Measurement of breath acetone concentrations by selected ion flow tube mass spectrometry in type 2 Diabetes

被引:76
|
作者
Storer, Malina [1 ]
Dummer, Jack [2 ]
Lunt, Helen [3 ]
Scotter, Jenny [1 ]
McCartin, Fiona [2 ]
Cook, Julie [2 ]
Swanney, Maureen [2 ]
Kendall, Deborah [3 ]
Logan, Florence [3 ]
Epton, Michael [2 ]
机构
[1] Syft Technol Ltd, Christchurch, New Zealand
[2] Christchurch Hosp, Dept Resp Med, Christchurch, New Zealand
[3] Christchurch Hosp, Ctr Diabet, Christchurch, New Zealand
关键词
BLOOD-GLUCOSE; ANALYZER; AMMONIA; SIFT;
D O I
10.1088/1752-7155/5/4/046011
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Selected ion flow tube-mass spectrometry (SIFT-MS) can measure volatile compounds in breath on-line in real time and has the potential to provide accurate breath tests for a number of inflammatory, infectious and metabolic diseases, including diabetes. Breath concentrations of acetone in type 2 diabetic subjects undertaking a long-term dietary modification programme were studied. Acetone concentrations in the breath of 38 subjects with type 2 diabetes were determined by SIFT-MS. Anthropomorphic measurements, dietary intake and medication use were recorded. Blood was analysed for beta hydroxybutyrate (a ketone body), HbA1c (glycated haemoglobin) and glucose using point-of-care capillary (fingerprick) testing. All subjects were able to undertake breath manoeuvres suitable for analysis. Breath acetone varied between 160 and 862 ppb (median 337 ppb) and was significantly higher in men (median 480 ppb versus 296 ppb, p = 0.01). In this cross-sectional study, no association was observed between breath acetone and either dietary macronutrients or point-of-care capillary blood tests. Breath analysis by SIFT-MS offers a rapid, reproducible and easily performed measurement of acetone concentration in ambulatory patients with type 2 diabetes. The high inter-individual variability in breath acetone concentration may limit its usefulness in cross-sectional studies. Breath acetone may nevertheless be useful for monitoring metabolic changes in longitudinal metabolic studies, in a variety of clinical and research settings.
引用
收藏
页数:5
相关论文
共 50 条
  • [41] Evaluation of the performance of field olfactometers by selected ion flow tube mass spectrometry
    Walgraeve, Christophe
    Van Hufffel, Katrijn
    Bruneel, Joren
    Van Langenhove, Herman
    BIOSYSTEMS ENGINEERING, 2015, 137 : 84 - 94
  • [42] Quantitative selected ion flow tube mass spectrometry: The influence of ionic diffusion and mass discrimination
    Spanel, P
    Smith, D
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2001, 12 (07) : 863 - 872
  • [43] In situ analysis of solvents on breath and blood:: a selected ion flow tube mass spectrometric study
    Wilson, PF
    Freeman, CG
    McEwan, MJ
    Milligan, DB
    Allardyce, RA
    Shaw, GM
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (05) : 427 - 432
  • [44] Quantification of breath isoprene using the selected ion flow tube mass spectrometric analytical method
    Spanel, P
    Davies, S
    Smith, D
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1999, 13 (17) : 1733 - 1738
  • [45] Selected ion flow tube mass spectrometry of 3-hydroxybutyric acid, acetone and other ketones in the headspace of aqueous solution and urine
    Wang, Tianshu
    Spanel, Patrik
    Smith, David
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2008, 272 (01) : 78 - 85
  • [46] Breast cancer detection using volatile compound profiles in exhaled breath via selected ion-flow tube mass spectrometry
    Nakayama, Yoshie
    Hanada, Mariko
    Koda, Hiroshi
    Sugimoto, Masahiro
    Takada, Masahiro
    Toi, Masakazu
    JOURNAL OF BREATH RESEARCH, 2023, 17 (01)
  • [47] Quantification of methyl thiocyanate in the headspace of Pseudomonas aeruginosa cultures and in the breath of cystic fibrosis patients by selected ion flow tube mass spectrometry
    Shestivska, Violetta
    Nemec, Alexandr
    Drevinek, Pavel
    Sovova, Kristyna
    Dryahina, Kseniya
    Spanel, Patrik
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2011, 25 (17) : 2459 - 2467
  • [48] A longitudinal study of ethanol and acetaldehyde in the exhaled breath of healthy volunteers using selected-ion flow-tube mass spectrometry
    Turner, C
    Spanel, P
    Smith, D
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2006, 20 (01) : 61 - 68
  • [49] Selected Ion Flow Tube Mass Spectrometry Analysis of Exhaled Breath for Volatile Organic Compound Profiling of Esophago-Gastric Cancer
    Kumar, Sacheen
    Huang, Juzheng
    Abbassi-Ghadi, Nima
    Spanel, Patrik
    Smith, David
    Hanna, George B.
    ANALYTICAL CHEMISTRY, 2013, 85 (12) : 6121 - 6128
  • [50] Concentrations of some metabolites in the breath of healthy children aged 7-18 years measured using selected ion flow tube mass spectrometry (SIFT-MS)
    Enderby, B.
    Lenney, W.
    Brady, M.
    Emmett, C.
    Spanel, P.
    Smith, D.
    JOURNAL OF BREATH RESEARCH, 2009, 3 (03)