Wear properties of poly-ether-ether-ketone bearing combinations under zero and cross shear kinematics in total knee arthroplasty

被引:11
作者
Chamberlain, Kathryn A. [1 ]
Rankin, Kenneth S. [2 ]
Briscoe, Adam [3 ]
Deehan, David [2 ]
Hyde, Philip J. [1 ]
机构
[1] Newcastle Univ, Sch Engn, Bioengn Grp, Newcastle Upon Tyne, Tyne & Wear, England
[2] Freeman Rd Hosp, Newcastle Upon Tyne, Tyne & Wear, England
[3] Hillhouse Int, Invibio Global Technol Ctr, Thornton Cleveleys FY5 4QD, Lancs, England
关键词
PEEK; wear; all-polymer wear; cross shear; pin-on-plate; MOLECULAR-WEIGHT POLYETHYLENE; TOTAL DISC REPLACEMENT; PIN-ON-PLATE; CFR-PEEK; CONTACT PRESSURE; TRANSFER FILM; JOINT; WETTABILITY; QUANTIFICATION; PERFORMANCE;
D O I
10.1002/jbm.b.34136
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Poly-ether-ether-ketone (PEEK), having shown favorable biocompatibility in spinal applications is being considered as an alternative biomaterial in orthopaedics, either as part of an all-polymer bearing couple, or a replacement for the metallic component in hard-on-soft bearings. Throughout the literature ultra-high molecular weight polyethylene exhibits a range of wear behavior dependent upon the amount of cross shear (CS) present in the bearing motion; in comparison, the behavior of PEEK bearing combinations subject to cross shear conditions is less understood. The aim of this study was to investigate the effect of cross shear on PEEK-on-PEEK and PEEK-on-Metal bearing couples. Wear tests were conducted using a four station pin-on-plate rig capable of uni-directional motion (zero cross shear) and multidirectional motion (cross shear); reciprocation (1 Hz), rotation (0 or 1 Hz), with gravimetric wear analysis used to determine the wear factors. The combined wear factors from the PEEK pins articulating on either PEEK or metal plates in this study suggest that it is preferable to use PEEK-on-Metal bearing couples under zero cross shear kinematic conditions and PEEK-on-PEEK for high cross shear applications. PEEK appears to demonstrate a CS dependency when articulating on hard surfaces. (c) 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 107B: 445-453, 2019.
引用
收藏
页码:445 / 453
页数:9
相关论文
共 45 条
  • [1] A comprehensive combined experimental and computational framework for pre-clinical wear simulation of total knee replacements
    Abdelgaied, A.
    Fisher, J.
    Jennings, L. M.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 78 : 282 - 291
  • [2] Quantification of the effect of cross-shear and applied nominal contact pressure on the wear of moderately cross-linked polyethylene
    Abdelgaied, Abdellatif
    Brockett, Claire L.
    Liu, Feng
    Jennings, Louise M.
    Fisher, John
    Jin, Zhongmin
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2013, 227 (H1) : 18 - 26
  • [3] The Biotribology of PEEK-on-HXLPE Bearings Is Comparable to Traditional Bearings on a Multidirectional Pin-on-disk Tester
    Baykal, Doruk
    Siskey, Ryan S.
    Underwood, Richard J.
    Briscoe, Adam
    Kurtz, Steven M.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (11) : 2384 - 2393
  • [4] The difference of material wettability as critical factor in the choice of a tribological prosthetic coupling without debris release
    Borruto, A
    Marrelli, L
    Palma, F
    [J]. TRIBOLOGY LETTERS, 2005, 20 (01) : 1 - 10
  • [5] Influence of contact pressure, cross-shear and counterface material on the wear of PEEK and CFR-PEEK for orthopaedic applications
    Brockett, C. L.
    Carbone, S.
    Abdelgaied, A.
    Fisher, J.
    Jennings, L. M.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 63 : 10 - 16
  • [6] PEEK and CFR-PEEK as alternative bearing materials to UHMWPE in a fixed bearing total knee replacement: An experimental wear study.
    Brockett, Claire L.
    Carbone, Silvia
    Fisher, John
    Jennings, Louise M.
    [J]. WEAR, 2017, 374 : 86 - 91
  • [7] The use of self-mating PEEK as an alternative bearing material for cervical disc arthroplasty: a comparison of different simulator inputs and tribological environments
    Brown, Tim
    Bao, Qi-Bin
    [J]. EUROPEAN SPINE JOURNAL, 2012, 21 : S717 - S726
  • [8] Wettability of porous surfaces.
    Cassie, ABD
    Baxter, S
    [J]. TRANSACTIONS OF THE FARADAY SOCIETY, 1944, 40 : 0546 - 0550
  • [9] Cooper J. R., 1993, Clinical Materials, V14, P295, DOI 10.1016/0267-6605(93)90016-Z
  • [10] PEEK-OPTIMA as an alternative to cobalt chrome in the femoral component of total knee replacement: A preliminary study
    Cowie, Raelene M.
    Briscoe, Adam
    Fisher, John
    Jennings, Louise M.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (11) : 1008 - 1015