Computational tools for thoracic endovascular aortic repair planning

被引:2
作者
Auricchio, Ferdinando [1 ]
Conti, Michele [1 ]
Romarowski, Rodrigo M. [2 ]
De Beaufort, Hector W. [3 ]
Grassi, Viviana [3 ]
Trimarchi, Santi [3 ,4 ]
机构
[1] Univ Pavia, Dept Civil Engn & Architecture, Pavia, Italy
[2] IRCCS Policlin San Donato, 3D & Comp Simulat Lab, Milan, Italy
[3] IRCCS Policlin San Donato, Thorac Aort Res Ctr, Via Morandi 30, I-20097 Milan, Italy
[4] Univ Milan, Dept Biomed Sci Hlth, Milan, Italy
来源
ITALIAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY | 2019年 / 26卷 / 01期
关键词
Aorta; thoracic; Endovascular procedures; Review; STENT-GRAFT DEPLOYMENT; FLUID-DYNAMICS; ENDOLEAK FORMATION; RISK; SIMULATION; HEMODYNAMICS; MULTICENTER; ANEURYSM; DESIGN;
D O I
10.23736/S1824-4777.18.01386-4
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Endovascular treatment in the thoracic aorta has mostly replaced open surgery in subjects with given anatomical requirements, showing excellent postoperative results. The selection of the right stent-graft is based on aortic morphology and operator skills, with the goal of guaranteeing good apposition and long-term stability. However, complications such as endoleak and migration can occur after thoracic endovascular aortic repair, mostly caused by a weakened mechanical interaction between the wall, the blood and the device. The goal of this review is to explore the current effort of the biomechanics community to predict in-hospital and long-term outcomes by means of computational simulations. Benchtop tests, finite element analysis and computational fluid dynamics will be explored together with several examples in order to show the current state-of-the-art for surgical planning and complications prediction. Our analysis shows the potentiality of numerical tools not only to foresee problems which cannot be diagnosed by current tools, but also that the proof-of-concept stage has been passed and that the technology is ready for the bedside.
引用
收藏
页码:51 / 58
页数:8
相关论文
共 52 条
[1]   Evaluation of the hemodynamic effectiveness of aortic dissection treatments via virtual stenting [J].
Alimohammadi, Mona ;
Bhattacharya-Ghosh, Benjamin ;
Seshadhri, Santhosh ;
Penrose, Justin ;
Agu, Obiekezie ;
Balabani, Stavroula ;
Díaz-Zuccarini, Vanessa .
International Journal of Artificial Organs, 2014, 37 (10) :753-762
[2]   Development of a patient-specific simulation tool to analyse aortic dissections: Assessment of mixed patient-specific flow and pressure boundary conditions [J].
Alimohammadi, Mona ;
Agu, Obiekezie ;
Balabani, Stavroula ;
Diaz-Zuccarini, Vanessa .
MEDICAL ENGINEERING & PHYSICS, 2014, 36 (03) :275-284
[3]   Morphological and stent design risk factors to prevent migration phenomena for a thoracic aneurysm: A numerical analysis [J].
Altnji, H. -E. ;
Bou-Said, B. ;
Walter-Le Berre, H. .
MEDICAL ENGINEERING & PHYSICS, 2015, 37 (01) :23-33
[4]  
Arthurs CJ, 2018, EUR J CARDIO THORACI
[5]   Patient-specific analysis of post-operative aortic hemodynamics: a focus on thoracic endovascular repair (TEVAR) [J].
Auricchio, F. ;
Conti, M. ;
Lefieux, A. ;
Morganti, S. ;
Reali, A. ;
Sardanelli, F. ;
Secchi, F. ;
Trimarchi, S. ;
Veneziani, A. .
COMPUTATIONAL MECHANICS, 2014, 54 (04) :943-953
[6]   Patient-specific aortic endografting simulation: From diagnosis to prediction [J].
Auricchio, F. ;
Conti, M. ;
Marconi, S. ;
Reali, A. ;
Tolenaar, Jip L. ;
Trimarchi, S. .
COMPUTERS IN BIOLOGY AND MEDICINE, 2013, 43 (04) :386-394
[7]   A Clinically Applicable Stochastic Approach for Noninvasive Estimation of Aortic Stiffness Using Computed Tomography Data [J].
Auricchio, Ferdinando ;
Conti, Michele ;
Ferrara, Anna ;
Lanzarone, Ettore .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (01) :176-187
[8]  
Berezowski M, 2018, EUR J VASC ENDOVASC
[9]   Thoracic Endovascular Aortic Repair of Aortic Arch Pathologies with the Conformable Thoracic Aortic Graft: Early and 2 year Results from a European Multicentre Registry [J].
Boeckler, D. ;
Brunkwall, J. ;
Taylor, P. R. ;
Mangialardi, N. ;
Huesing, J. ;
Larzon, T. .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2016, 51 (06) :791-800
[10]   Virtual evaluation of stent graft deployment: A validated modeling and simulation study [J].
De Bock, S. ;
Iannaccone, F. ;
De Santis, G. ;
De Beule, M. ;
Van Loo, D. ;
Devos, D. ;
Vermassen, F. ;
Segers, P. ;
Verhegghe, B. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 13 :129-139