Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station

被引:89
作者
Causey, Marlin Wayne [1 ]
Miller, Seth [1 ]
Foster, Andrew [2 ]
Beekley, Alec [1 ]
Zenger, David [2 ]
Martin, Matthew [1 ]
机构
[1] Madigan Army Med Ctr, Dept Surg, Tacoma, WA 98431 USA
[2] Madigan Army Med Ctr, Dept Anesthesia, Tacoma, WA 98431 USA
关键词
Hemoglobin; Noninvasive monitor; Masimo Radical-7; Hemorrhage; HAND ISCHEMIA;
D O I
10.1016/j.amjsurg.2011.01.020
中图分类号
R61 [外科手术学];
学科分类号
摘要
BACKGROUND: Hemoglobin levels must be obtained through blood draws, which are invasive, time-consuming, and provide only 1 data point at a time rather than continuous measurements. The Masimo Radical-7 SpHb Station (Masimo Corporation, Irvine, CA) has been shown by its manufacturers to provide accurate noninvasive hemoglobin measurements in physiologically normal patients. The objective of this study was to validate noninvasive hemoglobin measurements using the Masimo Radical-7 device. METHODS: Data were prospectively collected in 2 cohorts of patients: major operations requiring hemodynamic monitoring (operating room [OR]) and critically ill patients (intensive care unit [ICU]). Noninvasive hemoglobin measurements (SpHb) were recorded and were then compared with laboratory hemoglobin measurements. RESULTS: Data were collected on 60 patients (OR = 25 and ICU = 45). The overall correlation of the Masimo SpHb and the laboratory Hb was .77 (P < .001) in the OR group with a mean difference of .29 g/dL (95% confidence interval [CI], .08-.49). The overall correlation in the ICU group was .67 (P < .001) with a mean difference of .05 g/dL (95% CI, -.22 to -.31). CONCLUSIONS: Noninvasive hemoglobin monitoring is a new technology that correlated with laboratory values and supports the continued study of noninvasive hemoglobin monitoring. (C) 2011 Elsevier Inc. All rights reserved.
引用
收藏
页码:590 / 596
页数:7
相关论文
共 18 条
[1]  
[Anonymous], 2002, ANESTHESIOL CLIN N A
[2]  
Carlson K A, 1993, Anesthesiol Rev, V20, P173
[3]   Noninvasive monitoring of red blood cell transfusion in very low birthweight infants using diffuse optical spectroscopy [J].
Cerussi, A ;
Van Woerkom, R ;
Waffarn, F ;
Tromberg, B .
JOURNAL OF BIOMEDICAL OPTICS, 2005, 10 (05)
[4]   MEASUREMENT OF HEMOGLOBIN FLOW AND BLOOD-FLOW BY NEAR-INFRARED SPECTROSCOPY [J].
EDWARDS, AD ;
RICHARDSON, C ;
VANDERZEE, P ;
ELWELL, C ;
WYATT, JS ;
COPE, M ;
DELPY, DT ;
REYNOLDS, EOR .
JOURNAL OF APPLIED PHYSIOLOGY, 1993, 75 (04) :1884-1889
[5]   Continuous, noninvasive monitoring of total hemoglobin concentration by an optoacoustic technique [J].
Esenaliev, RO ;
Petrov, YY ;
Hartrumpf, O ;
Deyo, DJ ;
Prough, DS .
APPLIED OPTICS, 2004, 43 (17) :3401-3407
[6]   Continuous monitoring of hematocrit values during apheresis for allogeneic peripheral blood stem cell collection [J].
Kajiume, T ;
Kawano, Y ;
Takaue, Y ;
Abe, T ;
Okamoto, Y ;
Makimoto, A ;
Watanabe, T ;
Sato, J ;
Yokobayashi, A ;
Kuroda, Y .
JOURNAL OF HEMATOTHERAPY, 1998, 7 (06) :493-497
[7]  
KASLER M, 1990, HNO, V38, P375
[8]  
Macknet M, 2007, ANESTHESIOLOGY, V107, pA1545
[9]  
*MAS, MAS TECHN B
[10]   Noninvasive optical, electrical, and acoustic methods of total hemoglobin determination [J].
McMurdy, John W. ;
Jay, Gregory D. ;
Suner, Selim ;
Crawford, Gregory .
CLINICAL CHEMISTRY, 2008, 54 (02) :264-272