Design of a Novel Bimanual Robotic System for Single-Port Laparoscopy

被引:139
作者
Piccigallo, Marco [1 ]
Scarfogliero, Umberto [1 ]
Quaglia, Claudio [1 ]
Petroni, Gianluigi [1 ]
Valdastri, Pietro [1 ]
Menciassi, Arianna [1 ]
Dario, Paolo [1 ]
机构
[1] Scuola Super Sant Anna, CRIM Lab, I-56025 Pisa, Italy
关键词
Bimanual robot; miniature robotic arm; minimally invasive surgery; robotic surgery; single-port laparoscopy (SPL); ENDOSCOPIC SURGERY;
D O I
10.1109/TMECH.2010.2078512
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper presents the design and fabrication of Single-Port lapaRoscopy bImaNual roboT (SPRINT), a novel tele-operated robotic system for minimally invasive surgery. SPRINT, specifically designed for single-port laparoscopy, is a high-dexterity miniature robot, able to reproduce the movement of the hands of the surgeon, who controls the system through a master interface. It comprises two arms with six degrees of freedom (DOFs) that can be individually inserted and removed in a 30-mm-diameter umbilical access port. The system is designed to leave a central lumen free during operations, thus allowing the insertion of other laparoscopic tools. The four distal DOFs of each arm are actuated by on-board brushless dc motors, while the two proximal DOFs of the shoulder are actuated by external motors. The constraints generated by maximum size and power requirements led to the design of compact mechanisms for the actuation of the joints. The wrist is actuated by three motors hosted in the forearm, with a peculiar differential mechanism that allows us to have intersecting roll-pitch-roll axes. Preliminary tests and validations were performed ex vivo by surgeons on a first prototype of the system.
引用
收藏
页码:871 / 878
页数:8
相关论文
共 27 条
  • [21] Magnetic retraction in natural-orifice transluminal endoscopic surgery (NOTES): addressing the problem of traction and countertraction
    Ryou, M.
    Thompson, C. C.
    [J]. ENDOSCOPY, 2009, 41 (02) : 143 - 148
  • [22] Spaun GO, 2008, EUR SURG, V40, P211, DOI 10.1007/s10353-008-0428-7
  • [23] STASSEN HG, 2004, ENG PATIENT SAFETY I, P2
  • [24] Robotic single-site surgery
    White, Michael A.
    Haber, Georges-Pascal
    Kaouk, Jihad H.
    [J]. CURRENT OPINION IN UROLOGY, 2010, 20 (01) : 86 - 91
  • [25] Single-port Urological Surgery: Single-center Experience With the First 100 Cases
    White, Wesley M.
    Haber, Georges-Pascal
    Goel, Raj K.
    Crouzet, Sebastien
    Stein, Robert J.
    Kaouk, Jihad H.
    [J]. UROLOGY, 2009, 74 (04) : 801 - 804
  • [26] XU K, P 2009 IEEE RSJ INT, P5546
  • [27] Mechatronic design of a new robot for force control in minimally invasive surgery
    Zemiti, Nabil
    Morel, Guillaume
    Ortmaier, Tobias
    Bonnet, Nicolas