Patient result median monitoring for clinical laboratory quality control

被引:30
作者
Wilson, Andrew [1 ]
Roberts, William L. [1 ,2 ]
Pavlov, Igor [1 ]
Fontenot, Jo [1 ]
Jackson, Brian [1 ,2 ]
机构
[1] Inst Clin & Expt Pathol, ARUP Labs, Salt Lake City, UT USA
[2] Univ Utah, Hlth Sci Ctr, Dept Pathol, Salt Lake City, UT 84132 USA
关键词
Statistical process control; Median monitoring; Limiting distributions; SPECIFICATIONS; DESIGN; CHART;
D O I
10.1016/j.cca.2011.04.024
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Statistical process control is foundational in laboratory medicine. It typically uses artificial control specimens and can detect some, but not all, analytical defects. A practical, robust method to more directly detect trends in patient results, such as monitoring mean or median patient results, is desirable. Methods: We generated a simulated set of laboratory results from a normal distribution, and also downloaded sequential patient results for serum sodium and CA 19-9. For each of the three data sets we calculated the standard error of the mean and estimated the standard error of the median by bootstrapping. Results: The standard error of the mean is a practical, easily calculated summary statistic that can be used to construct control charts. The standard error of the median, cannot be reliably estimated without using bootstrap methods, but is more resistant to outliers. Our study confirms a simple relationship between the variance of the median and the variance of the mean, i.e., for Gaussian distributions, Var[Median]/Var[Median] = pi/2 We also confirm that for skewed distributions, the median is more stable than the mean, implying Var[Median]/Var[Mean] < 1. Finally, we establish a sample size of 200 individual patient results as sufficient for monitoring medians for data from approximately Gaussian distributions. Conclusions: Monitoring patient result medians represents a practical, statistically self-consistent approach to laboratory quality control. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:1441 / 1446
页数:6
相关论文
共 17 条
[1]  
Abu-Shawiesh Moustafa Omar Ahmed, 2008, Journal of Mathematics and Statistics, V4, P102, DOI 10.3844/jmssp.2008.102.107
[2]  
BULL B, 1982, METHODS HEMATOLOGY Q
[3]   CHARACTERIZATION OF THE PROTEIN MATRIX OF QUALITY-CONTROL SERA BY A HIGH-RESOLUTION TWO-DIMENSIONAL ELECTROPHORESIS TECHNIQUE [J].
CLARK, PMS ;
KRICKA, LJ ;
GOMO, ARZ ;
WHITEHEAD, TP ;
BULLOCK, DG ;
SAIDI, H .
CLINICA CHIMICA ACTA, 1980, 103 (02) :219-228
[4]  
EHRMEYER SS, 1995, AM J CLIN PATHOL, V104, pS40
[5]  
Figueiredo F, 2009, REVSTAT-STAT J, V7, P151
[6]  
Greene William H., 2012, ECONOMETRIC ANAL, DOI DOI 10.1016/J.JCORPFIN.2006.05.002
[7]   THE USE OF CONTROL CHARTS IN THE CLINICAL LABORATORY [J].
LEVEY, S ;
JENNINGS, ER .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1950, 20 (11) :1059-1066
[8]  
Linnet K, 2000, CLIN CHEM, V46, P867
[9]  
Lott JA, 1996, CLIN CHEM, V42, P888
[10]  
MARILYN H, 2002, STAT PROCESS CONTROL