Automation of Monte Carlo-based treatment plan verification for proton therapy

被引:8
作者
Kaluarachchi, Maduka [1 ]
Moskvin, Vadim [1 ]
Pirlepesov, Fakhriddin [1 ]
Wilson, Lydia J. [1 ]
Xie, Fang [1 ]
Faught, Austin M. [1 ]
机构
[1] St Jude Childrens Res Hosp, Dept Radiat Oncol, 332 N Lauderdale St, Memphis, TN 38105 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2020年 / 21卷 / 08期
关键词
automation; Monte Carlo; proton therapy; BRAIN-STEM; BRAGG PEAK; SIMULATION; HETEROGENEITIES; PARAMETERS; TOPAS;
D O I
10.1002/acm2.12923
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Independent calculations of proton therapy plans are an important quality control procedure in treatment planning. When using custom Monte Carlo (MC) models of the beamline, deploying the calculations can be laborious, time consuming, and require in-depth knowledge of the computational environment. We developed an automated framework to remove these barriers and integrate our MC model into the clinical workflow. Materials and Methods The Eclipse Scripting Application Programming Interface was used to initiate the automation process. A series of MATLAB scripts were then used for preprocessing of input data and postprocessing of results. Additional scripts were used to monitor the calculation process and appropriately deploy calculations to an institutional high-performance computing facility. The automated framework and beamline models were validated against 160 patient specific QA measurements from an ionization chamber array and using a +/- 3%/3 mm gamma criteria. Results The automation reduced the human-hours required to initiate and run a calculation to 1-2 min without leaving the treatment planning system environment. Validation comparisons had an average passing rate of 99.4% and were performed at depths ranging from 1 to 15 cm. Conclusion An automated framework for running MC calculations was developed which enables the calculation of dose and linear energy transfer within a clinically relevant workflow and timeline. The models and framework were validated against patient specific QA measurements and exhibited excellent agreement. Before this implementation, execution was prohibitively complex for an untrained individual and its use restricted to a research environment.
引用
收藏
页码:131 / 138
页数:8
相关论文
共 50 条
  • [41] A Monte Carlo dose calculation algorithm for proton therapy
    Fippel, M
    Soukup, M
    MEDICAL PHYSICS, 2004, 31 (08) : 2263 - 2273
  • [42] Monte Carlo-based scatter correction for the SMARTZOOM collimator
    Dietze, Martijn M. A.
    Kunnen, Britt
    Stella, Martina
    de Jong, Hugo W. A. M.
    EJNMMI PHYSICS, 2020, 7 (01)
  • [43] Proton therapy analysis using the Monte Carlo method
    Noshad, H
    Givechi, N
    RADIATION MEASUREMENTS, 2005, 39 (05) : 521 - 524
  • [44] Effective clinical applications of Monte Carlo-based independent secondary dose verification software for helical tomotherapy
    Kosaka, Takahiro
    Takatsu, Jun
    Inoue, Tatsuya
    Hara, Naoya
    Mitsuhashi, Taira
    Suzuki, Michimasa
    Shikama, Naoto
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 104 : 112 - 122
  • [45] An algorithm to assess the need for clinical Monte Carlo dose calculation for small proton therapy fields based on quantification of tissue heterogeneity
    Bueno, M.
    Paganetti, H.
    Duch, M. A.
    Schuemann, J.
    MEDICAL PHYSICS, 2013, 40 (08)
  • [46] Monte Carlo-based scatter correction for the SMARTZOOM collimator
    Martijn M. A. Dietze
    Britt Kunnen
    Martina Stella
    Hugo W. A. M. de Jong
    EJNMMI Physics, 7
  • [47] A fast Monte Carlo code for proton transport in radiation therapy based on MCNPX
    Jabbari, Keyvan
    Seuntjens, Jan
    JOURNAL OF MEDICAL PHYSICS, 2014, 39 (03) : 156 - 163
  • [48] Evaluation of proton therapy dose in breast cancer with boron agent based on GATE Monte Carlo code
    Bijarpasi, Fereshteh Mollaee
    Sadremomtaz, Alireza
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2025, 18 (01)
  • [49] RayStation/GATE Monte Carlo simulation framework for verification of proton therapy based on the 12N imaging
    Ahmadi Ganjeh, Zahra
    Zapien-Campos, Brian
    Traneus, Erik
    Both, Stefan
    Dendooven, Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (19)
  • [50] Independent dose verification system with Monte Carlo simulations using TOPAS for passive scattering proton therapy at the National Cancer Center in Korea
    Shin, Wook-Geun
    Testa, Mauro
    Kim, Hak Soo
    Jeong, Jong Hwi
    Lee, Se Byeong
    Kim, Yeon-Joo
    Min, Chul Hee
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (19) : 7598 - 7616