Comparison of Daily Prostate Positions during Conformal Radiation Therapy of Prostate Cancer Using an Integrated CT-linear Accelerator System: In-room CT Image versus Digitally Reconstructed Radiograph

被引:4
作者
Aoyama, Hideki [1 ,2 ]
Azuma, Yoshiharu [2 ]
Inamura, Keiji [1 ]
机构
[1] Okayama Univ Hosp, Cent Div Radiol, Okayama 7008558, Japan
[2] Okayama Univ, Grad Sch Hlth Sci, Okayama 7008558, Japan
关键词
BONY ANATOMY; GUIDED RADIOTHERAPY; SETUP ERRORS; ORGAN MOTION; VERIFICATION; IRRADIATION; LOCALIZATION; UNCERTAINTY; VARIABILITY; REDUCTION;
D O I
10.1269/jrr.10111
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Prostate position/Prostate cancer/CT-linear accelerator system/In-room CT image/Digitally reconstructed radiograph. Purpose: The purpose of this study is to quantify the magnitudes of the position shifts of internal structures together with the correlation between the day-to-day positioning of the prostate and the bony anatomy using an integrated CT-linear accelerator system for external beam radiation therapy for prostate cancer. Materials and methods: A total of 1176 pretreatment in-room CT images and their digitally reconstructed radio-graph (DRR) pairs from 33 patients were acquired over the course of the study. The differences between the isocenter of the prostate on in-room CT and the isocenter of the bony anatomy on DRR were analyzed. The agreement between positions in each direction was compared using Bland-Altman limits of agreement. Results: The 95% limits of agreement in lateral (LR), superoinferior (SI), and anteroposterior (AP) directions were -2.98 to 2.49 mm, -4.69 to 5.75 mm, and -8.23 to 7.30 mm, respectively. The isocenter was localized to within 3.0 mm on in-room CT images and DRR 99.0% in LR, 85.1% in SI, and 85.9% in AP. Conclusions: Considerable differences between in-room CT images and DRR exist. These data demonstrate that there is a significantly greater shift in the SI and AP directions than in the lateral direction for the entire patient group. Applications such as our image guide system will, with routine clinical use, continue to improve the precision of external beam radiation therapy for prostate cancer.
引用
收藏
页码:220 / 228
页数:9
相关论文
共 26 条
[1]   Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate [J].
Alasti, H ;
Petric, MP ;
Catton, CN ;
Warde, PR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03) :869-884
[2]   Variation in prostate position relative to adjacent bony anatomy [J].
Althof, VGM ;
Hoekstra, CJM ;
Loo, HJT .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (03) :709-715
[3]   Analysis of prostate and seminal vesicle motion: Implications for treatment planning [J].
Beard, CJ ;
Kijewski, P ;
Bussiere, M ;
Gelman, R ;
Gladstone, D ;
Shaffer, K ;
Plunkett, M ;
Costello, P ;
Coleman, CN .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (02) :451-458
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Modeling set-up error by daily MVCT for prostate adjuvant treatment delivered in 20 fractions: Implications for the assessment of the optimal correction strategies [J].
Broggi, Sara ;
Cozzarini, Cesare ;
Fiorino, Claudio ;
Maggiulli, Eleonora ;
Alongi, Filippo ;
Cattaneo, Giovanni Mauro ;
Di Muzio, Nadia ;
Calandrino, Riccardo .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) :246-252
[6]   On-line aSi portal imaging of implanted fiducial markers for the reduction of interfraction error during conformal radiotherapy of prostate carcinoma [J].
Chung, PWM ;
Haycocks, T ;
Brown, T ;
Cambridge, Z ;
Kelly, V ;
Alasti, H ;
Jaffray, DA ;
Catton, CN .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01) :329-334
[7]   A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload [J].
De Boer, HCJ ;
Heijmen, BJM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (05) :1350-1365
[8]   Comparison of prostate set-up accuracy and margins with off-line bony anatomy corrections and online implanted fiducial-based corrections [J].
Greer, P. B. ;
Dahl, K. ;
Ebert, M. A. ;
Wratten, C. ;
White, M. ;
Denham, J. W. .
JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2008, 52 (05) :511-516
[9]   Measurement of patient positioning errors in three-dimensional conformal radiotherapy of the prostate [J].
Hanley, J ;
Lumley, MA ;
Mageras, GS ;
Sun, J ;
Zelefsky, MJ ;
Leibel, SA ;
Fuks, Z ;
Kutcher, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (02) :435-444
[10]   Development of a semi-automatic alignment tool for accelerated localization of the prostate [J].
Hua, CH ;
Lovelock, DM ;
Mageras, GS ;
Katz, MS ;
Mechalakos, J ;
Lief, EP ;
Hollister, T ;
Lutz, WR ;
Zelefsky, MJ ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (03) :811-824