Clinical evaluation of Synthetic Aperture Sequential Beamforming and Tissue Harmonic Imaging

被引:0
作者
Brandt, Andreas Hjelm [1 ,2 ]
Hemmsen, Martin Christian [2 ]
Hansen, Peter Moller [1 ]
Hansen, Kristoffer Lindskov [1 ]
Lange, Theis [3 ]
Jensen, Jorgen Arendt [2 ]
Nielsen, Michael Bachmann [1 ]
机构
[1] Copenhagen Univ Hosp, Dept Radiol, Copenhagen, Denmark
[2] Tech Univ Denmark, Ctr Fast Ultrasound Imaging, Lyngby, Denmark
[3] Univ Copenhagen, Sect Biostat, DK-1168 Copenhagen, Denmark
来源
2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS) | 2014年
关键词
Imagine quality evaluation; Synthetic Aperture Sequential Beamforming Tissue Harmonic Imaging; Comparison of techniques; IN-VIVO EVALUATION; ULTRASOUND;
D O I
10.1109/ULTSYM.2014.0324
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This study determines if the data reduction achieved by the combination Synthetic Aperture Sequential Beamforming (SASB) and Tissue Harmonic Imaging (THI) affects image quality. SASB-THI was evaluated against the combination of Dynamic Received Focusing and Tissue Harmonic Imaging (DRF-THI). A BK medical UltraView 800 ultrasound scanner equipped with a research interface and an abdominal 3.5 MHz 3.5CL192-3ML convex array transducer was used and connected to a stand alone PC. SASB-THI and DRF-THI scan sequences were recorded interleaved and processed offline. Nineteen patients diagnosed with focal liver pathology were scanned to set a clinical condition, where ultrasonography is often performed. A total of 114 sequences were recorded and evaluated by five radiologists. The evaluators were blinded to the imaging technique, and each sequence was shown twice with different left-right positioning, resulting in 1140 evaluations. The program Image Quality Assessment Program (IQap) and a Visual Analog Scale (VAS) were applied for the evaluation. The scale ranged from -50 to 50, where positive values favored SASB-THI. SASB-THI and DRF-THI were evaluated alike in 49% of the evaluations, 28% favored SASB-THI and 23% favored DRF-THI. The average rating was 0.70 (Cl: -0.80 to 2.19). The statistical analysis, where the hypothesis of no differences between the techniques was tested, yielded a p-value of p=0.64, indicating no preference to any technique. This study demonstrates that SASB-THI and DRF-THI have equally good image quality although a data reduction of 64 times is achieved with SASB-THI.
引用
收藏
页码:1312 / 1315
页数:4
相关论文
共 17 条
[1]   Second harmonic imaging using synthetic aperture sequential beamforming [J].
Du, Yigang ;
Rasmussen, Joachim ;
Jensen, Henrik ;
Jensen, Jorgen Arendt .
2011 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2011, :2261-2264
[2]   The graphic rating scale [J].
Freyd, M .
JOURNAL OF EDUCATIONAL PSYCHOLOGY, 1923, 14 :83-102
[3]  
Hansen P.M., 2014, CLIN EVALUA IN PRESS
[4]  
Hemmsen M.C., 2010, P SPIE MED IMAGING, V7629
[5]   Implementation of a Versatile Research Data Acquisition System Using a Commercially Available Medical Ultrasound Scanner [J].
Hemmsen, Martin Christian ;
Nikolov, Svetoslav Ivanov ;
Pedersen, Mads Moller ;
Pihl, Michael Johannes ;
Enevoldsen, Marie Sand ;
Hansen, Jens Munk ;
Jensen, Jorgen Arendt .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2012, 59 (07) :1487-1499
[6]   IN VIVO EVALUATION OF SYNTHETIC APERTURE SEQUENTIAL BEAMFORMING [J].
Hemmsen, Martin Christian ;
Hansen, Peter Moller ;
Lange, Theis ;
Hansen, Jens Munk ;
Hansen, Kristoffer Lindskov ;
Nielsen, Michael Bachmann ;
Jensen, Jorgen Arendt .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2012, 38 (04) :708-716
[7]   Ultrasonographic evaluation of focal hepatic lesions: Comparison of pulse inversion harmonic, tissue harmonic, and conventional imaging techniques [J].
Jang, HJ ;
Lim, HK ;
Lee, WJ ;
Kim, SH ;
Kim, KA ;
Kim, EY .
JOURNAL OF ULTRASOUND IN MEDICINE, 2000, 19 (05) :293-299
[8]   Synthetic aperture ultrasound imaging [J].
Jensen, Jorgen Arendt ;
Nikolov, Svetoslav Ivanov ;
Gammelmark, Kim Lokke ;
Pedersen, Morten Hogholm .
ULTRASONICS, 2006, 44 :E5-E15
[9]   Evaluation of Ultrasound Synthetic Aperture Imaging Using Bidirectional Pixel-Based Focusing: Preliminary Phantom and In Vivo Breast Study [J].
Kim, Choye ;
Yoon, Changhan ;
Park, Jong-Ho ;
Lee, Yuhwa ;
Kim, Won Hwa ;
Chang, Jung Min ;
Choi, Byung Ihn ;
Song, Tai-Kyong ;
Yoo, Yang-Mo .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (10) :2716-2724
[10]  
Kim K.-S., 2000, KEY ENG MATER, V47, P1510