Mechanical comparison in cadaver specimens of three different 90-degree double-plate osteosyntheses for simulated C2-type distal humerus fractures with varying bone densities

被引:100
作者
Schuster, Ivo [1 ,2 ]
Korner, Jan [1 ,3 ]
Arzdorf, Michael [1 ]
Schwieger, Karsten [1 ]
Diederichs, Gerd [1 ]
Linke, Berend [1 ]
机构
[1] AO Res Inst, CH-7270 Davos, Switzerland
[2] Univ Freiburg, Dept Orthopaed & Trauma Surg, Freiburg, Germany
[3] Johannes Gutenberg Univ Mainz, Dept Trauma Surg, Mainz, Germany
关键词
distal humerus fracture; comminution; bone mineral density; biomechanics; screw pullout;
D O I
10.1097/BOT.0b013e3181632cf8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To investigate the bone-implant-anchorage of 90-degree double-plate osteosynthesis in simulated complete intra-articular distal humerus fractures using conventional reconstruction plates (CRP), locking compression plates (LCP), and distal humerus plates (DHP), depending on the bone mineral density (BMD) of the cadaver specimens. Methods: Groups (CRP, LCP, DHP, n = 8; LCP, DHP, n = 13) in distal humerus cadaver bones were created based on BMD. The fracture model was an unstable intraarticular distal humerus fracture with a transverse osteotomy gap representing metaphyseal comminution (AO type 13-C2.3). Flexion and extension stiffness as well as cycles until failure due to screw pullout under cyclic loading were evaluated. Estimates of BMD values, below which failure was likely to occur, were determined. Results: Stiffness values were not significantly different between groups (extension: P = 0.881, flexion: P = 0.547). Under cyclic loading, consistent screw pullout failure occur-red at BMD values below about 400 mg/cm(3) for CRP and below about 300 mg/cm(3) for LCP constructs. Comparing BMD-matched groups of 8 and 13 specimens respectively, the failure rate was significantly lower for the DHP (0/8) than for the CRP (5/8; P = 0.026) and tended to be lower for the DHP (0/13) as compared to the LCP (4/13; P = 0.096). Conclusion: Bone-implant anchorage was different between locking and nonlocking plate constructs and depended on BMD. While in good bone quality implant choice was not critical, both locking plates provided superior resistance against screw loosening as compared to the CRP at low BMD values (<420 mg/cm(3)). Based on our laboratory results, we conclude that locking plates such as the LCP and DHP are constructs designed to keep anatomical reduction in the presence of comminution and poor bone quality in a low intraarticular fracture of the distal humerus.
引用
收藏
页码:113 / 120
页数:8
相关论文
共 27 条
[1]   Assessment of bone quality in the proximal humerus by measurement of the contralateral site: a cadaveric analyze [J].
Diederichs, G ;
Korner, J ;
Goldhahn, J ;
Linke, B .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2006, 126 (02) :93-100
[2]   Mechanical properties of cancellous bone of the distal humerus [J].
Dunham, CE ;
Takaki, SE ;
Johnson, JA ;
Dunning, CE .
CLINICAL BIOMECHANICS, 2005, 20 (08) :834-838
[3]   Intercondylar fractures of the distal humerus in adults: a critical analysis of 55 cases [J].
Gupta, R ;
Khanchandani, P .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2002, 33 (06) :511-515
[4]   TRANSMISSION OF PRESSURES ACROSS ELBOW JOINT [J].
HALLS, AA ;
TRAVILL, A .
ANATOMICAL RECORD, 1964, 150 (03) :243-&
[5]   Treatment of distal humerus fractures in the elderly [J].
Hausman, M ;
Panozzo, A .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (425) :55-63
[6]  
Helfet D L, 1990, J Orthop Trauma, V4, P260, DOI 10.1097/00005131-199004030-00004
[7]  
Helfet David L, 2004, J Bone Joint Surg Am, V86-A Suppl 1, P18
[8]  
HELFET DL, 1993, CLIN ORTHOP RELAT R, V292, P26
[9]  
Henley M B, 1987, J Orthop Trauma, V1, P24, DOI 10.1097/00005131-198701010-00004
[10]   INTERCONDYLAR FRACTURES OF THE HUMERUS - AN OPERATIVE APPROACH [J].
JUPITER, JB ;
NEFF, U ;
HOLZACH, P ;
ALLGOWER, M .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (02) :226-239