A novel system for the direct measurement of the electron energy distribution function by Langmuir probe in ion sources

被引:6
作者
Homfray, D. [1 ]
Crowley, B. [1 ]
机构
[1] UKAEA Euratom Fus Assoc, Culham Sci Ctr, Abingdon OX14 3DB, Oxon, England
基金
欧盟地平线“2020”; 英国工程与自然科学研究理事会;
关键词
Langmuir probe; electron energy distribution function; Boyd and Twiddy; second derivative;
D O I
10.1016/j.fusengdes.2007.08.005
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
To realise a high yield H- (or D-) ion source capable of long pulse operation for the ITER neutral beam heating system, it is vital to understand and optimise the plasma parameters in the ion source. Calculating the electron energy distribution function (EEDF) in a non-Maxwellian plasma using conventional Langmuir probe numerical techniques is prone to error and generally produces poor results. Using a technique developed by Boyd and Twiddy [R.L.F. Boyd, N.D. Twiddy, Proc. R. Soc. A 53 (1959) 250] in the 1950s a new Langmuir probe system has been developed. This paper presents a detailed account of the experimental method, apparatus, software and analysis. The impact of the development is demonstrated by comparing data obtained from the ITER Negative Ion Beam Development Test Stand at CEA Cadarache using this method with similar data obtained with conventional methods. Crown Copyright (C) 2007 Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:829 / 835
页数:7
相关论文
共 10 条
[1]   THE COLLECTION OF POSITIVE IONS BY A PROBE IMMERSED IN A PLASMA [J].
ALLEN, JE ;
BOYD, RLF ;
REYNOLDS, P .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION B, 1957, 70 (03) :297-304
[2]   THEORY OF ELECTROSTATIC PROBES IN A LOW-DENSITY PLASMA [J].
BERNSTEIN, IB ;
RABINOWITZ, IN .
PHYSICS OF FLUIDS, 1959, 2 (02) :112-121
[3]  
Boyd R L F., 1959, P ROY SOC, V53, P250, DOI [10.1098/rspa.1959.0050, DOI 10.1098/RSPA.1959.0050]
[4]  
CROWLEY B, 1966, NUCL FUSION, V46, pS307
[5]  
Druyvesteyn M J, 1930, Z PHYS, V54, P790
[6]   EFFECT OF AC AMPLTIUDE ON MEASUREMENT OF ELECTRON ENERGY DISTRIBUTION FUNCTIONS [J].
KILVINGTON, AI ;
JONES, RP ;
SWIFT, JD .
JOURNAL OF SCIENTIFIC INSTRUMENTS, 1967, 44 (07) :517-+
[7]  
Laframboise J. G., 1966, Theory of spherical and cylindrical Langmuir probes in a collisionless, Maxwellian plasma at rest
[8]  
LANGMUIR I, 1924, GEN ELECTR REV, V27, P449, DOI DOI 10.1103/PHYSREV.28.727
[9]   Measurement of EEDF and distribution of primary electrons in a bucket ion source [J].
Miyamoto, S ;
Kanayama, R ;
Minami, T ;
Mune, T ;
Horiike, H .
NUCLEAR FUSION, 2006, 46 (06) :S313-S317
[10]   Negative ion sources for neutral beam injection into fusion machines [J].
Trainham, R ;
Jacquot, C ;
Riz, D ;
Simonin, A ;
Miyamoto, K ;
Fujiwara, Y ;
Okumura, Y .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (02) :926-928