Evaluation of Electromagnetic Noise Emitted from Light-Emitting Diode (LED) Lamps and Compatibility with Wireless Medical Telemetry Service

被引:4
作者
Ishida, Kai [1 ]
Wu, Ifong [1 ]
Gotoh, Kaoru [1 ]
Matsumoto, Yasushi [1 ]
机构
[1] Natl Inst Informat & Commun Technol, Koganei, Tokyo 1848795, Japan
关键词
wireless medical telemetry service; switching regulator; LED lamp; electromagnetic compatibility; receiver sensitivity; IMPULSIVENESS CORRECTION FACTOR; AMPLITUDE PROBABILITY-DISTRIBUTION; PERFORMANCE; IMPACT; INTERFERENCE; SYSTEM;
D O I
10.1587/transcom.2019HMP0003
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Wireless medical telemetry service (WMTS) is an important wireless communication system in healthcare facilities. Recently, the potential for electromagnetic interference by noise emitted by switching regulators installed in light-emitting diode (LED) lamps has been a serious problem. In this study, we evaluated the characteristics of the electromagnetic noise emitted from LED lamps and its effect on WMTS. Switching regulators generally emit wide band impulsive noise whose bandwidth reaches 400 MHz in some instances owing to the switching operation, but this impulsive nature is difficult to identify in the reception of WMTS because the bandwidth of WMTS is much narrower than that of electromagnetic noise. Gaussian approximation (GA) can be adopted for band-limited electromagnetic noise whose characteristics have no repetitive variation. On the other hand, GA with the impulsive correction factor (ICF) can be adopted for band-limited electromagnetic noise that has repetitive variation. We investigate the minimum receiver sensitivity of WMTS for it to be affected by electromagnetic noise emitted from LED lamps. The required carrier-to-noise power ratio (CNR) of Gaussian noise and electromagnetic noise for which GA can be adopted was approximately 15 dB, but the electromagnetic noise for which GA with the ICF can be adopted was 3 to 4 dB worse. Moreover, the spatial distribution of electromagnetic noise surrounding an LED lamp installation was measured. Finally, we roughly estimated the offset distance between the receiving antenna of WMTS and LED lamps when a WMTS signal of a certain level was added in a clinical setting using our experimental result for the required CNR.
引用
收藏
页码:637 / 644
页数:8
相关论文
共 17 条
[1]  
[Anonymous], 2015, STD21 RCR ASS RAD IN
[2]  
[Anonymous], 2018, 15 CISPR INT SPEC CO
[3]  
Electromagnetic Compatibility Conference Japan, 2016, EL COMP C JAP GUID U
[4]   A Simple Measurement Method to Derive the Impulsiveness Correction Factor for Communication Performance Estimation [J].
Fors, Karina M. ;
Wiklundh, Kia C. ;
Stenumgaard, Peter F. .
IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2013, 55 (05) :834-841
[5]   APD measurement for evaluating disturbances related to the performance of digital communication systems [J].
Gotoh, K ;
Matsumoto, Y ;
Yamanaka, Y ;
Shinozuka, T .
IEICE TRANSACTIONS ON COMMUNICATIONS, 2005, E88B (08) :3235-3241
[6]   Newly identified electromagnetic problems with medical telemetry systems [J].
Hanada, Eisuke ;
Ishida, Kai ;
Kudou, Takato .
PRZEGLAD ELEKTROTECHNICZNY, 2018, 94 (02) :21-24
[7]   Impact of LED lamp noise on receiver sensitivity of wireless medical telemetry system [J].
Ishida, Kai ;
Arie, Sazu ;
Wu, Ifong ;
Gotoh, Kaoru ;
Matsumoto, Yasushi .
IEICE COMMUNICATIONS EXPRESS, 2018, 7 (11) :421-426
[8]   Electromagnetic compatibility of wireless medical telemetry systems and light-emitting diode (LED) lamps [J].
Ishida, Kai ;
Arie, Sazu ;
Gotoh, Kaoru ;
Hanada, Eisuke ;
Hirose, Minoru ;
Matsumoto, Yasushi .
PRZEGLAD ELEKTROTECHNICZNY, 2018, 94 (02) :25-28
[9]   Evaluation of Electromagnetic Fields in a Hospital for Safe Use of Electronic Medical Equipment [J].
Ishida, Kai ;
Fujioka, Tomomi ;
Endo, Tetsuo ;
Hosokawa, Ren ;
Fujisaki, Tetsushi ;
Yoshino, Ryoji ;
Hirose, Minoru .
JOURNAL OF MEDICAL SYSTEMS, 2016, 40 (03) :1-11
[10]  
Japan Lighting Manufacturers Association, LIGHT VIS 2020 VER 2