Hybrid 3D T1-weighted gradient-echo sequence for fiducial marker detection and tumor delineation via magnetic resonance imaging in liver stereotactic body radiation therapy

被引:0
作者
Kato, Yutaka [1 ]
Kamomae, Takeshi [2 ]
Kumagai, Motoki [2 ]
Oie, Yumi [2 ]
Noguchi, Yumiko [1 ]
Okudaira, Kuniyasu [1 ]
Kawamura, Mariko [2 ]
Taoka, Toshiaki [2 ]
Naganawa, Shinji [2 ]
机构
[1] Nagoya Univ Hosp, Dept Radiol Technol, Nagoya, Aichi, Japan
[2] Nagoya Univ, Dept Radiol, Grad Sch Med, Nagoya, Aichi, Japan
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2022年 / 95卷
关键词
Liver stereotactic body radiation therapy; Gold fiducial marker; Marker detection; EOB-MRI; RADIOTHERAPY;
D O I
10.1016/j.ejmp.2022.01.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Gold fiducial markers are used to guide liver stereotactic body radiation therapy (SBRT) and are hard to detect by magnetic resonance imaging (MRI). In this study, the parameters of the three-dimensional Tl-weighted turbo gradient-echo (3D T1W-GRE) sequence were optimized for gold marker detection without degrading tumor delineation. Methods: Custom-made phantoms mimicking tumor and normal liver parenchyma were prepared and embedded with a gold marker. The 3D T1W-GRE was scanned by varying echo time (TE), bandwidth (BW), flip angle (FA), and base matrix size. The signal-to-noise ratio (SNR), contrast ratio (CR), and relative standard deviation (RSD) of the signal intensity in the area including the gold marker were evaluated, and the parameters were optimized accordingly. The modified 3D T1W-GRE (called HYBRID) was compared with the conventional T1W-GRE- and T2*-sequences in both phantom and clinical studies. In the clinical study of six patients with primary liver tumors, two observers visually assessed marker detection, tumor delineation, and overall image quality on a four-point scale. Results: In the phantom study, HYBRID showed significantly higher SNR and RSD than those of conventional T1W-GRE (P < 0.001). In the clinical study, HYBRID yielded significantly higher scores than conventional T1W-GRE did in terms of marker detection (P < 0.001). The scores of both sequences were not statistically different in terms of tumor delineation and overall image quality (P = 0.56 and P = 0.32). Conclusions: The proposed HYBRID sequence improved gold fiducial marker detection without degrading tumor delineation in MRI for SBRT of primary liver tumor.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 27 条
[1]   Qualitative Evaluation of Fiducial Markers for Radiotherapy Imaging [J].
Chan, Maria F. ;
Cohen, Gil'ad N. ;
Deasy, Joseph O. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2015, 14 (03) :298-304
[2]   Stereotactic Body Radiation Therapy for Liver Tumors: Current Status and Perspectives [J].
Doi, Hiroshi ;
Beppu, Naohito ;
Kitajima, Kazuhiro ;
Kuribayashi, Kozo .
ANTICANCER RESEARCH, 2018, 38 (02) :591-599
[3]   Visibility and artifacts of gold fiducial markers used for image guided radiation therapy of pancreatic cancer on MRI [J].
Gurney-Champion, Oliver J. ;
Lens, Eelco ;
van der Horst, Astrid ;
Houweling, Antonetta C. ;
Houweling, Antonetta C. ;
Klaassen, Remy ;
van Hooft, Jeanin E. ;
Stoker, Jaap ;
van Tienhoven, Geertjan ;
Nederveen, Aart J. ;
Bel, Arjan .
MEDICAL PHYSICS, 2015, 42 (05) :2638-2647
[4]   Diagnosis of Pathologically Early HCC with EOB-MRI: Experiences and Current Consensus [J].
Ichikawa, Tomoaki ;
Sano, Katsuhiro ;
Morisaka, Hiroyuki .
LIVER CANCER, 2014, 3 (02) :97-107
[5]   INTERNAL FIDUCIAL MARKERS AND SUSCEPTIBILITY EFFECTS IN MRI-SIMULATION AND MEASUREMENT OF SPATIAL ACCURACY [J].
Jonsson, Joakim H. ;
Garpebring, Anders ;
Karlsson, Magnus G. ;
Nyholm, Tufve .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (05) :1612-1618
[6]   Accuracy of device-specific 2D and 3D image distortion correction algorithms for magnetic resonance imaging of the head provided by a manufacturer [J].
Karger, Christian P. ;
Hoess, Angelika ;
Bendl, Rolf ;
Canda, Valer ;
Schad, Lothar .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (12) :N253-N261
[7]   Estimation of Liver Function Using T1 Mapping on Gd-EOB-DTPA-Enhanced Magnetic Resonance Imaging [J].
Katsube, Takashi ;
Okada, Masahiro ;
Kumano, Seishi ;
Hori, Masatoshi ;
Imaoka, Izumi ;
Ishii, Kazunari ;
Kudo, Masatoshi ;
Kitagaki, Hajime ;
Murakami, Takamichi .
INVESTIGATIVE RADIOLOGY, 2011, 46 (04) :277-283
[8]   Will Gd-EOB-MRI Change the Diagnostic Algorithm in Hepatocellular Carcinoma? [J].
Kudo, Masatoshi .
ONCOLOGY, 2010, 78 :87-93
[9]   Stereotactic Body Radiation Therapy for Primary and Metastatic Liver Tumors [J].
Liu, Erqi ;
Stenmark, Matthew H. ;
Schipper, Matthew J. ;
Balter, James M. ;
Kessler, Marc L. ;
Caoili, Elaine M. ;
Lee, Oliver E. ;
Ben-Josef, Edgar ;
Lawrence, Theodore S. ;
Feng, Mary .
TRANSLATIONAL ONCOLOGY, 2013, 6 (04) :442-446
[10]   Enhancement Effects of Hepatic Dynamic MR Imaging at 3.0 T and 1.5 T Using Gadoxetic Acid in a Phantom Study: Comparison with Gadopentetate Dimeglumine [J].
Morita, Kosuke ;
Namimoto, Tomohiro ;
Awai, Kazuo ;
Komi, Masanori ;
Hashida, Masahiro ;
Tsuji, Takashi ;
Hirai, Toshinori ;
Yamashita, Yasuyuki .
MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (01) :213-218