Comparison of ultrasonic pachymetry with orbscan in corneal haze

被引:14
作者
Altan-Yaycioglu, Rana
Pelit, Aysel
Akova, Yonca Aydin
机构
[1] Baskent Univ Adana, Teaching & Med Res Ctr, Dept Ophthalmol, TR-01250 Adana, Turkey
[2] Baskent Univ, Dept Ophthalmol, Fac Med, TR-06490 Ankara, Turkey
关键词
corneal topography; corneal opacity; keratitis; trachoma; corneal dystrophy;
D O I
10.1007/s00417-007-0578-5
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background To compare the results of ultrasonic and Orbscan pachymetry in corneal haze not related to surgery. Methods An institutionally based, prospective, controlled clinical trial included 40 eyes of 40 consecutive patients with corneal haze not related to surgery. Fifty normal eyes were used as a control group to calculate the customized acoustic factor. Corneal haze was graded with slit-lamp by a single examiner. Ultrasonic and Orbscan II pachymetry measurements were obtained. Paired t- and Kruskal-Wallis tests were used to evaluate the difference between ultrasonic and Orbscan pachymetry measurements. Correlation of the two methods and haze grade were determined with Kendall's tau-b correlation analysis. Results Eighty-five percent of patients had moderate haze. With the customized acoustic factor, Orbscan values were significantly smaller than ultrasonic pachymetry measurements (p=0.001, paired samplet-test). Increased haze grade resulted in smaller Orbscan pachymetry measurements however, this difference was insignificant (p=0.444, Kruskal-Wallis test). A positive correlation was found between two methods, though insignificant (p=0.064, Kendall's tau-b correlation). When grouped in respect of diagnoses, differences in pachymetry measurements as well as the correlation were not significant (p > 0.05). Conclusions An insignificant positive correlation was found between the two methods in the presence of corneal haze. Orbscan measurements were inversely related to haze grading, and were significantly lower than ultrasonic pachymetry. According to our results, we recommend that clinicians should not count on Orbscan topography alone in the presence of corneal haze.
引用
收藏
页码:1759 / 1763
页数:5
相关论文
共 15 条
  • [1] Accuracy of Orbscan optical pachymetry in corneas with haze
    Boscia, F
    La Tegola, MG
    Alessio, G
    Sborgia, C
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) : 253 - 258
  • [2] Accuracy of Orbscan II slit-scanning elevation topography
    Cairns, G
    McGhee, CNJ
    Collins, MJ
    Owens, H
    Gamble, GD
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (12) : 2181 - 2187
  • [3] Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes
    Chakrabarti, HS
    Craig, JP
    Brahma, A
    Malik, TY
    McGhee, CNJ
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (11) : 1823 - 1828
  • [4] Cheng ACK, 2006, J REFRACT SURG, V22, P363
  • [5] Comparison of corneal pachymetry using ultrasound and Orbscan II
    Fakhry, MA
    Artola, A
    Belda, JL
    Ayala, MJ
    Alió, JL
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) : 248 - 252
  • [6] WOUND-HEALING AFTER EXCIMER LASER KERATOMILEUSIS (PHOTOREFRACTIVE KERATECTOMY) IN MONKEYS
    FANTES, FE
    HANNA, KD
    WARING, GO
    POULIQUEN, Y
    THOMPSON, KP
    SAVOLDELLI, M
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1990, 108 (05) : 665 - 675
  • [7] Central and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry
    González-Méijome, JM
    Cerviño, A
    Yebra-Pimentel, E
    Parafita, MA
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01) : 125 - 132
  • [8] Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures
    Iskander, NG
    Penno, EA
    Peters, NT
    Gimbel, HV
    Ferensowicz, M
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (05) : 681 - 685
  • [9] Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound
    Lackner, B
    Schmidinger, G
    Pieh, S
    Funovics, MA
    Skorpik, C
    [J]. OPTOMETRY AND VISION SCIENCE, 2005, 82 (10) : 892 - 899
  • [10] The repeatability of corneal thickness measures
    Marsich, MM
    Bullimore, MA
    [J]. CORNEA, 2000, 19 (06) : 792 - 795