Clinical experience of photon counting breast tomosynthesis: comparison with traditional mammography

被引:43
作者
Svane, Gunilla [1 ,2 ]
Azavedo, Edward [1 ,2 ]
Lindman, Karin [3 ]
Urech, Mattias [3 ]
Nilsson, Jonas [3 ]
Weber, Niclas [3 ]
Lindqvist, Lars [3 ]
Ullberg, Christer [3 ]
机构
[1] Karolinska Univ Hosp, Dept Radiol, SE-17176 Stockholm, Sweden
[2] Karolinska Inst, Dept Mol Med & Surg, SE-17176 Stockholm, Sweden
[3] XCounter AB, SE-18233 Danderyd, Sweden
关键词
Tomosynthesis; photon counting; digital mammography; breast imaging; mammography screening; breast cancer; DIGITAL MAMMOGRAPHY; POPULATION; ACCURACY; WOMEN; ROC;
D O I
10.1258/ar.2010.100262
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: In two-dimensional mammography, a well-known problem is over- and underlying tissue which can either obstruct a lesion or create a false-positive result. Tomosynthesis, with an ability to layer the tissue in the image, has the potential to resolve these issues. Purpose: To compare the diagnostic quality, sensitivity and specificity of a single tomosynthesis mammography image and a traditional two-view set of two-dimensional mammograms and to assess the comfort of the two techniques. Material and Methods: One hundred and forty-four women, mainly chosen because of suspicious features on standard mammograms (76 malignant), had a single tomosynthesis image taken of one breast using a novel photon counting system. On average, the dose of the tomosynthesis images was 0.63 times that of the two-view images and the compression force during the procedure was halved. The resulting images were viewed by two radiologists and assessed both individually and comparing the two techniques. Results: In 56% of the cases the radiologists rated the diagnostic quality of the lesion details higher in the tomosynthesis images than in the conventional images (and in 91 % equal or higher), which means there is a statistically significant preference for the tomosynthesis technique. This included the calcifications which were rated as having better quality in 41 A) of the cases. While sensitivity was slightly higher for traditional mammography the specificity was higher for tomosynthesis. However, neither of these two differences was large enough to be statistically significant. Conclusion: The overall accuracy of the two techniques was virtually equal despite the radiologist's very limited experience with tomosynthesis images and vast experience with two-dimensional mammography. As the diagnostic quality of the lesion details in the tomosynthesis images was valued considerably higher this factor should improve with experience. The patients also favored the tomosynthesis examination, rating the comfort of the procedure as much higher than regular mammography which might affect screening attendance.
引用
收藏
页码:134 / 142
页数:9
相关论文
共 27 条
  • [11] Power estimation for the Dorfman-Berbaum-Metz method
    Hillis, SL
    Berbaum, KS
    [J]. ACADEMIC RADIOLOGY, 2004, 11 (11) : 1260 - 1273
  • [12] Recent developments in the Dorfman-Berbaum-Metz procedure for multireader ROC study analysis
    Hillis, Stephen L.
    Berbaum, Kevin S.
    Metz, Charles E.
    [J]. ACADEMIC RADIOLOGY, 2008, 15 (05) : 647 - 661
  • [13] A comparison of denominator degrees of freedom methods for multiple observer ROC analysis
    Hillis, Stephen L.
    [J]. STATISTICS IN MEDICINE, 2007, 26 (03) : 596 - 619
  • [14] The Norwegian Breast Cancer Screening Program: re-attendance related to the women's experiences, intentions and previous screening result
    Hofvind, SSH
    Wang, H
    Thoresen, S
    [J]. CANCER CAUSES & CONTROL, 2003, 14 (04) : 391 - 398
  • [15] MAIDMENT ADA, 2010, DIGITAL MAMMOGRAPHY, P144
  • [16] Evaluation of a photon-counting breast tomosynthesis imaging system
    Maidment, Andrew D. A.
    Ullberg, Christer
    Lindman, Karin
    Adelow, Leif
    Egerstrom, Johan
    Eklund, Mathias
    Francke, Tom
    Jordung, Ulf
    Kristoffersson, Tomas
    Lindqvist, Lars
    Marchal, Daniel
    Olla, Hans
    Penton, Erik
    Rantanen, Juha
    Solokov, Skiff
    Weber, Niclas
    Westerberg, Hans
    [J]. Medical Imaging 2006: Physics of Medical Imaging, Pts 1-3, 2006, 6142 : B1420 - B1420
  • [17] Proper binormal ROC curves: Theory and maximum-likelihood estimation
    Metz, CE
    Pan, XC
    [J]. JOURNAL OF MATHEMATICAL PSYCHOLOGY, 1999, 43 (01) : 1 - 33
  • [18] Digital Tomosynthesis in breast imaging
    Niklason, LT
    Christian, BT
    Niklason, LE
    Kopans, DB
    Castleberry, DE
    OpsahlOng, BH
    Landberg, CE
    Slanetz, PJ
    Giardino, AA
    Moore, R
    Albagli, D
    DeJule, MC
    Fitzgerald, PF
    Fobare, DF
    Giambattista, BW
    Kwasnick, RF
    Liu, JQ
    Lubowski, SJ
    Possin, GE
    Richotte, JF
    Wei, CY
    Wirth, RF
    [J]. RADIOLOGY, 1997, 205 (02) : 399 - 406
  • [19] Pain and discomfort associated with mammography among urban low-income African-American women
    Papas, MA
    Klassen, AC
    [J]. JOURNAL OF COMMUNITY HEALTH, 2005, 30 (04) : 253 - 267
  • [20] Digital breast tomosynthesis: Initial experience in 98 women with abnormal digital screening mammography
    Poplack, Steven P.
    Tosteson, Tor D.
    Kogel, Christine A.
    Nagy, Helene M.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (03) : 616 - 623