Behaviour of the phakic iris-claw intraocular lens (Artisan®/Verisyse®) during accommodation:: An optical coherence biometry study

被引:10
作者
Sekundo, W. [1 ]
Bissmann, W. [2 ]
Tietjen, A. [3 ]
机构
[1] Johannes Gutenberg Univ Mainz, Dept Ophthalmol, Mainz, Germany
[2] Carl Zeiss Med AG, Jena, Germany
[3] Deutsch Diagnost Ctr, Ophthalmol Unit, Wiesbaden, Germany
关键词
accomodation; intraocular; lens; movement;
D O I
10.1177/112067210701700606
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To evaluate variations in the position of the phakic iris claw lens (Artisan((R)), Verisyse((R))) and the natural crystalline lens during the accommodation process. METHODS. We measured changes in position of the iris claw lens and the crystalline lens during the accommodation using optical coherence biometry (AC Master((R))/Carl Zeiss Meditec, Germany) in 17 patients (28 eyes) with a phakic iris claw lens implanted for high myopia and/or myopic astigmatism. Accommodative effort was obtained using an internal optical target within the measuring device. RESULTS. There was a forward shift of the phakic iris claw lens with a mean of 70 mu m (8-178 mu m) of the optical path length (OPL). At the same time the anterior pole of the natural lens showed a forward mean movement of 85 mu m (4-260 mu m). CONCLUSIONS. An anterior displacement of the iris-claw phakic lens was shown in a series of eyes during the accommodation process. As this displacement goes along with the forward displacement of the anterior pole of the crystalline lens, the preoperative measurement of the latter might provide some additional information about the position of the iris claw lens in the accommodative state after implantation.
引用
收藏
页码:904 / 908
页数:5
相关论文
共 12 条
[1]  
Augustin AJ, 2007, AUGENHEILKUNDE, P823
[2]   Pigment dispersion and Artisan phakic intraocular lenses -: Crystalline lens rise as a safety criterion [J].
Baïkoff, G ;
Bourgeon, G ;
Jodai, HJ ;
Fontaine, A ;
Lellis, FV ;
Trinquet, L .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04) :674-680
[3]   Static and dynamic analysis of the anterior segment with optical coherence tomography [J].
Baikoff, G ;
Lutun, E ;
Ferraz, C .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (09) :1843-1850
[4]   Position of angle-supported, iris-fixated, and ciliary sulcus-implanted myopic phakic Intraocular lenses evaluated by scheimpflug photography [J].
Baumeister, M ;
Bühren, J ;
Kohnen, T .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 138 (05) :723-731
[5]   Multicenter study of the Artisan phakic intraocular lens [J].
Budo, C ;
Hessloehl, JC ;
Izak, M ;
Luyten, GPM ;
Menezo, JL ;
Sener, BA ;
Tassignon, MJ ;
Termote, H ;
Worst, JGF .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (08) :1163-1171
[6]  
Fechner P U, 1991, Refract Corneal Surg, V7, P286
[7]   Intermittent myopic shift of 4.0 diopters after implantation of an Artisan iris-supported phakic intraocular lens [J].
Kohnen, T ;
Cichocki, M ;
Bühren, J ;
Baumeister, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (07) :1444-1447
[8]   NEGATIVE IMPLANT - A RETROSPECTIVE STUDY [J].
LANDESZ, M ;
WORST, JGF ;
SIERTSEMA, JV ;
VANRIJ, G .
DOCUMENTA OPHTHALMOLOGICA, 1993, 83 (04) :261-270
[9]   Endothelial study of iris-claw phakic lens: Four year follow-up [J].
Menezo, JL ;
Cisneros, AL ;
Rodriguez-Salvador, V .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (08) :1039-1049
[10]  
Sekundo W, 2004, OPHTHALMOLOGE, V101, P246, DOI 10.1007/s00347-003-0970-4