Determination of Coefficient of Friction for Self-Expanding Stent-Grafts

被引:53
作者
Vad, Siddharth
Eskinazi, Amanda [3 ]
Corbett, Timothy [4 ,5 ]
McGloughlin, Tim [4 ,5 ]
Vande Geest, Jonathan P. [1 ,2 ]
机构
[1] Univ Arizona, Dept Biomed Engn, Dept Aerosp & Mech Engn, Tucson, AZ 85721 USA
[2] Univ Arizona, Inst Collaborat Res BIO5, Tucson, AZ 85721 USA
[3] Univ Arizona, Dept Agr & Biosyst Engn, Tucson, AZ 85721 USA
[4] Univ Limerick, CABER, MSSi, Limerick, Ireland
[5] Univ Limerick, Dept Aerosp & Mech Engn, Limerick, Ireland
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2010年 / 132卷 / 12期
关键词
endovascular repair; self-expanding stent-grafts; finite element analysis; coefficient of friction; contact pressure; ABDOMINAL AORTIC-ANEURYSMS; ENDOVASCULAR REPAIR; IN-VIVO; PROXIMAL FIXATION; NITINOL STENT; MIGRATION; MODEL; PRESSURE; STRENGTH; PLACEMENT;
D O I
10.1115/1.4002798
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Migration of stent-grafts (SGs) after endovascular aneurysm repair of abdominal aortic aneurysms is a serious complication that may require secondary intervention. Experimental, analytical, and computational studies have been carried out in the past to understand the factors responsible for migration. In an experimental setting, it can be very challenging to correctly capture and understand the interaction between a SG and an artery. Quantities such as coefficient of friction (COF) and contact pressures that characterize this interaction are difficult to measure using an experimental approach. This behavior can be investigated with good accuracy using finite element modeling. Although finite element models are able to incorporate frictional behavior of SGs, the absence of reliable values of coefficient of friction make these simulations unreliable. The aim of this paper is to demonstrate a method for determining the coefficients of friction of a self-expanding endovascular stent-graft. The methodology is demonstrated by considering three commercially available self-expanding SGs, labeled as A, B, and C. The SGs were compressed, expanded, and pulled out of polymeric cylinders of varying diameters and the pullout force was recorded in each case. The SG geometries were recreated using computer-aided design modeling and the entire experiment was simulated in ABAQUS 6.8/STANDARD. An optimization procedure was carried out for each SG oversize configuration to determine the COF that generated a frictional force corresponding to that measured in the experiment. The experimental pullout force and analytically determined COF for SGs A, B, and C were in the range of 6-9 N, 3-12 N, and 3-9 N and 0.08-0.16, 0.22-0.46, and 0.012-0.018, respectively. The computational model predicted COFs in the range of 0.00025-0.0055, 0.025-0.07, and 0.00025-0.006 for SGs A, B, and C, respectively. Our results suggest that for SGs A and B, which are exoskeleton based devices, the pullout forces increase upto a particular oversize beyond which they plateau, while pullout forces showed a continuous increase with oversize for SG C, which is an endoskeleton based device. The COF decreased with oversizing for both types of SGs. The proposed methodology will be useful for determining the COF between self-expanding stent-grafts from pullout tests on human arterial tissue. [DOI: 10.1115/1.4002798]
引用
收藏
页数:10
相关论文
共 55 条
  • [1] ALICEA LA, 2004, MECH BIOMATERIALS ST, pF1
  • [2] Iliac fixation length and resistance to in-vivo stent-graft displacement
    Arko, FR
    Heikkinen, M
    Lee, ES
    Bass, A
    Alsac, JM
    Zarins, CK
    [J]. JOURNAL OF VASCULAR SURGERY, 2005, 41 (04) : 664 - 670
  • [3] *ASTM, 2008, ASTM STAND SPEC WROU
  • [4] Effects of stent design parameters on normal artery wall mechanics
    Bedoya, Julian
    Meyer, Clark A.
    Timmins, Lucas H.
    Moreno, Michael R.
    Moore, James E., Jr.
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2006, 128 (05): : 757 - 765
  • [5] The effect of injectable biocompatible elastomer (PDMS) on the strength of the proximal fixation of endovascular aneurysm repair grafts: An in vitro study
    Bosman, Willem-Maarten P. F.
    van der Steenhoven, Tim J.
    Suarez, Daniel R.
    Valstar, Edward R.
    de Vries, Alexander C.
    Brom, Hans L. F.
    Jacobs, Michael J.
    Hamming, Jaap F.
    [J]. JOURNAL OF VASCULAR SURGERY, 2010, 52 (01) : 152 - 158
  • [6] Budynas Richard G., 2008, SHIGLEY'S MECHANICAL ENGINEERING DESIGN
  • [7] Proximal fixation of thoracic stent-grafts as a function of oversizing and increasing aortic arch angulation in human cadaveric aortas
    Canaud, Ludovic
    Alric, Pierre
    Laurent, Martrille
    Baum, Thierry-Pascal
    Branchereau, Pascal
    Marty-Ane, Charles Henri
    Berthet, Jean-Phillipe
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2008, 15 (03) : 326 - 334
  • [8] Endovascular Stent grafting of thoracic aortic aneurysms
    Carroccio, A
    Ellozy, S
    Spielvogel, D
    Marin, ML
    Hollier, L
    [J]. ANNALS OF VASCULAR SURGERY, 2003, 17 (04) : 473 - 478
  • [9] Catanese J, 1999, J BIOMED MATER RES, V48, P187, DOI 10.1002/(SICI)1097-4636(1999)48:2<187::AID-JBM13>3.3.CO
  • [10] 2-D