MR imaging of vascular stents: effects of susceptibility, flow, and radiofrequency eddy currents

被引:100
作者
Bartels, LW [1 ]
Smits, HFM [1 ]
Bakker, CJG [1 ]
Viergever, MA [1 ]
机构
[1] Univ Utrecht, AZU, Med Ctr, Dept Radiol,Image Sci Inst, NL-3584 CX Utrecht, Netherlands
关键词
eddy currents; flow; artifacts; MR; vascular studies; stents and prostheses;
D O I
10.1016/S1051-0443(07)61918-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: The purpose of this in vitro study was to examine the various sources of artifacts in magnetic resonance (MR) imaging and angiography of vascular stents. MATERIALS AND METHODS: Five low-artifact stents-Wallstent (cobalt alloy), Memotherm (nitinol), Perflex (stainless steel), Passager (tantalum), and Smart (nitinol)-were imaged in a vascular flow phantom, consisting of a thin-walled cellulose vessel model connected to a pump system. The echo time and the angulation of the stents with respect to the direction of the main magnetic field were varied. Spin echo and gradient echo images as well as three-dimensional MR angiograms were obtained to study the effects of flow magnetic susceptibility, and radiofrequency-induced eddy currents. RESULTS: Susceptibility artifacts were restricted to the stents' direct environment and were mildest at short echo times and with the stents aligned with the main magnetic field. Nitinol stents showed less artifacts than steel stents did. Radiofrequency artifacts obscuring the stent lumen and flow-related lumen displacement were seen in all stents. The extent to which these occurred depended on strut geometry and orientation. CONCLUSIONS: For low-artifact stents, the material the stent is made of is not the only important factor in the process of artifact formation. Susceptibility artifacts, radiofrequency eddy currents and flow-related artifacts all contribute to the image distortion, and are dependent on the geometry and orientation of the struts and on the orientation of the stent in the main magnetic field.
引用
收藏
页码:365 / 371
页数:7
相关论文
共 19 条
  • [1] Metallic artifacts of coronary and iliac arteries stents in mr angiography and contrast-enhanced CT
    Amano, Y
    Ishihara, M
    Hayashi, H
    Gemma, K
    Kawamata, H
    Amano, M
    Kumazaki, T
    [J]. CLINICAL IMAGING, 1999, 23 (02) : 85 - 89
  • [2] SIMULATION OF SUSCEPTIBILITY ARTIFACTS IN 2D AND 3D FOURIER-TRANSFORM SPIN-ECHO AND GRADIENT-ECHO MAGNETIC-RESONANCE-IMAGING
    BAKKER, CJG
    BHAGWANDIEN, R
    MOERLAND, MA
    RAMOS, LMP
    [J]. MAGNETIC RESONANCE IMAGING, 1994, 12 (05) : 767 - 774
  • [3] SUSCEPTIBILITY ARTIFACTS IN 2DFT SPIN-ECHO AND GRADIENT-ECHO IMAGING - THE CYLINDER MODEL REVISITED
    BAKKER, CJG
    BHAGWANDIEN, R
    MOERLAND, MA
    FUDERER, M
    [J]. MAGNETIC RESONANCE IMAGING, 1993, 11 (04) : 539 - 548
  • [4] ANALYSIS OF MACHINE-DEPENDENT AND OBJECT-INDUCED GEOMETRIC DISTORTION IN 2DFT MR IMAGING
    BAKKER, CJG
    MOERLAND, MA
    BHAGWANDIEN, R
    BEERSMA, R
    [J]. MAGNETIC RESONANCE IMAGING, 1992, 10 (04) : 597 - 608
  • [5] HEATING EFFECTS OF METALLIC IMPLANTS BY MRI EXAMINATIONS
    BUCHLI, R
    BOESIGER, P
    MEIER, D
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (03) : 255 - 261
  • [6] NONSUSCEPTIBILITY ARTIFACTS DUE TO METALLIC OBJECTS IN MR-IMAGING
    CAMACHO, CR
    PLEWES, DB
    HENKELMAN, RM
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (01): : 75 - 88
  • [7] Plain and covered stent-grafts: In vitro evaluation of characteristics at three-dimensional MR angiography
    Hilfiker, PR
    Quick, HH
    Debatin, JF
    [J]. RADIOLOGY, 1999, 211 (03) : 693 - 697
  • [8] MAGNETIC-RESONANCE ANGIOGRAPHY OF INTRAVASCULAR ENDOPROSTHESES - INVESTIGATION OF 3 DEVICES
    LAISSY, JP
    GRAND, C
    MATOS, C
    STRUYVEN, J
    BERGER, JF
    SCHOUMANCLAEYS, E
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1995, 18 (06) : 360 - 366
  • [9] Iliofemoral arterial occlusive disease: Contrast-enhanced MR angiography for preinterventional evaluation and follow-up after stent placement
    Link, J
    Steffens, JC
    Brossman, J
    Graessner, J
    Hackethal, S
    Heller, M
    [J]. RADIOLOGY, 1999, 212 (02) : 371 - 377
  • [10] Ludeke K M, 1985, Magn Reson Imaging, V3, P329, DOI 10.1016/0730-725X(85)90397-2