Improving accuracy of phakic intraocular lens sizing using high-frequency ultrasound biomicroscopy

被引:67
作者
Dougherty, Paul J. [1 ]
Rivera, Robert P. [2 ]
Schneider, David [3 ]
Lane, Stephen S. [4 ]
Brown, David [5 ]
Vukich, John [6 ]
机构
[1] Dougherty Laser Vis, Camarillo, CA 93010 USA
[2] Barnet Dulaney Perkins Eye Ctr, Phoenix, AZ USA
[3] Midw Eye Ctr, Cincinnati, OH USA
[4] Associated Eye Care, Stillwater, MN USA
[5] Eye Ctr Florida, Ft Myers, FL USA
[6] Davis Duehr Dean Med Ctr, Madison, WI USA
关键词
IMPLANTABLE CONTACT-LENS;
D O I
10.1016/j.jcrs.2010.07.014
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess the efficacy of high-frequency ultrasound biomicroscopy (UBM) in improving the accuracy of phakic intraocular lens (pIOL) sizing by increasing the incidence of acceptable postoperative vault. SETTING: Multiple private practices, United States. DESIGN: Evaluation of diagnostic test or technology. METHODS: This prospective multicenter clinical study evaluated eyes having pIOL (Visian Implantable Col lamer Lens) implantation. A retrospective data analysis was performed using UBM measurements (VuMax-II) of preoperative sulcus-to-sulcus (STS) distance and postoperative vault. The regression data and clinical input from investigators were then used to develop a pIOL sizing nomogram. The nomogram used only STS and pIOL power as variables to determine length. Inadequate vault (<90 mu m) and excessive vault (>1000 mu m) were defined based on peer-reviewed literature. Sizing recommendations using the nomogram were studied prospectively and compared with 2 sizing methods used in the United States that are based on white-to-white (WTW) measurements. RESULTS: One eye was excluded from the analysis because the wrong length pIOL was placed (12.6 mm instead of nomogram-recommended 13.2 mm), resulting in 51 mu m of vault. The mean postoperative vault in the remaining 72 cases was 340 mu m +/- 174 (SD) (range 90 to 952 pm); there were no cases of inadequate or excessive vault with the newly developed UBM nomogram. Sizing methods using WTW measurements would have resulted in different-sized pIOLs in 36% to 69% of cases compared with the STS method. CONCLUSION: There were no cases of inadequate or excessive vault when the UBM nomogram for pIOL sizing was used.
引用
收藏
页码:13 / 18
页数:6
相关论文
共 10 条
  • [1] Choi KH, 2007, J REFRACT SURG, V23, P362
  • [2] THE LENS OPACITIES CLASSIFICATION SYSTEM-III
    CHYLACK, LT
    WOLFE, JK
    SINGER, DM
    LESKE, MC
    BULLIMORE, MA
    BAILEY, IL
    FRIEND, J
    MCCARTHY, D
    WU, SY
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (06) : 831 - 836
  • [3] DOUGHERTY PJ, 2007, ANN M AM AC OPHTH NE
  • [4] Implantable contact lens for moderate to high myopia - Relationship of vaulting to cataract formation
    Gonvers, M
    Bornet, C
    Othenin-Girard, P
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (05) : 918 - 924
  • [5] Hernandez-Matamoros JL, 2007, ASCRS S CAT IOL REFR
  • [6] LOVISOLO CF, 2007, ASCRS S CAT IOL REFR
  • [7] Phakic intraocular lenses
    Lovisolo, CR
    Reinstein, DZ
    [J]. SURVEY OF OPHTHALMOLOGY, 2005, 50 (06) : 549 - 587
  • [8] *MOR RES CONS INC, PREVISE
  • [9] Correlation of Anterior Chamber Angle and Ciliary Sulcus Diameters With White-to-White Corneal Diameter in High Myopes Using Artemis VHF Digital Ultrasound
    Reinstein, Dan Z.
    Archer, Timothy J.
    Silverman, Ronald H.
    Rondeau, Mark J.
    Coleman, D. Jackson
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2009, 25 (02) : 185 - 194
  • [10] *STAAR SURG, VIS ICL PROD INF