BEDVH-A method for evaluating biologically effective dose volume histograms: Application to eye plaque brachytherapy implants

被引:22
作者
Gagne, Nolan L. [1 ]
Leonard, Kara L. [1 ]
Huber, Kathryn E. [1 ]
Mignano, John E. [1 ]
Duker, Jay S. [2 ]
Laver, Nora V. [2 ,3 ]
Rivard, Mark J. [1 ]
机构
[1] Tufts Univ, Sch Med, Dept Radiat Oncol, Boston, MA 02111 USA
[2] Tufts Univ, Sch Med, Dept Ophthalmol, Boston, MA 02111 USA
[3] Tufts Univ, Sch Med, Dept Pathol, Boston, MA 02111 USA
关键词
eye plaque; radiobiology; biologically effective dose; INTRAOCULAR TUMORS; RADIATION-THERAPY; OCULAR MELANOMA; UVEAL MELANOMA; I-125; JUXTAPAPILLARY; DISTRIBUTIONS; RADIOTHERAPY; PD-103; DESIGN;
D O I
10.1118/1.3679010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A method is introduced to examine the influence of implant duration T, radionuclide, and radiobiological parameters on the biologically effective dose (BED) throughout the entire volume of regions of interest for episcleral brachytherapy using available radionuclides. This method is employed to evaluate a particular eye plaque brachytherapy implant in a radiobiological context. Methods: A reference eye geometry and 16 mm COMS eye plaque loaded with Pd-103, I-125, or Cs-131 sources were examined with dose distributions accounting for plaque heterogeneities. For a standardized 7 day implant, doses to 90% of the tumor volume (D-TUMOR(90)) and 10% of the organ at risk volumes (D-OAR(10)) were calculated. The BED equation from Dale and Jones and published alpha/beta and mu parameters were incorporated with dose volume histograms (DVHs) for various T values such as T = 7 days (i.e., (TUMORBED10)-B-7 and (OARBED10)-B-7). By calculating BED throughout the volumes, biologically effective dose volume histograms (BEDVHs) were developed for tumor and OARs. Influence of T, radionuclide choice, and radiobiological parameters on TUMORBEDVH and OARBEDVH were examined. The nominal dose was scaled for shorter implants to achieve biological equivalence. Results: D-TUMOR(90) values were 102, 112, and 110 Gy for Pd-103, I-125, and Cs-131, respectively. Corresponding (TUMORBED10)-B-7 values were 124, 140, and 138 Gy, respectively. As T decreased from 7 to 0.01 days, the isobiologically effective prescription dose decreased by a factor of three. As expected, (TUMORBEDVH)-B-7 did not significantly change as a function of radionuclide half-life but varied by 10% due to radionuclide dose distribution. Variations in reported radiobiological parameters caused (TUMORBED10)-B-7 to deviate by up to 46%. Over the range of (OAR) alpha/beta values, (OARBED10)-B-7 varied by up to 41%, 3.1%, and 1.4% for the lens, optic nerve, and lacrimal gland, respectively. Conclusions: BEDVH permits evaluation of the relative biological effectiveness for brachytherapy implants. For eye plaques, TUMORBEDVH and OARBEDVH were sensitive to implant duration, which may be manipulated to affect outcomes. (C) 2012 American Association of Physicists in Medicine. [DOI: 10.1118/1.3679010]
引用
收藏
页码:976 / 983
页数:8
相关论文
共 27 条
[1]  
[Anonymous], 2003, MCNP GEN MONTE CARLO, V1.
[2]   Design and dosimetric considerations of a modified COMS plaque: The reusable "seed-guide" insert [J].
Astrahan, MA ;
Szechter, A ;
Finger, PT .
MEDICAL PHYSICS, 2005, 32 (08) :2706-2716
[3]   A radiobiological investigation on dose and dose rate for permanent implant brachytherapy of breast using 125I or 103Pd sources [J].
Baltas, Dimos ;
Lymperopoulou, Georgia ;
Loffler, Edgar ;
Mavroidis, Panayiotis .
MEDICAL PHYSICS, 2010, 37 (06) :2572-2586
[5]   The clinical radiobiology of brachytherapy [J].
Dale, RG ;
Jones, B .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (845) :465-483
[6]  
DIENERWEST M, 1993, CONTROL CLIN TRIALS, V14, P362
[7]   Finger's "slotted'' eye plaque for radiation therapy: treatment of juxtapapillary and circumpapillary intraocular tumours [J].
Finger, Paul T. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (07) :891-894
[8]   Palladium-103 Ophthalmic Plaque Radiation Therapy for Choroidal Melanoma: 400 Treated Patients [J].
Finger, Paul T. ;
Chin, Kimberly J. ;
Duvall, Greg .
OPHTHALMOLOGY, 2009, 116 (04) :790-796
[9]   The COMS randomized trial of iodine 125 brachytherapy for choroidal melanoma - IV. Local treatment failure and enucleation in the first 5 years after brachytherapy. COMS report No. 19 [J].
Jampol, LM ;
Moy, CS ;
Murray, TG ;
Reynolds, SM ;
Albert, DM ;
Schachat, AP ;
Diddie, KR ;
Engstrom, RE ;
Finger, PT ;
Hovland, KR ;
Joffe, L ;
Olsen, KR ;
Wells, CG .
OPHTHALMOLOGY, 2002, 109 (12) :2197-2206
[10]   Biological-effective versus conventional dose volume histograms correlated with late genitourinary and gastrointestinal toxicity after external beam radiotherapy for prostate cancer: a matched pair analysis [J].
Jani, AB ;
Hand, CM ;
Pelizzari, CA ;
Roeske, JC ;
Krauz, L ;
Vijayakumar, S .
BMC CANCER, 2003, 3 (1)