Minimum MU optimization (MMO): an inverse optimization approach for the PBS minimum MU constraint

被引:32
作者
Gao, Hao [1 ]
Clasie, Benjamin [2 ,3 ]
Liu, Tian [1 ]
Lin, Yuting [1 ]
机构
[1] Emory Univ, Dept Radiat Oncol, Winship Canc Inst, Atlanta, GA 30322 USA
[2] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
[3] Harvard Med Sch, Boston, MA 02115 USA
关键词
proton therapy; pencil beam scanning (PBS); intensity modulated proton therapy (IMPT); alternating direction method of multipliers (ADMM); CONE-BEAM CT; ALGORITHM;
D O I
10.1088/1361-6560/ab2133
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The deliverability of proton pencil beam scanning (PBS) treatment plans is subject to the minimum monitor unit (MU) constraint. This work introduces an inverse optimization approach to enforce the minimum MU constraint on planned spots, for accurate delivery of the planned dose. We formulate the minimum MU problem as an inverse optimization problem that accounts for the minimum MU constraint, i.e. minimum MU optimization (MMO). The MMO minimizes the difference between planned dose and deliverable dose while simultaneously enforcing the minimum MU constraint. Owing to the minimum MU constraint, MMO is nonconvex. Iterative convex relaxations are applied so that a sequence of convex subproblems of MMO need to be solved. The solution algorithm to the convex subproblem is developed based on alternating direction method of multipliers (ADMM). The MMO is validated in comparison with a greedy reassignment (GR) algorithm, and the gamma-index results suggest MMO can provide more accurate deliverable plans than GR. A simple-to-implement ADMM based MMO is developed to deal with the minimum MU constraint for high-quality deliverable PBS treatment plans.
引用
收藏
页数:8
相关论文
共 28 条
[1]  
[Anonymous], FOUND TRENDS MACH LE
[2]   Cine Cone Beam CT Reconstruction Using Low-Rank Matrix Factorization: Algorithm and a Proof-of-Principle Study [J].
Cai, Jian-Feng ;
Jia, Xun ;
Gao, Hao ;
Jiang, Steve B. ;
Shen, Zuowei ;
Zhao, Hongkai .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (08) :1581-1591
[3]   Incorporating deliverable monitor unit constraints into spot intensity optimization in intensity-modulated proton therapy treatment planning [J].
Cao, Wenhua ;
Lim, Gino ;
Li, Xiaoqiang ;
Li, Yupeng ;
Zhu, X. Ronald ;
Zhang, Xiaodong .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (15) :5113-5125
[4]   PBS machine interlocks using EWMA [J].
Clasie, Benjamin M. ;
Kooy, Hanne M. ;
Flanz, Jacob B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (01) :400-412
[5]   Numerical solutions of the γ-index in two and three dimensions [J].
Clasie, Benjamin M. ;
Sharp, Gregory C. ;
Seco, Joao ;
Flanz, Jacob B. ;
Kooy, Hanne M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (21) :6981-6997
[6]  
Gao H., 2011, UCLA CAM REPORT, P11
[7]  
Gao HS, 2010, 2ND CONFERENCE ON KEY TECHNOLOGY OF HORTICULTURE (CKTH 2010), P100
[8]   Principal component reconstruction (PCR) for cine CBCT with motion learning from 2D fluoroscopy [J].
Gao, Hao ;
Zhang, Yawei ;
Ren, Lei ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2018, 45 (01) :167-177
[9]   Robust fluence map optimization via alternating direction method of multipliers with empirical parameter optimization [J].
Gao, Hao .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (07) :2838-2850
[10]   Limited-view multi-source quantitative photoacoustic tomography [J].
Gao, Hao ;
Feng, Jing ;
Song, Liang .
INVERSE PROBLEMS, 2015, 31 (06)