Radiation dose measurements during kilovoltage-cone beam computed tomography imaging in radiotherapy

被引:15
作者
Kumar, A. Sathish [1 ]
Singh, I. Rabi Raja [1 ]
Sharma, Sunil Dutt [2 ]
John, Subhashini [1 ]
Ravindran, B. Paul [1 ]
机构
[1] Christian Med Coll & Hosp, Dept Radiotherapy, Vellore, Tamil Nadu, India
[2] Bhabha Atom Res Ctr, Radiol Phys & Advisory Div, Bombay, Maharashtra, India
关键词
Cone beam computed tomography; image-guided radiotherapy; in vivo dose measurement; MOSFET; FLAT-PANEL IMAGER; GUIDED RADIOTHERAPY; PROSTATE-CANCER; CT; THERAPY; PRECISION; DOSIMETRY; MOSFET; CBCT; MV;
D O I
10.4103/0973-1482.164699
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Objective: The use of image guidance during radiotherapy for accurate localization and setup has become the standard care of practice in radiotherapy. This mostly involves the use of kilovoltage-cone beam computed tomography (kV-CBCT) for verification of patient setup on the first few days and on a weekly basis. Some protocols require this to be performed daily and also before and after the treatment. Though the radiation due to this kV-CBCT is small, the repeated use could deliver a dose that could increase the probability of the stochastic effect. The main purpose of this work is to measure radiation dose during image guidance with kV-CBCT. Materials and Methods: In this work, we have attempted to measure the dose during kV-CBCT for different sites both on a humanoid phantom and on patients undergoing image-guided radiotherapy with MOSFETs calibrated against an ion chamber. Results: The dose measurement on patients during kV-CBCT resulted in mean doses of 0.19 and 0.3 cGy to the ipsilateral and contralateral eyes, 0.625 and 1.097 cGy to the surface of the ipsilateral and contralateral breasts, and 3.01 cGy to the surface of the pelvis. Conclusion: Radiation dose to the eye, breast, and the surface of the pelvis have been arrived at during CBCT. The doses measured on patients agreed closely with those measured on humanoid phantom and with published values.
引用
收藏
页码:858 / 863
页数:6
相关论文
共 26 条
[1]   A Monte Carlo investigation of cumulative dose measurements for cone beam computed tomography (CBCT) dosimetry [J].
Abuhaimed, Abdullah ;
Martin, Colin J. ;
Sankaralingam, Marimuthu ;
Gentle, David J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (04) :1519-1542
[2]   An investigation of image guidance dose for breast radiotherapy [J].
Alvarado, Rosemerie ;
Booth, Jeremy T. ;
Bromley, Regina M. ;
Gustafsson, Helen B. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (03) :25-38
[3]   Imaging doses from the Elekta Synergy X-ray cone beam CT system [J].
Amer, A. ;
Marchant, T. ;
Sykes, J. ;
Czajka, J. ;
Moore, C. .
BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (954) :476-482
[4]   A new radiotherapy surface dose detector: The MOSFET [J].
Butson, MJ ;
Rozenfeld, A ;
Mathur, JN ;
Carolan, M ;
Wong, TPY ;
Metcalfe, PE .
MEDICAL PHYSICS, 1996, 23 (05) :655-658
[5]   KILOVOLTAGE IMAGING DOSES IN THE RADIOTHERAPY OF PEDIATRIC CANCER PATIENTS [J].
Deng, Jun ;
Chen, Zhe ;
Roberts, Kenneth B. ;
Nath, Ravinder .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (05) :1680-1688
[6]   Accurate patient dosimetry of kilovoltage cone-beam CT in radiation therapy [J].
Ding, George X. ;
Duggan, Dennis M. ;
Coffey, Charles W. .
MEDICAL PHYSICS, 2008, 35 (03) :1135-1144
[7]   Dosimetric evaluation of the OneDose™ MOSFET for measuring kilovoltage imaging dose from image-guided radiotherapy procedures [J].
Ding, George X. ;
Coffey, Charles W. .
MEDICAL PHYSICS, 2010, 37 (09) :4880-4885
[8]   Cone-beam CT with megavoltage beams and an amorphous silicon electronic portal imaging device: Potential for verification of radiotherapy of lung cancer [J].
Ford, EC ;
Chang, J ;
Mueller, K ;
Sidhu, K ;
Todor, D ;
Mageras, G ;
Yorke, E ;
Ling, CC ;
Amols, H .
MEDICAL PHYSICS, 2002, 29 (12) :2913-2924
[9]   Online image-guided intensity-modulated radiotherapy for prostate cancer: How much improvement can we expect? A theoretical assessment of clinical benefits and potential dose escalation by improving precision and accuracy of radiation delivery [J].
Ghilezan, M ;
Yan, D ;
Liang, J ;
Jaffray, D ;
Wong, J ;
Martinez, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (05) :1602-1610
[10]   A performance comparison of flat-panel imager-based MV and kV cone-beam CT [J].
Groh, BA ;
Siewerdsen, JH ;
Drake, DG ;
Wong, JW ;
Jaffray, DA .
MEDICAL PHYSICS, 2002, 29 (06) :967-975