Estimation of intraocular pressure in rabbits using noncontact tonometer: A comparative evaluation with Schiotz tonometer

被引:8
作者
Gupta, S. K.
Saxena, R.
Agarwal, R.
Galpalli, N. D.
Srivastava, S.
Agrawal, S. S.
机构
[1] Delhi Inst Pharmaceut Sci & Res, Inst Clin Res India, Dept Pharmacol, New Delhi, India
[2] All India Inst Med Sci, Dr Rajendra Prasad Ctr Ophthalm Sci, New Delhi 110029, India
[3] Delhi Inst Pharmaceut Sci & Res, Dept Pharmacol, New Delhi, India
来源
METHODS AND FINDINGS IN EXPERIMENTAL AND CLINICAL PHARMACOLOGY | 2007年 / 29卷 / 06期
关键词
intraocular pressure; noncontact tonometer; rabbits; Schiotz tonometer;
D O I
10.1358/mf.2007.29.6.1119161
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Estimation of intraocular pressure in rabbits is often done by indentation tonometry using Schiotz tonometer. The use of Schiotz tonometer for estimation of intraocular pressure requires the use of topical anesthesia, poses risk of infection and injury to cornea and puts the animal under stress. Therefore, in this study noncontact tonometer was used to estimate intraocular pressure in rabbits and the technique is described for the first time. Twenty-five rabbits were subjected to intraocular pressure estimation using Schiotz as well as noncontact tonometer by two independent observers. All intraobserver and interobserver differences were within two standard deviation from the mean, indicating good repeatability and reproducibility of the two methods. A good linear correlation was observed between measurements done by Schiotz and noncontact tonometers with the correlation coefficient (r(2)) of 0.9122 and regression equation y = 1.632x + 4.8684. No significant difference was observed in percent change in intraocular pressure in timolol-treated, water-loaded rabbits using noncontact tonometer and Schiotz tonometer. Noncontact tonometer provides an easy, quick and safe method for intraocular pressure estimation, which is comparable to the conventional method of Schiotz tonometry. Therefore, despite the cost, noncontact tonometry seems to have a greater scope in experimental studies. (c) 2007 Prous Science. All rights resmed.
引用
收藏
页码:405 / 409
页数:5
相关论文
共 11 条
  • [1] EVALUATION OF KEELER PULSAIR NONCONTACT TONOMETRY - RELIABILITY AND REPRODUCIBILITY
    BONOMI, L
    BARAVELLI, S
    COBBE, C
    TOMAZZOLI, L
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1991, 229 (03) : 210 - 212
  • [2] Goyal J K, 1989, Indian J Ophthalmol, V37, P176
  • [3] Jani A., 2005, IRANIAN J PHARM THER, V4, P95
  • [4] Prospective comparison of the new indentation tonometer TGdC-01, the non-contact tonometer PT100 and the conventional Goldmann applanation tonometer
    Müller, A
    Godenschweger, L
    Lang, GE
    Kampmeier, J
    [J]. KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2004, 221 (09) : 762 - 768
  • [5] Patikulsila Direk, 2003, J Med Assoc Thai, V86, P467
  • [6] Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study
    Regine, F
    Scuderi, GL
    Cesareo, M
    Ricci, F
    Cedrone, C
    Nucci, C
    [J]. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2006, 26 (01) : 33 - 39
  • [7] Oculohypotensive effect of angiotensin-converting enzyme inhibitors in acute and chronic models of glaucoma
    Shah, GB
    Sharma, S
    Mehta, AA
    Goyal, RK
    [J]. JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 2000, 36 (02) : 169 - 175
  • [8] NON-CONTACT TONOMETER - ITS VALUE AND LIMITATIONS
    SHIELDS, MB
    [J]. SURVEY OF OPHTHALMOLOGY, 1980, 24 (04) : 211 - 219
  • [9] Methylcellulose, a healing inhibitor factor in an animal model of trabeculectomy
    Shouman, Ayman A. M.
    Helal, Ahmed
    Marzouk, Mohamed A.
    Zaki, Eman M. A.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2006, 47 (06) : 2515 - 2519
  • [10] VEMON SA, 1985, BR J OPHTHALMOL, V73, P247