ;
Vizzotto, Mariana Boessio ;
Liedke, Gabriela Salatino ;
Dias da Silveira, Heraldo Luis ;
Montagner, Francisco ;
Dias da Silveira, Heloisa Emilia .
DENTAL TRAUMATOLOGY, 2013, 29 (01) :41-46
[6]  
De Man B, 1999, IEEE T NUCL SCI, V46, P691, DOI 10.1109/23.775600
[7]   Beam hardening artefacts occur in dental implant scans with the NewTom® cone beam CT but not with the dental 4-row multidetector CT [J].
Draenert, F. G. ;
Coppenrath, E. ;
Herzog, P. ;
Muller, S. ;
Mueller-Lisse, U. G. .
DENTOMAXILLOFACIAL RADIOLOGY, 2007, 36 (04) :198-203
[8]  
Esmaeili Farzad, 2012, J Dent Res Dent Clin Dent Prospects, V6, P49
[9]   Principles of CT: Radiation Dose and Image Quality [J].
Goldman, Lee W. .
JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2007, 35 (04) :213-225
[10]   State-of-the-art on cone beam CT imaging for preoperative planning of implant placement [J].
Guerrero, ME ;
Jacobs, R ;
Loubele, M ;
Schutyser, F ;
Suetens, P ;
van Steenberghe, D .
CLINICAL ORAL INVESTIGATIONS, 2006, 10 (01) :1-7