Background We compared the Omron 725 CIC device (Omron Healthcare Inc., Vernon Hills, Illinois, USA), which is designed to register the blood pressure on the arm, with a mercury sphygmomanometer. In addition, we evaluated the possible impact that this device might have on the decisions made in a hypertension clinic. Methods Patients (n=183) older than 18 years (range 18-84 years) with a wide range of systolic (87-197 mmHg) and diastolic (48-108 mmHg) blood pressures were included, Some of the standards of the Association for the Advancement of Medical Instrumentation and of the British Hypertension Society were used to evaluate the results of the automated device in clinical practice. Results Using Bland-Altman analysis, an underestimation of both measures was observed with the automated Omron 725 CIC device; the systolic pressure was 3.6 +/- 8.8 mmHg too low with a very wide range of - 13.7 to 20.9 and the diastolic pressure was also 4.4 +/- 6.3 mmHg too low with a range of - 8.1 to 16.9. Clinical decisions could have been changed in 24 of the 116 hypertensive patients (20.6%) if the readings of the automated device had been used instead of using the readings of a mercury sphygmomanometer. These could have included modifying the dosage or changing the medicine used. Conclusion The blood pressure measurements by Omron 725 CIC are different from those of blood pressure readings taken with a mercury sphygmomanometer and this could affect clinical decisions in the diagnosis and follow-up of a hypertensive patient in an office environment.
引用
收藏
页码:321 / 327
页数:7
相关论文
共 50 条
  • [1] Accuracy of the Omron RX-M, an automated blood pressure measuring device, measuring blood pressure at the wrist, according to a modified British Hypertension Society protocol
    Braam, RL
    Aslan, B
    Thien, T
    BLOOD PRESSURE MONITORING, 2004, 9 (01) : 25 - 30
  • [2] Accuracy of the OMRON M4 automatic blood pressure measuring device
    JE Naschitz
    L Loewenstein
    R Lewis
    D Keren
    L Gaitini
    A Tamir
    D Yeshurun
    Journal of Human Hypertension, 2000, 14 : 423 - 427
  • [3] Accuracy of the OMRON M4 automatic blood pressure measuring device
    Naschitz, JE
    Loewenstein, L
    Lewis, R
    Keren, D
    Gaitini, L
    Tamir, A
    Yeshurun, D
    JOURNAL OF HUMAN HYPERTENSION, 2000, 14 (07) : 423 - 427
  • [4] Accuracy of the Omron RX-M, an automated wrist blood pressure measuring device
    Braam, R
    Aslan, B
    Thien, T
    JOURNAL OF HYPERTENSION, 2003, 21 : S22 - S23
  • [5] Validation of the Omron® RX-3 blood pressure measuring device at the wrist level according to the international protocol of the European Society of Hypertension
    El-Assaad, MA
    Orobinskaya, L
    Topouchian, JA
    Asmar, RG
    AMERICAN JOURNAL OF HYPERTENSION, 2005, 18 (05) : 29A - 29A
  • [6] Evaluation of the OMRON HEM-907 automated blood pressure device: comparison with office and ambulatory blood pressure readings
    Woodcock-Smith, Jean
    Miles, Karen
    Senaratne, Nipuna
    Wilkinson, Ian
    McEniery, Carmel
    JOURNAL OF HUMAN HYPERTENSION, 2016, 30 (10) : 655 - 655
  • [7] Evaluation of the Omron HEM-907 automated blood pressure device: comparison with office and ambulatory blood pressure readings
    D'Sa, Lauren
    Senaratne, Nipuna
    Woodcock-Smith, Jean
    Miles, Karen M.
    Wilkinson, Ian B.
    McEniery, Carmel M.
    HYPERTENSION RESEARCH, 2019, 42 (01) : 52 - 58
  • [8] Evaluation of the Omron HEM-907 automated blood pressure device: comparison with office and ambulatory blood pressure readings
    Lauren D’Sa
    Nipuna Senaratne
    Jean Woodcock-Smith
    Karen M Miles
    Ian B Wilkinson
    Carmel M McEniery
    Hypertension Research, 2019, 42 : 52 - 58
  • [9] EVALUATION OF AN AUTOMATED BLOOD-PRESSURE MEASURING DEVICE
    HOSSACK, KF
    GROSS, B
    BRUCE, RA
    AUSTRALIAN AND NEW ZEALAND JOURNAL OF MEDICINE, 1980, 10 (03): : 388 - 389
  • [10] EVALUATION OF A COIN OPERATED BLOOD-PRESSURE MEASURING DEVICE
    WHIPPLE, IT
    BERKSON, DM
    SHIREMAN, L
    BROWN, MC
    RAYNOR, W
    SHEKELLE, RB
    CIRCULATION, 1978, 58 (04) : 94 - 94