Semi-quantitative and simulation analyses of effects of γ rays on determination of calibration factors of PET scanners with point-like 22Na sources

被引:8
作者
Hasegawa, Tomoyuki [1 ]
Sato, Yasushi [2 ]
Oda, Keiichi [3 ]
Wada, Yasuhiro [4 ]
Murayama, Hideo [5 ]
Yamada, Takahiro [6 ]
机构
[1] Kitasato Univ, Sch Allied Hlth Sci, Minami Ku, Kanagawa 2520373, Japan
[2] Natl Inst Adv Ind Sci & Technol, Tsukuba, Ibaraki 3058568, Japan
[3] Tokyo Metropolitan Inst Gerontol, Itabashi Ku, Tokyo 1730022, Japan
[4] RIKEN Ctr Mol Imaging Sci, Chuo Ku, Kobe, Hyogo 6500047, Japan
[5] Natl Inst Radiol Sci, Inage Ku, Chiba 2638555, Japan
[6] Japan Radioisotope Assoc, Bunkyo Ku, Tokyo 1138941, Japan
关键词
POSITRON-EMISSION-TOMOGRAPHY; IONIZATION-CHAMBER; PERFORMANCE; RECONSTRUCTION; LIMIT;
D O I
10.1088/0031-9155/56/18/016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The uncertainty of radioactivity concentrations measured with positron emission tomography (PET) scanners ultimately depends on the uncertainty of the calibration factors. A new practical calibration scheme using point-like Na-22 radioactive sources has been developed. The purpose of this study is to theoretically investigate the effects of the associated 1.275 MeV gamma rays on the calibration factors. The physical processes affecting the coincidence data were categorized in order to derive approximate semi-quantitative formulae. Assuming the design parameters of some typical commercial PET scanners, the effects of the gamma rays as relative deviations in the calibration factors were evaluated by semi-quantitative formulae and a Monte Carlo simulation. The relative deviations in the calibration factors were less than 4%, depending on the details of the PET scanners. The event losses due to rejecting multiple coincidence events of scattered gamma rays had the strongest effect. The results from the semi-quantitative formulae and the Monte Carlo simulation were consistent and were useful in understanding the underlying mechanisms. The deviations are considered small enough to correct on the basis of precise Monte Carlo simulation. This study thus offers an important theoretical basis for the validity of the calibration method using point-like Na-22 radioactive sources.
引用
收藏
页码:6031 / 6045
页数:15
相关论文
共 40 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]  
[Anonymous], 2008, NATL COMPREHENSIVE C, P1
[3]  
[Anonymous], 873597 NBSIR NIST
[4]  
Bailey DL, 1998, DEV NUC MED, V32, P55
[5]  
Bailey DL, 1997, EUR J NUCL MED, V24, P660
[6]   STUDY OF POSITRONIUM IN WATER AND ICE FROM 22 TO -144 DEGREES C BY ANNIHILATION QUANTA MEASUREMENTS [J].
COLOMBINO, P ;
FISCELLA, B ;
TROSSI, L .
NUOVO CIMENTO, 1965, 38 (02) :707-+
[7]   ON THE ANGULAR DISTRIBUTION OF 2-PHOTON ANNIHILATION RADIATION [J].
DEBENEDETTI, S ;
COWAN, CE ;
KONNEKER, WR ;
PRIMAKOFF, H .
PHYSICAL REVIEW, 1950, 77 (02) :205-212
[8]  
DERENZO SE, 1977, J NUCL MED, V18, P491
[9]  
EICHLING JO, 1977, J NUCL MED, V18, P845
[10]  
Firestone R.B., 2004, WWW TABLE RADIOACTIV